JaeJin Cho1, Dongchan Kim1, Hyunseok Seo1, Kinam Kwon1, Seohee So1, and HyunWook Park1
1Korea Advanced Institute of Science and Technology (KAIST), Daejeon, Korea, Republic of
Synopsis
Blipped-CAIPI imaging is widely used for fast
imaging, which is one of the simultaneous multi-slice imaging methods. Conventional
water-fat separation methods can be combined with the blipped-CAIPI technique. However,
it results in the chemical-shift ghost artifact because fat signal on slightly
shifted position is exited in the slice-selection process. This geometric error
in slice-selection generates additional phase cycling, which causes the ghost
artifacts on each slice’s fat image. In this abstract, a SENSE-based water-fat
separation method is proposed, which considers the additional phase cycling on
fat signal and obtains more accurate water-fat separated images.Introduction
Blipped controlled aliasing parallel (blipped-CAIPI) imaging is widely
used for fast imaging, which is one of the simultaneous multi-slice (SMS)
imaging methods
[1]. Water-fat separation methods such as DIXON and iterative
decomposition of water and fat with echo asymmetry and least-squares estimation
[2] can be
combined with the blipped-CAIPI technique. However, it results in the chemical-shift ghost artifact
because fat signal on slightly shifted position is exited in the
slice-selection process. This geometric error in slice-selection generates
additional phase cycling, which causes the ghost artifacts on each slice’s fat
image. In this abstract, a sensitivity encoding (SENSE)-based water-fat
separation method is proposed, which considers the additional phase cycling on
fat signal and obtains more accurate water-fat separated images.
Method
A slice-selection process excites fat signal on
shifted position from the region of water signal due to chemical shift. The
geometrical slice-selection error can be calculated as follows.
\[z_{shift} = \frac{f_{fat}}{\gamma G_{ss}}\tag1\]
where zshift is the distance of fat shift from the water slice, ffat is the chemical
shift frequency between fat and water, γ is the gyromagnetic
ratio, and Gss is a slice-selection
gradient. This geometric error generates additional phase
cycling of fat signal by gradient blip of the blipped-CAIPI (Figure 1). This
additional phase cycling makes the ghost artifact on fat images in
phase-encoding direction. The additional phase is calculated as follows.
\[\alpha = 2\pi \gamma M_{blip}z_{shift}=\frac{2\pi f_{fat}M_{blip}}{G_{ss}}\tag2\]
where α is an additional phase, and Mblip is the moment by blip gradient.
Figure 2 shows the phase variation in k-space and
its point spread function (PSF) for two-slices blipped-CAIPI. Unwanted peak in
fat signal is generated on N/2-field of view (FOV) shifted position. Therefore,
the ghost of first slice’s fat signal is overlaid on the second slice’s fat
signal. The ghost artifacts cannot be separated by using conventional water-fat
separation methods. However, the additional phase α can be calculated
from equation (2), from which the ghost artifacts can be easily decomposed by
SENSE-based least square method.
\[\rho = \underset{\rho}{\operatorname{argmin}}{\parallel S-C\cdot \Gamma \cdot B \cdot A\cdot F_u \cdot \rho \parallel}\tag3\]
where ρ is the reconstructed signal intensity, S is the acquired signal, C is a coil sensitivity, Γ is the additional phase, B is the blipped phase for blipped-CAIPI, A is chemical shift, and Fu is the under-sampled Fourier transform.
Figure 3 shows a simulation results representing
the chemical shift effect in the blipped-CAIPI. This simulation is based on the
blipped-CAIPI echo planar imaging (EPI) sequence for two-slices SMS imaging. Water
and fat images are separated using the fat shift in phase-encoding direction caused
by the phase variation between water and fat, which is generated during echo
train. Because the distance of fat shift can be calculated, water-fat separated
images are well reconstructed by using the conventional parallel imaging, which
is referred as EPI chemical shift separation[3]. However, the conventional
-unknown EPI chemical shift separation method
cannot reconstruct the accurate water and fat images from the acquired image
due to chemical shift ghost artifact. The proposed method reconstructs each image
correctly.
Experiments
were conducted on a 3.0 T MRI scanner (Siemens Verio, Germany) and 32-channel
head coil. Blipped-CAIPI multi-shot fly-back EPI sequence was implemented for two-slices
SMS imaging with the imaging parameters of 4000ms/25ms/1560Hz/Px(TR/TEeffective/BWreadout), image size of 64×128, voxel size of 2.5×2.5×5.0mm3, the distance between slices of 20mm, and Gss of
4.95mT/m, which means α is
18.52°
.
Results
Figure 4 shows experimental results. For
comparison, multi-shot fly-back EPI sequence is used, and images are
reconstructed by EPI chemical shift separation method, which uses the phase
variation during echo train and coil sensitivity. Without considering
additional phase cycling, the chemical-shift ghost artifact in fat images is
shown as expected. The proposed reconstruction method separates water and fat
images correctly with the calculated α of
18.52°.
Discussion and Conclusion
Simulation
and phantom experiment show that there is the chemical-shift ghost artifact in fat
images from the conventional water-fat separation method. This artifact is generated
by additional phase cycling of fat signal. However, the additional phase cycling
generated by chemical shift can be calculated, and the correct water-fat
separated images can be acquired by the proposed SENSE-based reconstruction
method.
Acknowledgements
This research was partly supported by the Brain
Research Program through the National Research Foundation of Korea (NRF) funded
by the Ministry of Science, ICT & Future Planning (2014M3C7033999) and
Korea Health Technology R&D Project through the Korea Health Industry
Development Institute (KHIDI), funded by the Ministry of Health & Welfare,
Republic of Korea (grant number : HI14C1135).References
[1]
K. Setsompop et al. Blipped-controlled aliasing in parallel imaging for
simultaneous multislice echo planar imaging with reduced g-factor penalty. MRM. 2012;67(5):1210-1224.
[2]
S.B. Reader et al. Multicoil Dixon
Chemical Species Separation With an Iterative Least-Squares Estimation Method. MRM. 2004;51:35-45.
[3] P.J. Shin et
al. Chemical Shift Separation with Controlled Aliasing for Hyperpolarized
13C Metabolic Imaging. MRM.
2015;74:978-989.