Charlotte Debus1, Ralf Floca2, Amir Abdollahi1, Jürgen Debus3, and Michael Ingrisch4
1Translational Radiation Oncology, German Cancer Research Center (DKFZ), Heidelberg, Germany, 2Software development for Integrated Diagnostics and Therapy, German Cancer Research Center (DKFZ), Heidelberg, Germany, 3Department of Radiology, University of Heidelberg Medical School, Heidelberg, Germany, 4Institute for Clinical Radiology, Ludwig-Maximilians-University Hospital, Munich, Germany
Synopsis
The two-compartment exchange model is a tracer-kinetic model that is defined by two coupled first-order differential equations. These can be solved analytically or by direct integration. In this simulation study, we compared both strategies for different parameter scenarios in synthetic 4D images. The sum of squared residuals was calculated either by numeric integration with the Runge-Kutta method or by numeric convolution.
The resulting parameter estimates were evaluated in terms of accuracy, precision and computational speed. Both approaches yield similar results in parameter determination, the convolution excelled in computational speed.Purpose
Parameter estimation
in dynamic contrast-enhanced MRI (DCE MRI) is usually performed by
non-linear least square fitting of a tracer-kinetic model to a
measured concentration-time curve. The two-compartment exchange model
(2CXM) describes the compartments „plasma“ and „interstitial
volume“ with fractional volume v
p
and v
e
and their exchange in terms of plasma flow F
p
and permeability-surface area product PS. The model function is
defined by a system of two coupled differential equations
1,
2. A closed-form analytic solution as
convolution of a biexponential residue function with the arterial
concentration (AIF derives from indicator dilution theory. Several
studies on model feasibility and comparison between different models
have been performed
3, 4, 5,
however, little attention was given to technical aspects of the
fitting routine
6
. Parameters are usually estimated by fitting the model function to measured data using standard non-linear least squares methods. The residuals can be calculated by either numeric convolution of the closed-form solution
3 or by directly solving the differential equations
2 (e.g. using the Runge Kutta
algorithm). We
studied both strategies in terms of accuracy, robustness and
computation speed by fitting simulated concentration time curves in
image space, and quantitative and qualitative evaluation of the
resulting parameter maps.
Methods
Time-resolved
concentration images with gaussian noise and contrast-to-noise ratio
CNR=300 (expressed as ratio between the maximum of the AIF and the standard deviation of the noise) were
simulated, using a measured arterial input function resulting from a
double-bolus injection (temporal resolution 2.11s, 169 timepoints, Ref. 7).
All curves were
fitted with an in-house written software module developed within the
Medical Imaging and Interaction Toolkit MITK8,
which uses the levenberg-marquard optimizer implementation of the
visual numeric library (VNL). The differential equations were solved
using the boost implementation of the Runge-Kutta method. For the
convolution approach, the arterial input function was interpolated
linearly. Parameter constraints were applied to assure model
consistency : 0<vp,ve<
1, vp+ve<1,
0<Fp<100
ml/min/100ml, 0<PS< 100 ml/min/100ml. The optimizer
configuration and start parameters were kept constant for all fits.
Accuracy and robustness
were evaluated on concentration-curves using 5 different parameter
combinations (Figure 1a). For each of these combinations, a
homogeneous 30x30 pixel concentration image of 169 time points with
random noise was simulated. The resulting fitted parameters were
compared to the input value by calculating the bias, e.g. the
difference between input value and the mean of the fitting result,
and the variance of the mean.
For illustration a
50x50 pixel concentration image of 169 timepoints with different
lettering for Fp
and vp
and constant PS, ve
was fitted with both approaches.
Results
The convolution and
the differential equation approach yielded similar results for a
given parameter combination (Figure 2). The mean estimates of Fp
compared to the true value showed no considerable deviation except for
low Fp.
(Figure 3). For the reference situation as well as low PS and low ve
the parameter estimates of PS and vp
were close to the original values. The fitting result for ve
yielded a poor estimate with many outliers. The mean and median
parameter estimates differed strongly. For
low values of Fp,
both fitting strategies yielded poor fit stability, large
uncertainties and many outliers.
The convolution approach
excelled in terms of computational time (Figure 4).
In the visual
approach (Figure 5) the text „ISMRM“ and „Singapore“ from the
input parameter maps could be reproduced by both fitting routines.
However, the value of vp
showed interference with parameter estimate of PS and ve
as fragments of the textline „Singapore“ could be observed in the
corresponding maps. PS showed good fit stability in the rest of the
image, whereas ve
again presented poor fitting results.
Conclusion
The convolution and
differential equation definition of the model function of the 2CXM
yield very similar results in terms of accuracy and precision.
Fitting with the closed form solution is superior in computational
time. Both approaches are limited in accuracy for situations with low
blood flow. The model parameter v
e
shows great instability and little reliability in all cases. This is
expected to improve considerably by usage of longer total acquisition
time. More parameter combinations need to be evaluated for a detailed
error estimation. Further studies should include variations in the
starting parameters as well as different configurations for the
optimizer.
Acknowledgements
No acknowledgement found.References
[1] Sourbron, S. P., and D. L. Buckley.
"Tracer kinetic modelling in MRI: estimating perfusion and
capillary permeability." Physics in medicine and biology
57.2 (2012): R1.
[2] Brix, Gunnar, et al. "Tracer
kinetic modelling of tumour angiogenesis based on dynamic
contrast-enhanced CT and MRI measurements." European journal
of nuclear medicine and molecular imaging 37.1 (2010): 30-51
[3] Sourbron, Steven, et al.
"Quantification of cerebral blood flow, cerebral blood volume,
and blood–brain-barrier leakage with DCE-MRI." Magnetic
Resonance in Medicine 62.1 (2009): 205-217.
[4] Luypaert, Robert, et al. "The
Akaike information criterion in DCE-MRI: Does it improve the
haemodynamic parameter estimates?." Physics in medicine and
biology 57.11 (2012): 3609.
[5] Brix, Gunnar, et al.
"Pharmacokinetic analysis of tissue microcirculation using
nested models: multimodel inference and parameter identifiability."
Medical physics 36.7 (2009): 2923-2933.
[6] Luypaert, R., et al. "Error
estimation for perfusion parameters obtained using the
two-compartment exchange model in dynamic contrast-enhanced MRI: a
simulation study." Physics in medicine and biology 55.21
(2010): 6431.
[7] Ingrisch, Michael, et al.
"Quantification of perfusion and permeability in multiple
sclerosis: dynamic contrast-enhanced MRI in 3D at 3T."
Investigative radiology 47.4 (2012): 252-258.
[8] Wolf, Ivo, et al. "The medical
imaging interaction toolkit (MITK): a toolkit facilitating the
creation of interactive software by extending VTK and ITK."
Medical Imaging 2004. International Society for Optics and
Photonics, 2004.