Jan Sedlacik1, Andreas M. Frölich1, Johanna Spallek2, Nils D. Forkert3, Tobias D. Faizy1, Franziska Werner4,5, Tobias Knopp4,5, Dieter Krause2, Jens Fiehler1, and Jan-Hendrik Buhk1
1Neuroradiology, UKE, Hamburg, Germany, 2Product Development and Mechanical Engineering Design, TUHH, Hamburg, Germany, 3University of Calgary, Calgary, AB, Canada, 4Biomedical Imaging, UKE, Hamburg, Germany, 5Biomedical Imaging, TUHH, Hamburg, Germany
Synopsis
Magnetic
particle imaging (MPI) was compared with dynamic magnetic resonance
imaging (MRI) and dynamic subtraction angiography (DSA) in a
realistic 3D printed aneurysm model. All three methods clearly
depicted a distinct pulsatile flow pattern and a delayed contrast
agent outflow from the aneurysm. Despite the disadvantages of a much
lower temporal resolution of the dynamic MRI and the 2D projection of
the DSA, all three methods are valid tools for characterizing the
hemodynamics of aneurysms. Especially the radiation free, 3D, high
temporal resolution MPI method seems to be a very promising tool for
imaging and characterization of hemodynamics.Introduction
Magnetic
particle imaging (MPI) is capable of acquiring 3D datasets with high
temporal resolution
1, which may be especially beneficial
for in vivo hemodynamic imaging. The characterization of the
hemodynamics of aneurysms is of particular interest
2,
since treatment planning and follow-up diagnosis may benefit from
this new imaging technique. The purpose of this work was to compare
MPI with dynamic magnetic resonance imaging (MRI), and dynamic
subtraction angiography (DSA) in a realistic 3D printed aneurysm
model
3 and to evaluate the capabilities of the different
methods.
Materials
and Methods
The
3D printed aneurysm model was derived from a
static
3D subtraction angiography
of a patient
with an incidental Internal Carotid Artery (ICA) aneurysm of saccular
morphology (ca.5mm diameter). The model was printed with 254µm thick
layers of acrylonitrile butadiene styrene at fused deposition
modeling using the HP Designjet 3D printer and impregnated with
Nano-Seal (Jeln Imprägnierung, Schwalmstedt, Germany)4.
The aneurysm model was connected to a peristaltic pump, which was set
to deliver a physiological flow and pulsation rate of about
250mL/min and 70/s, respectively. However, due to the peristaltic
nature of the pump, the pulsation profile is not comparable with
physiology. 4D phase contrast flow quantification (4D pc-fq) and
dynamic MRI, i.e. time-resolved contrast enhanced angiography with
stochastic trajectories, was performed using a 7T Bruker Clinscan
small animal MRI. The 4D pc-fq was triggered with the pump pulsation
and measured over 4 hours to obtain sufficient spatial resolution
(500µm isotropic) and signal to noise ratio. The dynamic MRI
measurement was optimized for fast dynamic MRI acquisition (270ms)
while administering a bolus of 3mL 0.05mol(Gd-DOTA)/L with a rate of
1mL/s using a syringe pump and an angiographic catheter with 1mm
inner diameter. The tip of the catheter was placed close to the
aneurysm model (ca.5cm upstream) to reduce bolus dispersion.
The first commercially available
preclinical MPI scanner (Bruker/Philips)
was used to acquire 1mm isotropic 3D data with 21.5ms temporal
resolution while administering a bolus with
50mmol(Fe)/L
(MM4,
TOPASS GmbH, Berlin,
Germany)
similarly as for the dynamic MRI measurement. DSA was acquired using
a Philips Allura FD20 with a temporal resolution of 33.3ms during
bolus injection of 150mg(iodine)/mL
(Imeron)
similar as for the dynamic MRI and MPI measurements. Image post
processing and visualization was done with in
house written software using
Matlab.
Results
Distinct
pulsatile
flow as well as lower flow velocities and a vortex inside
the aneurysm were clearly detected using 4D pc-fq (Fig.1-3). Dynamic
MRI, MPI and DSA also showed a clear pulsation with higher signal or
attenuation, i.e. contrast agent concentration, during the low flow
pulsation phases as well as a delayed contrast agent outflow from the
aneurysm (Fig.1). Single frames around maximum contrast agent
concentration allow depicting the contrast agent passage through the
model for MPI and DSA but not for dynamic MRI (Fig.4).
Discussion
and Conclusion
4D
pc-fq enables a nearly perfect depiction and characterization of flow
patterns, which is very helpful for better understanding the
contrast agent dynamics of our aneurysm model. However, extreme long
scan times and averaging over thousands of pulsation cycles renders
clinical application impossible, where short dynamic real time
measurements are required. All three dynamic methods (MRI, MPI, and
DSA) showed the same distinct pulsation as detected with 4D pc-fq and
a delayed contrast agent outflow
from the aneurysm caused by the vortex inside the aneurysm.
The higher contrast agent concentration during the low flow pulsation
phase is caused by the constant bolus injection rate resulting in
less diluted contrast agent during the low flow and more diluted
contrast agent during high flow pulsation phases. The MPI even
depicts the secondary low flow
pulsation phase by a low broad signal maximum in between two
consecutive high sharp signal peaks. This additional fact
demonstrates the superior dynamic capabilities of the MPI method with
respect to dynamic MRI and DSA. However, all three methods were able
to detect the delayed contrast agent outflow from the aneurysm and
are, therefore, valid tools to characterize the hemodynamics of
aneurysms.
Acknowledgements
We wish to thank the German Research Foundation (DFG), grant no.
AD 125 / 5-4, and
the Forschungszentrum Medizintechnik Hamburg (fmthh) for financial
support and Philips Healthcare for the support and realization of the
“Hermann-Zeumer Research Laboratory” including a Philips
AlluraClarity Angiography system.References
1
Weizenecker J, Gleich B, Rahmer J, Dahnke H, Borgert J.
(2009)Three-dimensional real-time in vivo magnetic particle imaging.
Phys Med Biol; 54: L1-L10.
2 Jeong W, Rhee K. Hemodynamics of cerebral aneurysms: computational
analyses of aneurysm progress and treatment. Comput Math Methods Med.
2012;2012:782801.
3
Anderson JR, Thompson WL, Alkattan AK, Diaz O, Klucznik R, Zhang YJ,
Britz GW, Grossman RG, Karmonik C. Three-dimensional printing of
anatomically accurate, patient specific intracranial aneurysm models.
J Neurointerv Surg. 2015 Apr 10
4 Frölich AM, Spallek J, Brehmer L, Buhk JH, Krause D,
Fiehler J, Kemmling A. 3D Printing of Intracranial Aneurysms Using
Fused Deposition Modeling Offers Highly Accurate Replications. AJNR
Am J Neuroradiol. 2015 Aug 20.