Synopsis
Gradient impulse response functions were recently proposed to characterize MR
gradient systems with high accuracy. However, changes of the impulse response, e.g. due to thermal drifts, can limit its accuracy and hence applicability. To overcome this problem, we present a novel method
where the gradient response is continuously characterized during MR sequences
from repeatedly performed field probe measurements. The benefit of this method is
demonstrated by obtaining the continuous gradient output of MR sequences and first
imaging results are presented.Introduction
Gradient impulse response functions have recently been
introduced to characterize MR gradient systems with high accuracy (1).
This information can be employed to estimate gradient trajectories for enhanced image
reconstruction (2,3) and RF-pulse design (4),
or to tune the gradient system’s pre-emphasis settings. The underlying
assumption is that the system is linear and time-invariant. However, thermal
drifts of the system can cause a violation of the latter assumption
(5–7), thereby
limiting the accuracy/applicability of the method.
To overcome this problem, we present a novel method
where the gradient response is continuously characterized. Here, field
measurements are repeatedly performed during the MR sequence. From multiple field
measurements (harvest) and the known MR sequence gradients, the gradient
response is calculated. Frequency components that are not present in the MR sequence input are not characterized. Conversely, these components are not
required when utilizing the gradient response for the same sequence. The benefit of the method is demonstrated by obtaining the continuous gradient output
of MR sequences and first imaging results are presented.
Methods
All experiments were conducted on a 3T-MR system (Philips Healthcare, Netherlands). Eight 19F-based field probes (8) (T2=~2ms) were mounted to a head receive coil array (Fig.1left).
For probe excitation
and probe/coil data acquisition a dedicated RF-pulse generator and spectrometer were used.
The position of the probes were obtained in an initial calibration step (9). Thereafter, a 2D-SSh-EPI (resolution=(1.3mm)2, TE=61ms, TR=150ms, 20 slices) with 128
dynamics (duration=6:27min) was performed. During the MR sequence, the probes were repeatedly excited (TRprobe=6.005ms) followed by a readout of 500µs (Fig.1right). From the probes data
the gradient output during each readout was calculated (9), filtered to a bandwidth of ±50kHz and gridded to the same
time resolution (6.4µs) as the gradient input. The gradient response was subsequently determined
by solving the linear system
$$o=Ic$$
Here $$$o$$$ denotes a vector (Nx1) of multiple output
gradient readouts. $$$I$$$ denotes a matrix (NxM) with
each row holding a timeframe of the input waveform $$$i_{cont}$$$, each row being shifted
by the time relating to the corresponding sample in $$$o$$$. The vector $$$c$$$ (Mx1) denotes the unknown gradient response
(Fig.2). For each gradient response calculation, 1300 output readouts were used
(N=~50000, total harvesting time of 7.8s). The length of the gradient response
function was chosen to be 40ms (M=6025).
Finally, the continuous gradient output $$$o_{cont}$$$ was calculated by multiplying
a continuous input matrix $$$I_{cont}$$$ with the obtained response:
$$o_{cont}=I_{cont}c$$
For demonstration, the gradient responses were evaluated
for the first $$$c^{(first)}$$$ and last $$$c^{(last)}$$$ portion of the EPI sequence for the read gradient axis.
The method was repeated for a gradient echo sequence (TR=50ms,
TE=5ms, resolution=(1.4mm)2). Gridding based image reconstruction
was performed using the obtained k-space trajectory.
Results
The obtained continuous gradient output $$$o_{cont}$$$ matched the measured gradients $$$o$$$ up to the noise level (Fig.3), indicating the validity of the linearity and time-invariance assumption during the total acquisition time of $$$o$$$. $$$o_{cont}$$$ showed effects from eddy currents and fine mechanical vibrations, demonstrating the sensitivity of the method. The
response functions calculated from the beginning $$$c^{(first)}$$$ and the end $$$c^{(last)}$$$ of the EPI sequence, showed shifts of oscillatory
terms as well as a global broadening in the frequency domain (Fig.4left), which were also reflected in the gradient time courses $$$o_{cont}^{(first)}$$$ and $$$o_{cont}^{(last)}$$$ (Fig.4right). The obtained k-space data for the gradient echo scan allowed for a faithful
image reconstruction (Fig.5).
Discussion and
Conclusion
We introduced the concept of gradient response
harvesting. Unlike gradient response measurements using a dedicated calibration
sequence, the approach assumes/tolerates slow variations of the gradient response.
An important application is the possibility to obtain continuous
field data, which was also demonstrated in this work, including first image reconstruction results. Here, the approach
benefits from allowing gaps between probe data readouts. This drastically
reduces probe hardware requirements, which have so far prohibited alternative
continuous monitoring approaches (10) to
be performed simultaneously with the MR experiment.
The method can be straightforwardly extended to also account for gradient cross-terms. Other means
of calculating the gradient response, e.g. that include additional priors such exponential decay of eddy currents, may as well be employed. Moreover, the
temporal variability/update and required frequency resolution of the gradient responses are to be further investigated. Here, a combination with field feedback control (11) to address slow non-linear field deviations may be attractive.
Apart from obtaining continuous gradient/trajectory information, the
method may also be tested for real-time pre-emphasis tuning or
continuous system diagnostics.
The
accuracy and general applicability of this method offers the possibility to increase system performance and enhance
a wide range of MR applications.
Acknowledgements
No acknowledgement found.References
(1) Vannesjo SJ,
Haeberlin M, Kasper L, Pavan M, Wilm BJ, Barmet C, Pruessmann KP. Gradient
System Characterization by Impulse Response Measurements with a Dynamic Field
Camera. Magn. Reson. Med. Off. J. Soc. Magn. Reson. Med. Soc. Magn. Reson. Med.
2013;69:583–593.
(2) Vannesjo
SJ, Graedel NN, Kasper L, Gross S, Busch J, Haeberlin M, Barmet C, Pruessmann
KP. Image Reconstruction Using a Gradient Impulse Response Model for Trajectory
Prediction. Magn. Reson. Med. 2015,DOI10.1002/mrm.25841.
(3) Campbell-Washburn
AE, Xue H, Lederman RJ, Faranesh AZ, Hansen MS. Real-Time Distortion Correction
of Spiral and Echo Planar Images Using the Gradient System Impulse Response
Function. Magn. Reson. Med. 2015,DOI10.1002/mrm.25788.
(4) Cavusoglu
M, Dietrich B, Brunner D, Pruessmann K. Optimization of Parallel RF
Transmission Enabled by Concurrent Recording of RF and Gradient Fields. Proc.
23th Annu. Meet. ISMRM Tor. Can. 2015 p.:1819.
(5) Busch
J, Vannesjo SJ, Barmet C, Pruessmann KP, Kozerke S. Analysis of Temperature
Dependence of Background Phase Errors in Phase-Contrast Cardiovascular Magnetic
Resonance. J. Cardiovasc. Magn. Reson. 2014;16,DOI10.1186/s12968-014-0097-6.
(6) Kasper
L, Bollmann S, Vannesjo SJ, Gross S, Haeberlin M, Dietrich BE, Pruessmann KP.
Monitoring, Analysis, and Correction of Magnetic Field Fluctuations in Echo
Planar Imaging Time Series. Magn. Reson. Med. 2014
(7) Wilm
BJ, Nagy Z, Barmet C, Vannesjo SJ, Kasper L, Haeberlin M, Gross S, Dietrich BE,
Brunner DO, Schmid T, Pruessmann KP. Diffusion MRI with Concurrent Magnetic
Field Monitoring. Magn. Reson. Med. 2015
(8) De Zanche N, Barmet C,
Nordmeyer-Massner JA, Pruessmann KP. NMR Probes for Measuring
Magnetic Fields and Field Dynamics in MR Systems. Magn. Reson. Med. Off. J.
Soc. Magn. Reson. Med. Soc. Magn. Reson. Med. 2008;60:176–186.
(9) Barmet C, De Zanche N, Pruessmann
KP. Spatiotemporal Magnetic Field Monitoring for MR. Magn.
Reson. Med. Off. J. Soc. Magn. Reson. Med. Soc. Magn. Reson. Med.
2008;60:187–197.
(10) Dietrich
BE, Brunner DO, Wilm BJ, Barmet C, Gross S, Kasper L, Haeberlin M, Schmid T,
Vannesjo SJ, Pruessmann KP. A Field Camera for MR Sequence Monitoring and
System Analysis. Magn.
Reson. Med. 2015
(11) Duerst
Y, Wilm BJ, Dietrich BE, Vannesjo SJ, Barmet C, Schmid T, Brunner DO,
Pruessmann KP. Real-Time Feedback for Spatiotemporal Field Stabilization in MR
Systems. Magn.
Reson. Med. 2015;73:884–893.