Synopsis
Bone water exists in the form of free water
in the Haversian canals or lacunar-canalicular system, as well as bound water either loosely
bound to collagen or more tightly bound to mineral. Ultrashort echo time (UTE)
sequences with TEs as short as 8 µs can potentially detect signal from pore
water and loosely bound water. In this study, we introduce an approach for fast
volumetric mapping of bound and pore water content in vivo using a clinical 3T
MR scanner.Introduction
Bone
is a composite material consisting of mineral, collagen and water in a complex
hierarchical structure
1. Bone water exists in the form of free
water in the Haversian canals or lacunar-canalicular system, as well as bound water either loosely
bound to collagen or more tightly bound to mineral
2. Pore water
content provides a surrogate measure of cortical porosity, which is difficult
to measure directly in vivo due to the limited spatial resolution of most imaging techniques
3.
Loosely bound water content provides a surrogate measure of collagen content,
which is inaccessible with x-ray based techniques
4. Ultrashort echo
time (UTE) sequences with TEs as short as 8 µs can potentially detect signal
from pore water and loosely bound water
5-7. In this study, we
introduce an approach for fast volumetric mapping of bound and pore water
content in vivo using a clinical 3T MR scanner.
Methods
A 3D UTE Cones sequence (Figure 1) was implemented on a GE 3T Signa TwinSpeed MR scanner. The sequence was further combined with an
adiabatic inversion preparation pulse (IR-Cones) for selective imaging of bound
water
7. T
1 of pore water (T
1PW) can be
measured with dual-echo 3D Cones acquisitions with a series of TRs (TRs = 7.4,
10, 15, 20 ms) based on Fig.2 Eq. [1]. The T
1 of bound water (T
1BW)
can be measured with IR-Cones acquisitions with a series of TR/TI combinations
(e.g., TR/TI = 50/24; 100/45; 200/81; 300/110 ms) based on Fig 2. Eq. [2], on the
condition that each TR/TI combination satisfies the inversion and nulling of
pore water requirement (so that only bound water is detected)
8. Total
water concentration (WC
Total) can be measured by comparing Cones
signal intensity of bone with that of a water phantom (water doped with MnCl
2
with T
2* similar to that of bone) based on Fig 2. Eq. [3]. Bound water
concentration (WC
Bound) can be measured by comparing the IR-Cones signal intensity of
bone with that of the water phantom based on Fig 2. Eq. [4]. Pore water concentration
(WC
Pore) can be calculated as the difference between total and bound water
concentration based on Fig 2. Eq. [5]. η is a coil sensitivity correction. Typical imaging parameters included a flip angle of 20°,
a bandwidth of 250kHz, a FOV of 15cm, matrix of 192×192, slice thickness of 8cm, 10 slices, a total scan time of 20min including 7min for T
1CW
measurement, 5min for T
1PW measurement, 4 min for total
water content measurement and 4 min for bound water content measurement. Five
healthy volunteers were recruited for this feasibility study.
Results
Figure 2 IR-UTE imaging of the tibia-midshaft of a 27y healthy male
volunteer. Exponential fitting of the IR-Cones
signal with variable TR/TI combinations shows a short T1BW
of 137±7 ms for bound water.
Figure 3 shows volumetric imaging and mapping of total,
collagen-bound and pore water in the tibial midshaft of the same volunteer
using a clinical 3T scanner. Both Cones and IR-Cones sequences show high signal
for cortical bone and a water phantom. Further analysis shows single component
signal decay for the 3D IR-UTE signal (results not shown), confirming that only
collagen-bound water is detected. A mean bound water concentration of ~18% and
pore water concentration of ~4% was found in the tibial mid-shaft of the
healthy volunteers.
Figure 4 shows volumetric total and bound water mapping in ulna and radius of the forearm. Chemical shift effect leads to increased total water content, especially for the thin cortical cortex in the ulna. Higher spatial resolution and sampling bandwidth are expected to minimize this error.
Discussion and Conclusion
We have demonstrated that
collagen-bound water and pore water in cortical bone can be measured with 3D
Cones and IR-Cones sequences using a clinical whole-body scanner, with
correction of T
1PW measured with dual-echo Cones
acquisitions with variable TRs, and T
1CW measured with a series of
IR-Cones acquisitions with different TR/TI combinations. Volumetric mapping of total, collagen-bound and pore water concentration in vivo, including T
1CW measurement for collagen-bound water, can be achieved in less than 15 minutes. Further clinical studies
will be necessary to evaluate the diagnostic power of this method, including
studies of patients with osteoporosis, osteomalacia, osteopenia, Paget disease,
insufficiency fractures in the setting of biophosphonate therapy, raloxifene treatment, etc.
Acknowledgements
The authors acknowledge grant support from GE
Healthcare and NIH (1R01 AR062581-01A1, 1 R01 AR068987-01 and 1R21 AR063894-01A1).References
1. American Society for
Bone and Mineral Research ASBMR Bone Curriculum 2004.
http://depts.washington.edu/bonebio/ASBMRed/ASBMRed.html.
2. Wehrli
FW, Song HK, Saha PK, Wright AC. Quantitative MRI for the assessment of bone
structure and function. NMR Biomed 2006;19: 731-64.
3. Li
C, Seifert AC, Rad HS, Bhagat Y, Rajapakse CS, Sun W, Benny Lam SC, Wehrli FW.
Cortical Bone Water Concentration: Dependence of MR Imaging Measures on Age and
Pore Volume Fraction. Radiology 2014: 132585.
4. Cao
H, Nazarian A, Ackerman JL, Snyder BD, Rosenberg AE, Nazarian RM, Hrovat MI,
Dai G, Mintzopoulos D, Wu Y. Quantitative 31P NMR spectroscopy and 1H MRI
measurements of bone mineral and matrix density differentiate metabolic bone
diseases in rat models. Bone 2010; 46:1582-1590.
5. Biswas
R, Bae CW, Diaz E, Masuda K, Chung CB, Bydder GM, Du J. Ultrashort echo time
(UTE) imaging with bi-component analysis: bound and free water evaluation of
bovine cortical bone subject to sequential drying. Bone 2012; 50:749-755.
6. Horch R, Gochberg D, Nyman J, Does M. Clinically-compatible
MRI strategies for discriminating bound and pore water in cortical bone. Magn
Reson Med 2012; 68:1774-1784.
7. Carl
M, Bydder GM, Du J. UTE imaging with simultaneous water and fat signal
suppression using a time-efficient multispoke inversion recovery pulse
sequence. Magn Reson Med 2015; Aug 26.
8. Du J, Sheth V, He Q,
Carl M, Chen J, Corey-Bloom J, Bydder
GM. Measurement of T1 of the ultrashort T2* components in white matter of the
brain at 3T. PLOS One 2014, 9(8): e103296.