Matthew Restivo1, Alexander Raaijmakers1, Cornelis A.T. van den Berg1, Pedro Crespo-Valero2, Peter Luijten1, and Hans Hoogduin1
1Center for Imaging Sciences, University Medical Center Utrecht, Utrecht, Netherlands, 2Zurich Med Tech, Zurich, Switzerland
Synopsis
We propose a technique where we measure the
real S-matrix of the array/subject setup in-situ and then closely match it in simulation using circuit
co-simulation with a modified cost function. We show that by accurately
simulating coupling, the B1+ and thus the SAR can be better predicted using
FDTD simulations. Better pTx SAR predictions will ensure RF safety while
reducing the overly conservative pTx SAR predictions that are used currently.Introduction
Predicting peak local SAR (pSAR) in a standard MR experiment is a
difficult task due to a variety of factors including anatomical variation. Parallel
transmit (PTx) further complicates pSAR prediction because simulated SAR values
are also dependent on array coupling. While the effect of subject variability
on pSAR has been investigated1, there has been little attention to the
effect of array coupling on pSAR predictions. Due to all these unknowns, pSAR
predictions are often based on a conservative worst case setting. Understanding
how SAR is related to coupling, combined with more accurate subject-specific
body models, will reduce the overly conservative B1+ and duty cycle
restrictions on pTx arrays.
The current practice in PTx SAR prediction is to rely on numerical FDTD
simulations. Circuit co-simulation2,3 has been introduced as a way to quickly
achieve optimal tuning, matching, and coupling of an array in the simulation
environment with a given load. However, real physical array coupling is rarely
optimal and is also load dependent. PTx simulations where the coupling does not
match the physical coupling are, in most cases, incorrectly estimating pSAR. We
propose a technique to measure the real S-matrix of the in-situ array/subject
setup and then match it in simulation using circuit
co-simulation with a modified cost function.
Methods
The proposed method was tested at 7T using a 4-ch Tx/Rx head array
consisting of four rectangular loop coils distributed 90° apart (modeled in Fig.1).
The array was loaded with a 10 cm diameter spherical phantom with known dielectric
properties.
Directional couplers were installed in the transmit path of a Philips 7T
MRI. The couplers were calibrated to measure the forward and reflected
propagating waves at the coil connection plane. The S-matrix is computed from a
quick dynamic sequence where forward power is delivered to one channel at a
time and the reflections are monitored on all channels.
The experimental setup was modeled and simulated using Sim4Life (ZMT AG,
Zurich). For co-simulation, all capacitors and sources were simulated as
transmitting ports while the remaining ports were terminated with 50 ohms.
In co-simulation, the ideal capacitor values are found through
optimizing the S-matrix based on a target cost function. A typical cost
function for optimal array tuning is the 2-norm of the simulated S-matrix (i.e.
minimum reflected power). Here we define a modified cost function as the 2-norm
of the difference between the simulation output S-matrix ($$$S_{out}$$$) and
the measured S-matrix.
$$cost=\|S_{out}(C)-S_{measured}\|$$
The
optimization returns a vector of capacitor values, C, such that the complex
difference between measured and simulated S-matrix is minimized. Co-simulation
was implemented using python scripting in conjunction with Sim4Life.
Simulations
were compared to measured B1+ maps. B1+ was calculated using low flip angle gradient
echo (low FA GE) images for good dynamic range. The sequence used a
sufficiently long TR (50ms) and a 10° flip angle to ensure signal intensity is
proportional to B1+. B1- was divided out after determining B1- as the ratio of
a quadrature dual TR B1+ map4 and quadrature low FA GE image.
Results
The
realistic co-simulation method approximates the measured S-matrix with good
agreement - All entries of the simulated S-matrix differ from the measured
S-matrix by less than -19.5 dB (negligible difference in power coupling). In
contrast, using co-simulation to simulate an array tuned for minimum reflected
power produces a noticeably different S-matrix.
Fig.3 shows the measured B1+ maps compared to simulated B1+ maps using
both co-simulated S-matrices. The simulated B1+ maps include pSAR calculated
for the corresponding drive configuration.
Discussion and Conclusion
Although
the B1+ pattern looks relatively similar for each drive, there is a difference
in pSAR values of up to 43%, highlighting the effect of coupling on pSAR
prediction. Based on the similarity of the B1+ measurements compared to
realistically tuned simulation (Fig.3), we confirm that matching simulated and
physical array coupling tends to improve simulation accuracy.
In this example, the physical array is fairly well tuned (i.e. similar
to the minimum reflection case). However, manufacturing error or uneven loading
could potentially lead to less ideal coupling. To demonstrate this situation,
we performed the proposed method with an S-matrix measured during array
assembly, when one element was still poorly tuned (Fig.4a). This coupling
condition yields pSAR that exceed that of the ideally coupled array in one case
by over 40% (Fig.4b).
Using this modified circuit co-simulation method, SAR can be
re-evaluated before every scan session based on a measured S-matrix in under a
couple minutes. This method, used in conjunction with detailed, subject
specific body models1, would facilitate accurate pSAR predictions in pTx
experiments.
Acknowledgements
This work was supported by the Initial Training Network, HiMR, funded by the FP7 Marie Curie Actions of the European Commission (FP7-PEOPLE-2012-ITN-316716)References
1. Homann,
et. al., MRM 2011; 66(6):1767-76
2. Kozlov et. al., JMR 2009; 200(1):147-52
3. Beqiri et. al., MRM 2015; 74(5):1423-34
4. Yarnykh, MRM 2007; 57(1):192-200.