Janot Tokaya1, A.J.E. Raaijmakers1, J.F. Bakker2, P.R. Luijten1, and C.A.T. van den Berg1
1Imaging Division, UMC Utrecht, Utrecht, Netherlands, 2Medtronic, Eindhoven, Netherlands
Synopsis
Tissue heating induced by
sharply peaked scattered electric fields at the tip of elongated implants is a
severe safety hazard refraining patients with active implants from undergoing
MRI examinations. Transfer functions (TFs) are widely used in modern safety
standards to assess implant safety. Currently, dedicated setups are required to
determine TFs in challenging and time consuming experiments. We introduce a new
experimental technique based on the principle of reciprocity and exploiting the
ability to map induced currents with MRI. The proposed method can accurately
determine TFs with high spatial resolution in a single, quick and relatively
simple measurement. It furthermore has the potential to be applied in
heterogeneous media allowing safety assessment in more realistic scenarios
where the conventional methods become inapplicable.Introduction
The patient population with active implanted medical devices
has a relatively high probability of needing MRI investigations and is rapidly growing.
Tissue heating induced by sharply peaked scattered RF electric fields at the
tip of elongated implants is a severe safety hazard. The introduction of the
transfer function concept (Park et al$$$^1$$$) greatly expedited RF safety
assessment studies as impinging tangential electric fields assessed by
simulations could be combined with measurements of the transfer function (TF)
to determine the scattered electric field at the tip. Measurements of TFs are
widely used in modern safety standards (ISO TS 10974$$$^2$$$). Currently a
dedicated bench setup is required to apply, sequentially repositioned,
localized tangential electric field exposures along an implant and measure the
scattered field around the tip, called piecewise excitation method (PWE), which
is both experimentally challenging and time consuming.
Recently the reciprocate approach was
shown to be equally adequate in determining the TF$$$^3$$$. Contrary to the
conventional PWE method, the excitation occurs at the tip (by soldering it to a
coax cable with drive signal) and the resultant electric field distribution
along the structure displays the TF. We introduce a new approach to
measure the TF with this second method using MRI. We exploit
the ability to map induced currents on elongated implants using MRI. The
proposed method can accurately determine TFs with high spatial resolution in a
single relatively simple and time efficient measurement. It furthermore has the
potential to measure TFs of (curved) implanted devices in hetrogeneous media allowing experimental safety assessment in more realistic
scenarios, where conventional methods become inapplicable.
Theory
The relationship between the scattered electric field$$$\,\vec{E_s}(\vec{r})\,$$$in proximity of a tip and
the tangential incident field $$$E_{\text{tan}}\,$$$along an implant of length $$$L$$$, whose trajectory is parameterized
by $$$q,\,$$$can be generically written as$$$^1$$$,$$\vec{E_s}(\vec{r})=\int_{0}^{L}\vec{S}(q,\vec{r})E_{\text{tan} }(q)\text{d}q=\vec{E_\text{tip}}\int_{0}^{L}S(q)E_{\text{tan}}(q)\text{d}q.$$In this equation the complex quantity $$$S(q)\,$$$is known as the transfer function. The last
equality is valid because the spatial pattern$$$\,\vec{E_s}(\vec{r})\,$$$around the tip is independent on
the excitation along the implant.
By virtue of the principle of reciprocity an
applied electric field at the tip$$$\,\vec{E_i}(\vec{r})\,$$$and
measurements of the resultant $$$E_{\text{tan}}\,$$$along
the implant (or induced current $$$I(q)\,$$$) allows determination of TFs.
Methods
Generic implants (insulated copper wires of 10, 20 and 30cm length), with TFs known from literature$$$^{1,3}\,$$$were placed in an elliptic ASTM phantom filled with CuSO$$$_4\,$$$doped saline (conductivity 0.47S/m and relative permittivity 78). These implants were locally excited by an open ended coax cable where the inner conductor was soldered to the insulated wire. The coax cable was connected to a 1.5T MR scanner (Achieva, Philips Healthcare) and used as a transceive antenna. Although the structures are non-matched (hence have large reflections), the induced currents on the wire generate $$$B_1^+\,$$$field distribution and, by reciprocity, also prescribe the $$$B_1^{-}\,$$$field distribution, i.e. the wire’s sensitivity pattern. The product is $$$B_1^+B_1^{-*}\,$$$is directly proportional to the MRI signal level (in the low-flip angle regime). 3D spoiled gradient echo MRI data, with 1mm isotropic resolution with 2$$$^o\,$$$flip angle was acquired for each structure. Because the spatial current trajectory is known upon visual inspection (see fig.2) the law of Biot-Savart can be applied to reconstruct the current profile along the wire.
The wire trajectories were discretized in $$$N\,$$$3mm straight line segments assuming piecewise constant currents $$$I_k\,$$$which transform the Biot-Savart integral equation in a Riemann sum. A least squares approximation minimizing the signal from the Biot-Savart law with respect to actually acquired MRI data was used to determine the current distribution in the implant and hence the TF. I.e.,$$\min_{\widetilde{I}}\left\| M_{\text{BS}}\cdot \widetilde{I}-d_{\text{meas}}\right\|_2^2,$$where $$$M_{\text{BS}}\,$$$is a matrix containing the $$$N\,$$$contributions of each $$$I_k\,$$$,contained in $$$\widetilde{I},\,$$$ to every voxel and $$$d_{\text{meas}}\,$$$ a vector with voxel signals.
To verify the measurement outcome, the TF of all structures was simulated using Sim4Life (ZMT, Zurich).The phase of the MRI data was corrected
for B$$$_0\,$$$and eddy current contributions by using dual echo
gradient sequences different readout polarities (Van
Lier$$$^4$$$).
Results
The equivalence between the coax excitation and the PWE method was demonstrated with simulations.The normalized TFs determined with the
PWE method showed excellent agreement with the ones determined from simulations of excitation by a coax cable and with literature (fig.4).
The MRI measured TFs are in good agreement
with simulations (fig.5). With the proposed method, the electric field at the
tip is determined with a relative error of maximally 8% for a worst case
incident electric field distribution. This research shows that TFs can accurately be determined using MRI techniques. This avoids elaborate bench
setups and extensive measurement procedures. Furthermore, it allows testing of
implants under relevant and more realistic MRI conditions.
Acknowledgements
This work was supported by the DeNeCor project being part of the ENIAC Joint Undertaking.References
1. S.M. Park, R. Kamondetdacha, J.A. Nyenhuis, Calculation of MRI-inducedheating of an implanted medical lead wire with an electric eld transfer function. JMRI, 2007.
2.ISO/TS 10974, International Standards Organization Technical Specication,Requirements for the safety of magnetic resonance imaging for patients withan active implantable medical device., 2011.
3. S. Feng, R. Quing, W. Kainz, J. Chen, A technique to evaluate MRI-Induced Electric Fields at the Ends of Practical Implanted Lead, IEEE, 2015.
4. A.L. van Lier, D.O. Brunner, K.P. Pruessmann,D.W. Klomp, P.R. Luijten, J.J. Lagendijk, C.A. van den Berg, B$$$_1^+$$$phase mapping at 7
T and its application for in vivo electrical conductivity mapping. Magn Reson Med. 2012.