Synopsis
The use of non-conventional gradient waveforms has brought renewed interest in non-invasive estimation of the axon diameter from diffusion MRI. Using conventional single diffusion encoding (SDE) at clinical MRI scanners, axons smaller than the resolution limit (4-5 microns) were indistinguishable from each other. We here predict the resolution limit for arbitrary gradient waveforms in systems with (i) parallel axons and (ii) orientation dispersion. Results show that SDE is optimal for parallel fibers, but that multiple diffusion encodings (MDE) are preferred where there is orientation dispersion. With 300 mT/m and MDE, the resolution limit was 2.8 microns for dispersed fibers.Introduction
Estimation of the axon diameter from diffusion MRI (dMRI) may provide a valuable tool for non-invasive studies of brain connectivity [1]. However, accurate quantification of axon diameters in the human brain is challenging, since the MR signal curves are indistinguishable for axon with diameters below the resolution limit (Fig. 1) [2,3]. The purpose of this study was to develop a framework to estimate the resolution limit. To address the growing interest in non-conventional diffusion encodings that go beyond single diffusion encoding (SDE) [4], we generalised the framework to permit prediction of the resolution limit regardless of the gradient waveform.
Theory
We consider a simple model where the MR signal ($$$S$$$) originates only from water within parallel cylinders, representing axons. We define the resolution limit as the diameter ($$$d_{\rm{min}}$$$) for which $$$S$$$ becomes indistinguishable from that of an infinitesimally thin axon. Hence, the difference in signal between these cases ($$$\Delta{S}$$$) approaches the detection limit ($$$\sigma$$$) for $$$d_{\rm{min}}$$$: $$\Delta{S}=S(b\,|\,d_{\rm{min}})-S(b\,|\,d\,\rightarrow\,0)=\sigma\Rightarrow\Delta{S}=1-\exp\big(-bD_\bot(d_{\rm{min}})\big)\approx\,bD_\bot(d_{\rm{min}}),~~~(1)$$where $$$b$$$ is the b-value and $$$D_\bot$$$ is the apparent diffusivity perpendicular to the axons. The approximation is valid close to the resolution limit, where the attenuation is low. A novel contribution of this work is the notion that $$$D_\bot$$$ can be calculated for arbitrary gradient waveforms by utilising the low-frequency approximation of the diffusion spectrum, i.e., $$$D_\bot(\omega)\propto\omega^2$$$ [5], according to$$D_\bot(d)=k\,D_0^{-1}V_{\omega}\,d^4,~~~(2)$$where $$$k\,=\,(2^73^15^3)^{-1/2}$$$, $$$D_0$$$ is the bulk diffusivity, and $$$V_\omega$$$ is the variance of the encoding spectrum. Omitting the derivation, we found$$bV_\omega\,=\,\gamma^2\,\int\,g^2(t){\rm\,dt},~~~(3)$$where $$$g(t)$$$ is the gradient waveform. The resolution limit for parallel axons $$$d_{\rm{min}}^{\rm (para)}$$$ is now given from Eqs. 1-3 as$$d_{\rm{min}}^{\rm(para)}=\left(\frac{\sigma\,D_0}{k}\frac{1}{bV_\omega}\right)^{1/4}.$$In the case of full orientation dispersion [6], $$S(b|d)=\exp(-bD_\bot)\,h(A)$$where $$$h(A)=\sqrt{\pi/4}~{\rm{erf}}(A)/A$$$, $$$A^2=b(D_\parallel-D_\bot)$$$, and $$$D_\parallel$$$ is the axial diffusivity. In dispersed axons, $$$d_{\rm{min}}^{\rm(disp)}$$$ is thus given by $$d_{\rm{min}}^{\rm(disp)}=d_{\rm{min}}^{\rm(par)}\,h(A)^{-1/4},$$where we assume $$$D_\bot(d_{\rm{min}})\,\approx\,0$$$ and thus $$$h(A)$$$ becomes independent of $$$d$$$. Since $$$h(A)<1$$$, the resolution is worse in the dispersed case.
Methods
We first compared predictions of $$$D_\bot$$$ from Eq. 2 to results from Monte Carlo simulations [7]. By investigating a multitude of gradient waveforms, we concluded that multiple diffusion encoding (MDE) by square wave gradients yielded the lowest possible value of $$$d_{\rm{min}}$$$. We then searched for the number of diffusion encodings that minimised $$$d_{\rm{min}}$$$. The analysis was performed for parallel axons and for full orientation dispersion (assuming $$$D_\parallel=2$$$ μm
2/ms), for various gradient amplitudes and field strengths (3T and 7T). We assumed $$$\sigma=1\%$$$, which is a reasonable detection limit on a clinical MRI system. The waveform duration was 80 ms for 3T and 50 ms for 7T, which accounts for shorter T2 relaxation times at 7T.
Results
The low-frequency approximation is demonstrated in Fig. 2. Figure 3 shows $$$\Delta S$$$ for $$$d=3.5$$$ μm, versus the gradient oscillation frequency. Parallel axons show highest value of $$$\Delta S$$$ for the lowest frequency (i.e. SDE). For dispersed axons, $$$\Delta S$$$ is at its highest for a frequency of approximately 100 Hz. Figure 4 shows gradient waveforms that were optimised to minimise $$$d_{\rm{min}}$$$, with corresponding values of $$$d_{\rm{min}}$$$ shown in Figure 5. Waveforms with higher oscillation frequencies were preferred for dispersed axons. Such waveforms featured lower values of $$$b$$$. Orientation dispersion worsened the resolution limit by up to 50\%, and reduced the relative benefit of strong gradients (Table 1). 7T provides higher SNR than 3T, which translates to better resolution only if the gradient amplitude is the same as at 3T.
Discussion
We assessed the lowest axon diameter that can reliably be recovered from the signal attenuation of intra-axonal water. We label this diameter ’the resolution limit’. Our analysis allows the resolution limit to be assessed for arbitrary gradient waveforms, or conversely, as an approach to minimising the resolution limit by optimising the gradient waveform.
For parallel axons, we found that the SDE experiment yields a better resolution than more complex waveforms. However, axons are rarely parallel, not even in the corpus callosum [8], but rather run in tortuous configurations. In the presence of full orientation dispersion, the effective SNR decrease with higher $$$b$$$, since the intra-axonal water signal is attenuated proportional to the relative alignment of the axon to the direction of the diffusion encoding. Accordgingly, the resolution is at its optimum for lower b-values for dispersed compared to parallel axons. This finding is in agreement with [9]. Our framework provides valuable knowledge for the design of dMRI experiments, although we acknowledge that the model is an oversimplification [10,11].
In conclusion, we recommend MDE by oscillating square waves rather than SDE where there is axonal orientation dispersion, in order to obtain the best possible diameter resolution.
Acknowledgements
The authors acknowledge the NIH grants R01MH074794, P41EB015902, P41EB015898, and the Swedish Research Council (VR) grants 2012-3682, 2014-3910, and Swedish Foundation for Strategic Research (SSF) grant AM13-0090. References
1. Alexander DC, et al. (2010) Orientationally invariant indices of axon diameter and density from diffusion MRI. NeuroImage 52(4):1374–1389.
2. Nilsson M, Alexander DC (2012) Investigating tissue microstructure using diffusion MRI: How does the resolution limit of the axon diameter relate to the maximal gradient strength? Processings of the ISMRM, Melbourne. 3567.
3. Dyrby TB, Søgaard LV, Hall MG, Ptito M, Alexander DC (2012) Contrast and stability of the axon diameter index from microstructure imaging with diffusion MRI. Magn Reson Med. doi:10.1002/mrm.24501.
4. Shemesh N, et al. (2015) Conventions and nomenclature for double diffusion encoding NMR and MRI. Magn Reson Med. doi:10.1002/mrm.25901.
5. Stepisnik J, Lasic S, Mohoric A, Sersa I, Sepe A (2006) Spectral characterization of diffusion in porous media by the modulated gradient spin echo with CPMG sequence. J Magn Reson 182(2):195–199.
6. Eriksson S, Lasic S, Nilsson M, Westin C-F, Topgaard D (2015) NMR diffusion-encoding with axial symmetry and variable anisotropy: Distinguishing between prolate and oblate microscopic diffusion tensors with unknown orientation distribution. J Chem Phys 142(10):104201.
7. Hall M, Alexander DC (2009) Convergence and Parameter Choice for Monte-Carlo Simulations of Diffusion MRI. IEEE Trans Med Imaging 28(9):1354–1364.
8. Ronen I, et al. (2013) Microstructural organization of axons in the human corpus callosum quantified by diffusion-weighted magnetic resonance spectroscopy of N-acetylaspartate and post-mortem histology. Brain Struct Funct. doi:10.1007/s00429-013-0600-0.
9. Drobnjak I, Zhang H, Ianus A, Kaden E, Alexander DC (2015) PGSE, OGSE, and sensitivity to axon diameter in diffusion MRI: Insight from a simulation study. Magn Reson Med. doi:10.1002/mrm.25631.
10. Nilsson M, van Westen D, Ståhlberg F, Sundgren PC, Lätt J (2013) The role of tissue microstructure and water exchange in biophysical modelling of diffusion in white matter. Magn Reson Mater Phy 26(4):345–370.
11. Burcaw LM, Fieremans E, Novikov DS (2015) Mesoscopic structure of neuronal tracts from time-dependent diffusion. NeuroImage 114(C):18–37.