Russell Butler1, Guillaume Gilbert2, and Kevin Whittingstall3
1University of Sherbrooke, Sherbrooke, QC, Canada, 2MR Clinical Science - Philips Healthcare, Markham, ON, Canada, 3Diagnostic Radiology, University of Sherbrooke, Sherbrooke, QC, Canada
Synopsis
Precise and accurate knowledge of EEG sensors relative to
underlying cortical tissue enhances simultaneous EEG-FMRI studies, but to date
no specialized sequence for providing these locations exists. We propose an
ultra-short echo time sequence (UTE) to highlight the plastic casing and wiring
of a 64 channel MR compatible EEG cap. We show that the UTE resolves electrode
components up to 6mm from the surface of the scalp, allowing to locate the
precise contact point of electrode with skin and direction of wire leading away
from the electrode in all subjects (n=8). Purpose:
To obtain accurate and precise EEG electrode locations with
minimal scan time and no additional experimental setup through a UTE sequence optimised
specifically to highlight materials from which the electrodes are composed (1).
Methods
Sequence parameters:
All
experiments were performed on a clinical 3T system (Ingenia, Philips
Healthcare, Best, the Netherlands), using a 32-channel head coil for signal
reception. Ultra-short TE (UTE) imaging was performed using an axial 3D
radial stack-of-stars sequence with parameters: repetition time (TR) = 8ms,
echo time (TE) = 0.14ms, flip angle = 10 degrees, field-of-view (FoV) = 240 mm
x 240 mm x 170 mm, spatial resolution = 2 mm x 2 mm x 2 mm, acquisition time =
3min29sec. To evaluate the impact of spatial resolution on the electrode
position identification process, the same sequence was also repeated on one
subject with isotropic spatial resolutions of 3 mm (acquisition time =
1min33sec), 2.5 mm (acquisition time = 2min14sec) and 1.5 mm (acquisition time =
6min11sec).Being a low-flip angle gradient-echo sequence, this UTE sequence
leads to a specific absorption rate (SAR) and a RF duty cycle that are very
comparable to ones obtained for a typical 3D T1-weighted gradient-echo
acquisition, thus making the UTE sequence safe for use with an EEG cap present.
An axial anatomical 3D T1-weighted gradient-echo sequence
was also performed with parameters: TR = 7.9ms, TE = 3.5ms, inversion time (TI)
= 950ms, flip angle = 8 degrees, FoV = 240 mm x 240 mm x 150 mm, spatial
resolution = 1 mm x 1 mm x 1 mm, SENSE factor =2.4, acquisition time =
6min01sec.
Image Processing:
UTE: Raw UTE
images were upsampled to 1mm isotropic. A head mask was defined using in-house
software, and scalp layers were created at distances 0-6mm from the surface of
the head mask by iteratively dilating the mask. Scalp layers were averaged in
increments of 2mm, to create three separate layers at 0-2mm, 2-4mm, and 4-6mm
from the scalp. The three layers were projected to a pancake view (1), and an
RGB image was created by combining the three layers (red=0-2mm, green=2-4mm,
blue=4-6mm) (figure 1).
T1: a similar procedure to the UTE was performed, except the layers were
defined from 0-3mm inwards from the surface of the scalp (red=2-3mm,
green=1-2mm, blue=0-1mm), in order to isolate the gel artifacts (2).
Labeling:
Hand
labeling was done on the pancake projection with a single mouse click by
centering a crosshair over the electrode location. After all electrodes have
been labeled, the coordinates are transformed back to volumetric space.
Results
Reproducibility: UTE resolves electrode locations in a subject
independent manner, while gel artifacts vary from subject to subject (figure 2).
The inter subject variability in gel artifacts on the T1 was not due to image
processing deficiencies, as visual inspection of the raw T1 showed that in one
subject the gel artifacts were almost entirely absent (figure 3).
Voxel sizes: all electrodes are visible from 1.5mm to
2.5mm, but become more difficult to resolve at 3mm (figure 4).
Validation: visual inspection of back-projected locations
show that hand labeled positions closely coincide with electrode locations in
MRI space (figure 5).
Discussion
While we are not the first to use MRI to resolve EEG electrode locations, the variability of the T1 gel artifacts in (2) may be due to a dephasing effect stemming from magnetic susceptibility differences at the gel-skin interface. If this is the case, the T1 gel artifacts would be expected to vary in a subject and scan specific manner, based on the amount of applied gel, subject hair thickness, subject skin composition, and possibly other parameters. Our method, on the other hand, directly images the EEG electrodes and hence provides electrode locations in a subject independent manner.
We tested four separate resolutions (3mm, 2.5mm, 2mm and 1.5mm isotropic), and while it is possible to see most electrodes even at 3mm isotropic, we suggest keeping resolution below 2.5mm in order to obtain the most precise coordinates possible. The 2.5mm resolution scan is 2 minutes and 14 seconds, which is a negligible time increase in the context of most EEG-FMRI studies.
In summary, a fast and
reliable method to resolve EEG electrode locations in a subject independent
manner with no additional setup time and minimal scan time is presented. The
entire pipeline, from acquisition to hand labeling takes less than 10 minutes. We have created an image processing script to enable hand labeling, and encourage the addition of UTE to all simultaneous EEG-FMRI protocols.
Acknowledgements
No acknowledgement found.References
Springer, Fabian, Martirosian, Petros, Schwenzer, Nina F, Szimtenings,
Michael, Kreisler, Peter, Claussen, Claus D, Schick, Fritz MD, Three-Dimensional
Ultrashort Echo Time Imaging of Solid Polymers on a 3-Tesla Whole-Body MRI
Scanner. Investigative Radiology, 2008.
de Munck, J.C., van Houdt, P.J., Verdaasdonk, R.M.,
Ossenblok, P.P.W., 2012. A semi-automatic method to determine electrode
positions and labels from gel artifacts in EEG/fMRI-studies. Neuroimage 59,
399–403. doi:10.1016/j.neuroimage.2011.07.021