Valentina Taviani1, Shreyas S. Vasanawala1, and Brian A. Hargreaves1
1Stanford University, Stanford, CA, United States
Synopsis
A diffusion-weighted
(DW) imaging method was developed to mitigate off-resonance-induced distortion
and signal loss, which are problematic for body applications. A 2D RF pulse is
used in place of the conventional spectral-spatial excitation used for DW spin
echo echo-planar imaging. In the presence of off-resonance, a narrow band of
frequencies is excited due to the different bandwidths between excitation and
refocusing pulses. By progressively shifting the center frequency, the whole
range of off-resonance can be excited and a composite image, corrected for
off-resonance-induced distortion, can be reconstructed by estimating the field
map from the spectral information. Purpose
B0 inhomogeneity is a major
challenge encountered in body diffusion-weighted (DW) echo-planar imaging (EPI),
leading to severe signal loss and distortion. The numerous air-tissue
interfaces in the torso, the difficulty to accurately shim large volumes and
the presence of fat often result in suboptimal image quality. The use of 2D RF
pulses in conjunction with parallel imaging (PI) for coronal whole-body DW-EPI
has been described
1. While the 2D spatial selectivity can both improve
PI performance by phase encoding along the direction with the largest variation
in coil sensitivities and reduce distortion by limiting the FOV to the anatomy
of interest, the low bandwidth of the 2D RF pulse makes this technique extremely
sensitive to off-resonance, resulting in severe signal loss
2. We
present a method that can uniformly excite the whole imaging volume while
correcting residual off-resonance-induced distortion in clinically acceptable
scan times. Phantom and volunteer experiments were performed to demonstrate
feasibility.
Theory
2D RF pulses can be used to control the extent of the excited volume along
both the slice-select and phase-encoding direction. Here, the blipped direction
during excitation corresponds to the slice-select direction during imaging, so
that only the main excitation lobe is refocused by the 180° refocusing pulse
3. Due to the low slice-select bandwidth of
the 2D RF pulse (220Hz) and the relatively high bandwidth of the refocusing
pulse (1.2kHz), off-resonance spins are shifted outside of the volume refocused
by the 180º, so that the combination of the 2 pulses effectively produces a 2D
spatial/1D spectral excitation of finite bandwidth. By progressively shifting
the center frequency, the whole range of off-resonance can be excited, as
demonstrated in Figure 1, where a 30Hz/cm shim gradient was used to simulate off-resonance.
Root-sum-of-squares (RSOS) combination of the spectral components restores signal
over the whole FOV but does not correct off-resonance-induced distortion.
However, knowledge of the spectral components on a pixel-by-pixel basis allows
estimation of the off-resonance field (i.e. field map), which, in turn, can be
used to correct off-resonance induced distortion (d=ΔB
0/prBW;
d=shift along phase-encode direction in pixels, prBW=pseudo receiver bandwidth
per pixel). A center of mass (COM) approach was used to estimate the field map
from the frequency “bin” images
4 (cfr. Figure 1).
Materials and Methods
Phantom experiments were performed to assess accuracy of the
estimated field map and effectiveness of the distortion correction algorithm. A
separate IDEAL acquisition was performed to obtain a reference field map (6 TEs;
TE1/DTE=0.9/0.7ms; flip angle=3º;
BW=111kHz; FOV=44cm; phase FOV=50%; 7mm slice thickness; matrix size=160×160; 16 slices). The known shape of the phantoms
was used to evaluate the performance of the distortion correction algorithm. Healthy
volunteers were scanned using the upper 20 elements of a 32-channel receive-only
phased-array coil (NeoCoil, Pewaukee,
WI) and an 8-channel receive-only neurovascular coil (GE Healthcare, Waukesha,
WI) to demonstrate feasibility. All imaging was performed at 3T (GE MR750,
Waukesha, WI).
Results and Discussion
Figure 2 shows RSOS and
multispectral (MS) reconstructions using IDEAL and estimated field maps. Both
field maps were able to successfully correct the distortion induced by the shim
gradient. Note that distortion in the original acquired images is reflected in
the COM field map, while the IDEAL-generated field is undistorted. In vivo
experiments showed similar off-resonance distributions across different
subjects for a specific anatomy. In all cases, 4-5 frequency bins were
sufficient to cover the range of off-resonance, which could be achieved with
scans under five minutes. Figure 3 shows on-resonance and composite MS images
of 2 axial abdominal slices. Note that in this case the frequency spread was
minimal in the first slice but large enough to cause significant signal loss, successfully
recovered by the MS reconstruction, in a different slice (arrow). Figure 4 and 5
show coronal and axial images of the brachial plexus, where conventional on-resonance
imaging was non-diagnostic. Overall, good suppression of on-resonance fat was
achieved in the MS images (cfr. RSOS, Figure 4), although some off-resonance fat
(arrow) remains due to errors in the estimated field map, that could be
corrected by acquiring more frequency bins while keeping scan time constant by
reducing the number of signal averages per bin.
Conclusion
We developed a DW-EPI method
to reduce off-resonance-induced signal loss and distortion, which are
especially problematic for body applications. The main limitation consists in
the longer duration of the acquisition (proportional to the number of acquired
bins). While a few bins are usually sufficient to cover the whole off-resonance
range, multi-slice capabilities can be added to the proposed method to extend
coverage/reduce scan time.
Acknowledgements
Dr. Maggie Fung for useful discussion; GE Healthcare; NIH P41-EB015891-18. References
1. Obara P, Taviani V, Loening AM, Iagaru A, Hargreaves BA, Vasanawala SS. Comparison between whole-body coronal and axial DWI performed during PET-MR. Proc. Intl. Soc. Mag. Reson. Med. 2015; 23:3996.
2. Fung M, Estkowski L, Xu D, Banerjee S, Worters P, Wu G, Skinner J, Bayram E. Coronal whole body diffusion imaging with 2D spatially selective excitation (FOCUS). Proc. Intl. Soc. Mag.
Reson. Med. 2014; 22:2238.
3.
Saritas
EU, Cunningham CH, Lee JH, Han ET, Nishimura DG. DWI of the spinal cord with
reduced FOV single-shot EPI. Magn Reson Med 2008; 60:468-473.
4. Koch KM, Brau AC, Chen W, Gold GE, Hargreaves BA, Koff M, McKinnon GC, Potter HG, King KF.
Imaging Near Metal with a MAVRIC-SEMAC Hybrid. Magn Reson Med 2011; 65:71-82.