Synopsis
Osteoporosis is a disease
characterized by loss of bone mineral density, caused by loss of the equilibrium
between osteogenesis and adipogenesis. In this work T2, T1 and ADC were measured
using a low-field NMR scanner, for the detection of bone marrow changes related
to osteoporosis. Results showed that this method is capable of significantly classifying
between bones of rats that were ovariectomized, ovariectomized and treated with
parathyroid hormone, and sham-operated rats.INTRODUCTION
Osteoporosis is
characterized by decreased bone mass that may lead to an increased risk of
fractures. In addition, osteoporosis and ageing is associated with a reciprocal
decrease of osteogenesis and an increase of adipogenesis in bone marrow, being
driven by respective key molecules Runx2 and PPARγ.
1 Parathyroid
Hormone (PTH), a bone anabolic agent, increases osteogenic cells, thereby
leading to a rapid increase in trabecular bone, in addition to adipocytes
number reduction, in bone marrow.
2-3 Osteoporosis is commonly
diagnosed by assessing Bone Mineral Density (BMD) with dual-energy X-ray
absorptiometry (DEXA).
4 This method is effective when significant loss
of bone is present. Although MRI scan is capable of quantifying the amount
of adipocytes in marrow, the procedure is costly. Here we present, a low-field
NMR measurement that is capable of detecting changes in adipogenic
population in the bone marrow during early stages of ovarectomy (OVX) induced
osteoporosis in rats. Moreover, OVX rats treated with PTH showed significant
decrease of adipogenesis correlated with osteogenesis of trabecular bone.
METHODS
NMR
scans were performed on an NMR-MOUSE (Mobile Universal Surface Explorer, Magritek,
NZ). The device consists of a permanent U-shaped magnet, resulting in B
0=0.3T, with highly flat sensitive
volume at a distance of 25 mm parallel to the magnet surface (XY plane), and with
a strong stray-field gradient of ~8T/m (see Figure 1). An NMR signal profile was
acquired along the radial axis of the bone, for definition of the scanning zones
(either the center or periphery of the marrow cavity, shown in Figure 2). Then,
T2 was acquired using a CPMG-like sequence with 200 averages, 700 echoes, TR=1800
ms, TE=65
, scan time ~ 6 min. An estimation
method based on statistical signal processing was applied to the acquired CPMG
train.
5 T1 was acquired using a saturation recovery pulse sequence
with 18 different recovery times spaced exponentially between 0-2200 ms with 36
averages, 350 echoes, TR=2500 ms, TE=65
, scan time ~ 27 min. ADC was acquired
using a static gradient stimulated echo with 14 different B-values between 50 and 1000 $$$\frac{s}{\mu m ^2}$$$ with 200 averages, 250 echoes, TR=400 ms,
TE=65
, scan time ~19 min.
The
ability of the low-field portable NMR scanner to detect bone marrow changes was
tested. A comparable time series experiment was performed on 3-month-old,
female Sprague Dawley rats (n=32), which were divided into two equal groups ovariectomized (OVX) and sham-operated (sham) rats. The femur and tibia from
both hind limbs were isolated and underwent ex-vivo NMR scans, BMD
measurements, and histology. In the first experiment the scans were performed in
the central zone of the marrow cavity at 1 week, 1.5, 3.5 and 4.5 months post operations.
A second experiment was performed aimed to
find whether treatment with PTH during five weeks, starting one week post operation prevented the adipogenic pathway that is observed in OVX and whether it can be detected
early by NMR scans in both the central and peripheral zones.
RESULTS
Significant changes in the center of the marrow cavity of OVX bones were observed after 3.5 and 4.5 months, relative to
sham operated rats. Two NMR relaxation times, T1 and ADC, decreased in the OVX bones, while T2 increased.
The significant separation of the groups revealed by a multidimensional analysis
justifies an automated classification for future measurements, as shown in
figure 4. In the PTH treatment
experiment significant differences are shown after only 1.5 months, when
measured in the peripheral zone. T2 values of OVX bones are significantly higher than PTH
treated and sham operated rats (see figure 5). There are some indications that
by analyzing separately femur and tibia bones, the latter allows a more
sensitive detection of changes, however it requires a broader sample group. These
results are in correlation with a drop in BMD value and decreased red bone marrow,
as shown in figure 6.
DISCUSSION
This
study suggests a potential use of a tabletop NMR device for detecting the early
bone marrow changes that occur as osteoporosis develops. Clinical translation
resulting in an affordable version of this device in the future could
facilitate continuous monitoring of treatment efficacy. Future work will focus
on expanding peripheral results, with additional NMR relaxation times and will
further examine varying effects
of the disease and PTH treatment on the femur and tibia
bones.
Acknowledgements
No acknowledgement found.References
[1] Kawai, M., de Paula, F. J., & Rosen,
C. J. (2012). New insights into osteoporosis: the bone–fat
connection. Journal of internal medicine, 272(4), 317-329. [ [2Lu, R., Wang, Q.,
Han, Y., Li, J., Yang, X. J., & Miao, D. (2014). Parathyroid hormone
administration improves bone marrow microenvironment and partially rescues
haematopoietic defects in bmi1-null mice. PloS one, 9(4), e93864. [3]
Hodsman, A. B., Bauer, D. C., Dempster, D. W., Dian, L., Hanley, D. A., Harris,
S. T., ... & Yuen, C. K. (2005). Parathyroid hormone and teriparatide for
the treatment of osteoporosis: a review of the evidence and suggested
guidelines for its use. Endocrine reviews, 26(5), 688-703. [4] Kanis,
J. A., Burlet, N., Cooper, C., Delmas, P. D., Reginster, J. Y., Borgstrom, F.,
& Rizzoli, R. (2008). European guidance for the diagnosis and management of
osteoporosis in postmenopausal women. Osteoporosis international, 19(4),
399-428. [5] Bergman, E., Yeredor, A., & Nevo, U. (2014). An
estimation method for improved extraction of the decay curve signal from
CPMG-like measurements with a unilateral scanner. Journal of Magnetic
Resonance, 245, 87-93.