Jiaxin Shao1, Shams Rashid1, Kim-Lien Nguyen2,3, and Peng Hu1,4
1UCLADepartment of Radiological Sciences, David Geffen School of Medicine, University of California, Los Angeles, CA, United States, 2Department of Medicine, Division of Cardiology, David Geffen School of Medicine, University of California, Los Angeles, CA, United States, 3Division of Cardiology, Veterans Affairs Greater Los Angeles Healthcare System, Los Angeles, CA, United States, 4Biomedical Physics Inter-Departmental Graduate Program, University of California, Los Angeles, CA, United States
Synopsis
Current
cardiac T1 mapping techniques, including the modified Look-Locker
inversion-recovery (MOLLI), cannot be used effectively in patients with implanted
cardiac devices due to large off-resonance induced by the device. To eliminate
the device-induced image artefacts, we developed a technique by modifying the
MOLLI sequence to use spoiled gradient echo readout and a wideband inversion
pulse, with a new acquisition scheme and T1 estimation algorithm. The
feasibility of our
new technique was tested in phantom studies and validated in eight healthy
volunteers and ten patients with implanted cardiac devices at 1.5 Tesla.Background
With
recent development of a wideband LGE CMR sequence (1), detection of focal fibrosis is
feasible in patients with implanted cardiac devices, such as implantable
cardioverter defibrillators (ICDs). One main drawback of LGE is its limited
ability to detect diffuse fibrosis. Many patients with cardiac devices have
non-ischemic cardiomyopathy resulting from conditions associated with diffuse
fibrosis. Therefore, myocardial T1 mapping techniques (2–4) may help further diagnose and
monitor the disease processes in these patients. Unfortunately, current
widely used cardiac T1 mapping
techniques, including the modified Look-Locker inversion-recovery (MOLLI) (5) sequence, cannot be used to
image patients with ICDs. The large off-resonance induced by the device and associated
banding artifacts when using bSSFP readouts limit the reliability and accuracy
of existing techniques. Accordingly, we sought to develop and validate an
improved myocardial T1 mapping technique that would mitigate the device-induced
image artefacts in patients with ICDs.
Methods
A MOLLI-based
pulse sequence, named Wideband-FLASH-MOLLI, was developed by incorporating a
fast low angle shot (FLASH) readout (6) and a
wideband inversion pulse that was previously developed for wideband LGE (1). Figure 1 demonstrates the
acquisition scheme of the proposed Wideband-FLASH-MOLLI and the bandwidth of
the wideband inversion pulse compared to the traditional inversion pulse used
in MOLLI. With Wideband-FLASH-MOLLI, ten FLASH images were acquired over 13
heart beats during a single breath-hold with a resolution of 1.9×2.5×8.0 mm3.
The Bloch equation simulation with slice profile correction (BLESSPC) T1
estimation algorithm, which was originally developed for FLASH readout at 3.0T (6), was used in our current
technique to reconstruct the Wideband-FLASH-MOLLI T1 maps. The bSSFP-MOLLI with
InSiL fitting (7) and FLASH-MOLLI with BLESSPC
fitting (6) was implemented for comparisons,
both of which uses the conventional inversion pulse and a 3-(3)-3-(3)-5
acquisition scheme.
Wideband-FLASH-MOLLI
was evaluated using eight phantoms and validated in eight healthy volunteers
and ten patients with ICDs using a 1.5T MR scanner (Avanto, Siemens Healthcare;
Erlangen, Germany). The effects of off-resonance frequency variation, heart
rate variation, and presence of ICD on T1 estimation accuracy were investigated
using phantom studies. To mimic device-induced image artifacts, an ICD
generator was attached to the body coil and close to the left shoulder of
healthy volunteers. The bSSFP-MOLLI and FLASH-MOLLI images were acquired in all
healthy volunteers before and after the ICD attachment and in three patients
with ICDs for comparisons.
The coefficient of
determination (R2) was used to determine the quality of the fit for
each pixel. T1 values with R2 < 0.95 were set to zero in T1 maps.
Results
Wideband-FLASH-MOLLI
generated consistent T1 values over a wide range of off-resonance frequencies
(Figure 2) and showed no dependence on heart rate variation. The maximum T1
estimation errors using wideband-FLASH-MOLLI with and without an ICD were less
than 3% for T1 range of 212ms - 1673ms. For all eight healthy volunteers
without ICD, the average myocardial T1 values in septal region measured using
Wideband-FLASH-MOLLI was close to that measured using bSSFP-MOLLI with InSiL
fitting (1173±24 ms vs. 1175±30 ms, p=0.8). No statistically significant
difference in T1 values was found between four different myocardial regions (septal,
anterior, lateral and inferior) of left ventricular (LV) using
Wideband-FLASH-MOLLI (p>0.2). After ICD attachment in healthy volunteers,
the Wideband-FLASH-MOLLI T1 values were not significantly changed in any of the
four myocardial regions (p>0.2).
Due to
the presence of an ICD, the magnitude images acquired using bSSFP-MOLLI and
FLASH-MOLLI showed hyper-intensity and dark band artifacts within the
myocardium (Figure 3-4). The dark bands in the bSSFP-MOLLI are expected
off-resonance banding related to bSSFP readouts. The dark bands in the
FLASH-MOLLI images occurred at the boundary of inverted and non-inverted
regions, presumably due to signal cancelations within the boundary pixels. In
contrast, device-induced artifacts were eliminated within the myocardial
regions of all healthy volunteers and patients with ICDs when using
Wideband-FLASH-MOLLI.
Figure 5
shows the pre- and post-contrast Wideband-FLASH-MOLLI T1 maps and the wideband
LGE image acquired in a patient with an ICD, who was referred for assessment of
cardiac amyloidosis. The average native T1 value within the entire LV myocardium
was 1336±86.6 ms, which is ~14% greater than the average myocardial T1 values
in healthy volunteers. This finding is concordant with those reported in the
literature for cases of amyloidosis.
Conclusion
This study demonstrates the
feasibility of using Wideband-FLASH-MOLLI to eliminate device-induced image
artifacts in patients with implanted cardiac devices and can be used for
accurate myocardial T1 mapping. Wideband-FLASH-MOLLI enables assessment of
diffuse fibrosis in patients with implanted cardiac devices.
Acknowledgements
NoneReferences
1. Rashid S, Rapacchi S, Vaseghi M, et al.: Improved late gadolinium enhancement MR imaging for patients with implanted cardiac devices. Radiology 2014; 270:269–74.
2. Burt JR, Zimmerman SL, Kamel IR, Halushka M, Bluemke DA: Myocardial T1 mapping: techniques and potential applications. Radiographics 2014; 34:377–95.
3. Iles LM, Ellims AH, Llewellyn H, et al.: Histological validation of cardiac magnetic resonance analysis of regional and diffuse interstitial myocardial fibrosis. Eur Heart J Cardiovasc Imaging 2015; 16:14–22.
4. Karamitsos TD, Piechnik SK, Banypersad SM, et al.: Noncontrast T1 mapping for the diagnosis of cardiac amyloidosis. JACC Cardiovasc Imaging 2013; 6:488–97.
5. Messroghli DR, Radjenovic A, Kozerke S, Higgins DM, Sivananthan MU, Ridgway JP: Modified Look-Locker inversion recovery (MOLLI) for high-resolution T1 mapping of the heart. Magn Reson Med 2004; 52:141–6.
6. Shao J, Rapacchi S, Nguyen K-L, Hu P: Myocardial T1 mapping at 3.0 tesla using an inversion recovery spoiled gradient echo readout and bloch equation simulation with slice profile correction (BLESSPC) T1 estimation algorithm. J Magn Reson Imaging 2015. doi: 10.1002/jmri.24999.
7. Shao J, Nguyen K-L, Natsuaki Y, Spottiswoode B, Hu P: Instantaneous signal loss simulation (InSiL): an improved algorithm for myocardial T1 mapping using the MOLLI sequence. J Magn Reson Imaging 2015; 41:721–9.
8. Rashid S, Rapacchi S, Shivkumar K, Plotnik A, Finn JP, Hu P: Modified wideband three-dimensional late gadolinium enhancement MRI for patients with implantable cardiac devices. Magn Reson Med 2015. doi: 10.1002/mrm.25601.