André Ahlgren1, Ronnie Wirestam1, Freddy Ståhlberg1,2,3, and Linda Knutsson1
1Department of Medical Radiation Physics, Lund University, Lund, Sweden, 2Department of Diagnostic Radiology, Lund University, Lund, Sweden, 3Lund University Bioimaging Center, Lund University, Lund, Sweden
Synopsis
Deconvolution is an
ill-posed and ill-conditioned inverse problem that often yields non-physiological residue functions in perfusion MRI. Deconvolution methods
based on Fourier transform or matrix decomposition often yield solutions with spurious
oscillations. Although the perfusion value, estimated from the peak of the
tissue impulse response function, may still be useful, any estimate that
depends on the actual shape of the residue function will be prone to errors. To
obtain physiologically reasonable residue functions in perfusion MRI, we
propose the use of Bézier curves, and demonstrate initial experiences from the
application to DSC-MRI in vivo data.Purpose
To propose and assess a new method for
model-free quantification in perfusion MRI.
Introduction
Estimation of haemodynamic properties using
bolus-tracking perfusion MRI is typically accomplished by deconvolution of the
tissue concentration curve ($$$C_t$$$) with an arterial input function (AIF) in
order to yield the tissue impulse response function (TRF), which is the residue
function scaled with tissue perfusion
, i.e., $$$TRF(t)=f_t\cdot
R(t)$$$
. The fact that
deconvolution of noisy signals is an ill-posed and
ill-conditioned problem has led to the proposal of many different deconvolution
methods in perfusion MRI. The most commonly used is the singular value
decomposition (SVD) based method1, often accompanied by fix or adaptive
truncation. To produce more physiologically plausible residue functions without
spurious oscillations, methods based on, e.g., Tikhonov regularization2,
stochastic processes3, and control point interpolation4 (CPI),
the latter two implemented in Bayesian frameworks, have also been proposed.
In this work, we introduce a deconvolution method based on Bézier curves5,
which are inherently continuous and smooth in general. The concept of Bézier
curves was developed by de Casteljau (at Citroën), and independently by Bézier
(at Rénault) during the 1960s, to produce smooth curves and surfaces for car
design. Bézier curves are also common in computer graphics, for example, in
vector graphics software. The proposed deconvolution method, here referred to
as ‘Bézier deconvolution’ (BzD), resembles the CPI method4, although
BzD employs control polygons rather than pure control points. This means that
the curve does not pass through all control points, although the curve is
always contained within the convex hull of the control points. Due to their
smooth characteristics and the possibility to control their shape, we believe
that Bézier curves could allow for physiologically realistic residue function
shapes with rather few parameters. In this work, we demonstrate initial
experiences from the application of BzD to dynamic susceptibility contrast MRI
(DSC-MRI) in vivo data.
Methods
A Bézier curve of the nth order is given by
$$\mathbf{B}_n(\tau)=\sum_{i=0}^n\left(\begin{array}{c}n\\ i\end{array}\right)\left(1-\tau\right)^{n-1}\tau^i\mathbf{P}_i$$
where $$$\tau \in [0,1]$$$ and $$$\mathbf{P}_i$$$ denotes the coordinates of control point $$$i$$$. Cubic (B3) and quartic (B4) Bézier curves (see Fig. 1) can thus be expressed as follows:
$$\mathbf{B}_3(\tau)=(1-\tau)^3\mathbf{P}_0+3(1-\tau)^2\tau\mathbf{P}_1+3(1-\tau)\tau^2\mathbf{P}_2+\tau^3\mathbf{P}_3$$
$$\mathbf{B}_4(\tau)=(1-\tau)^4\mathbf{P}_0+4(1-\tau)^3\tau\mathbf{P}_1+6(1-\tau)^2\tau^2 \mathbf{P}_2+4(1-\tau)\tau^3\mathbf{P}_3+\tau^4\mathbf{P}_4$$
$$$\mathbf{P}_0=\left\{0,1\right\}$$$ and $$$P_{n,y}=0$$$ were fixed to ensure $$$R(0)=1$$$ as well as bounded-input, bounded-output stability (see Fig. 1c). To obtain a non-negative monotonically decreasing one-to-one solution, the conditions $$$0\leq P_{i,y}\leq 1$$$ and $$$0\leq P_{i,x}\leq P_{n,x}$$$ were also required. This resulted in 5 and 7 parameters for the cubic and quartic Bézier curves, respectively. The solution was resampled at the time points of the measured data before convolution with the AIF. The algorithm was implemented in a Bayesian framework6, using a non-informative prior for the CBF.
For proof of concept, the BzD method was applied to DSC-MRI data from a healthy volunteer, acquired with the following protocol: Single-shot gradient-echo EPI, matrix=128×128, voxel size=1.72×1.72×5 mm3, 20 slices, FA=60º and acquisition time=1:30 min. The project was approved by the local ethics committee, and written informed consent was obtained. A global AIF ($$$C_{AIF}$$$) was selected automatically, and the TRF was estimated based on the following model: $$$C_t(t)=\kappa f_t\left[C_{AIF}(t+\delta)\otimes R(t)\right]$$$, where $$$\kappa$$$ is a hematocrit and tissue density scaling factor (set to 0.701) and $$$\delta$$$ is a parameter taking into account the delay between tissue and arterial input curves. Mean transit time (MTT) was estimated as the time-integral of the residue function.
Results
Figure 2 displays parametric maps from the in vivo analysis, obtained using BzD and oSVD. The CBF values were
higher with BzD, compared to oSVD. MTT and delay maps were substantially
different between BzD and oSVD, and oSVD produced generally longer MTT and
delay estimates. Cubic and quartic BzD produced very similar results overall.
Figure 3 shows density plots of all residue functions from the slice in Figure
2. BzD produced more realistic residue function shapes than oSVD.
Discussion and Conclusion
We
implemented a new deconvolution method based on Bézier curves for quantitative
perfusion MRI. The BzD method produced higher perfusion values than the commonly used oSVD method (which is prone to non-physiological residue function
shapes and underestimation due to truncation of singular values). The MTT and
delay maps were also markedly different between BzD and oSVD.
A systematic comparison between BzD and the
CPI method4 would be of considerable interest, to assess whether
control polygons show advantages over control points. Furthermore, realistic model-free residue function shapes can
also yield capillary transit time distributions, which can be used for
model-free estimation of oxygen extraction capacity and metabolic rate of
oxygen capacity7 (by application of the capillary transit time heterogeneity
model8).
Acknowledgements
This work was supported by the Swedish Research Council.References
[1] Wu et al. Magn Reson
Med 2003;50:164-174. [2] Calamante et al. Magn Reson Med 2003;50:1237-1247. [3]
Zanderigo et al. IEEE Trans Biomed Eng 2009;56:1287-1297. [4] Mehndiratta et
al. NeuroImage 2013;64:560-570. [5] Forrest. Comp J 1972;15:71-79. [6] Okell et
al. Magn Reson Med 2012;68:969-979. [7] Mehndiratta et al. Proc ISMRM 2014, #348.
[8] Jespersen et al. J Cereb Blood Flow Metab 2012;32:264-277.