Gradient response harvesting for continuous system characterization during MR sequences
Bertram J. Wilm1, Benjamin E. Dietrich1, Jonas Reber1, S. Johanna Vannesjo1, and Klaas P. Pruessmann1

1Institute for Biomedical Engineering, University of Zurich and ETH Zurich, Zurich, Switzerland

Synopsis

Gradient impulse response functions were recently proposed to characterize MR gradient systems with high accuracy. However, changes of the impulse response, e.g. due to thermal drifts, can limit its accuracy and hence applicability. To overcome this problem, we present a novel method where the gradient response is continuously characterized during MR sequences from repeatedly performed field probe measurements. The benefit of this method is demonstrated by obtaining the continuous gradient output of MR sequences and first imaging results are presented.

Introduction

Gradient impulse response functions have recently been introduced to characterize MR gradient systems with high accuracy (1). This information can be employed to estimate gradient trajectories for enhanced image reconstruction (2,3) and RF-pulse design (4), or to tune the gradient system’s pre-emphasis settings. The underlying assumption is that the system is linear and time-invariant. However, thermal drifts of the system can cause a violation of the latter assumption (5–7), thereby limiting the accuracy/applicability of the method.

To overcome this problem, we present a novel method where the gradient response is continuously characterized. Here, field measurements are repeatedly performed during the MR sequence. From multiple field measurements (harvest) and the known MR sequence gradients, the gradient response is calculated. Frequency components that are not present in the MR sequence input are not characterized. Conversely, these components are not required when utilizing the gradient response for the same sequence. The benefit of the method is demonstrated by obtaining the continuous gradient output of MR sequences and first imaging results are presented.

Methods

All experiments were conducted on a 3T-MR system (Philips Healthcare, Netherlands). Eight 19F-based field probes (8) (T2=~2ms) were mounted to a head receive coil array (Fig.1left). For probe excitation and probe/coil data acquisition a dedicated RF-pulse generator and spectrometer were used. The position of the probes were obtained in an initial calibration step (9). Thereafter, a 2D-SSh-EPI (resolution=(1.3mm)2, TE=61ms, TR=150ms, 20 slices) with 128 dynamics (duration=6:27min) was performed. During the MR sequence, the probes were repeatedly excited (TRprobe=6.005ms) followed by a readout of 500µs (Fig.1right). From the probes data the gradient output during each readout was calculated (9), filtered to a bandwidth of ±50kHz and gridded to the same time resolution (6.4µs) as the gradient input. The gradient response was subsequently determined by solving the linear system

$$o=Ic$$

Here $$$o$$$ denotes a vector (Nx1) of multiple output gradient readouts. $$$I$$$ denotes a matrix (NxM) with each row holding a timeframe of the input waveform $$$i_{cont}$$$, each row being shifted by the time relating to the corresponding sample in $$$o$$$. The vector $$$c$$$ (Mx1) denotes the unknown gradient response (Fig.2). For each gradient response calculation, 1300 output readouts were used (N=~50000, total harvesting time of 7.8s). The length of the gradient response function was chosen to be 40ms (M=6025).

Finally, the continuous gradient output $$$o_{cont}$$$ was calculated by multiplying a continuous input matrix $$$I_{cont}$$$ with the obtained response:

$$o_{cont}=I_{cont}c$$

For demonstration, the gradient responses were evaluated for the first $$$c^{(first)}$$$ and last $$$c^{(last)}$$$ portion of the EPI sequence for the read gradient axis.

The method was repeated for a gradient echo sequence (TR=50ms, TE=5ms, resolution=(1.4mm)2). Gridding based image reconstruction was performed using the obtained k-space trajectory.

Results

The obtained continuous gradient output $$$o_{cont}$$$ matched the measured gradients $$$o$$$ up to the noise level (Fig.3), indicating the validity of the linearity and time-invariance assumption during the total acquisition time of $$$o$$$. $$$o_{cont}$$$ showed effects from eddy currents and fine mechanical vibrations, demonstrating the sensitivity of the method. The response functions calculated from the beginning $$$c^{(first)}$$$ and the end $$$c^{(last)}$$$ of the EPI sequence, showed shifts of oscillatory terms as well as a global broadening in the frequency domain (Fig.4left), which were also reflected in the gradient time courses $$$o_{cont}^{(first)}$$$ and $$$o_{cont}^{(last)}$$$ (Fig.4right). The obtained k-space data for the gradient echo scan allowed for a faithful image reconstruction (Fig.5).

Discussion and Conclusion

We introduced the concept of gradient response harvesting. Unlike gradient response measurements using a dedicated calibration sequence, the approach assumes/tolerates slow variations of the gradient response.

An important application is the possibility to obtain continuous field data, which was also demonstrated in this work, including first image reconstruction results. Here, the approach benefits from allowing gaps between probe data readouts. This drastically reduces probe hardware requirements, which have so far prohibited alternative continuous monitoring approaches (10) to be performed simultaneously with the MR experiment.

The method can be straightforwardly extended to also account for gradient cross-terms. Other means of calculating the gradient response, e.g. that include additional priors such exponential decay of eddy currents, may as well be employed. Moreover, the temporal variability/update and required frequency resolution of the gradient responses are to be further investigated. Here, a combination with field feedback control (11) to address slow non-linear field deviations may be attractive.

Apart from obtaining continuous gradient/trajectory information, the method may also be tested for real-time pre-emphasis tuning or continuous system diagnostics. The accuracy and general applicability of this method offers the possibility to increase system performance and enhance a wide range of MR applications.

Acknowledgements

No acknowledgement found.

References

(1) Vannesjo SJ, Haeberlin M, Kasper L, Pavan M, Wilm BJ, Barmet C, Pruessmann KP. Gradient System Characterization by Impulse Response Measurements with a Dynamic Field Camera. Magn. Reson. Med. Off. J. Soc. Magn. Reson. Med. Soc. Magn. Reson. Med. 2013;69:583–593.

(2) Vannesjo SJ, Graedel NN, Kasper L, Gross S, Busch J, Haeberlin M, Barmet C, Pruessmann KP. Image Reconstruction Using a Gradient Impulse Response Model for Trajectory Prediction. Magn. Reson. Med. 2015,DOI10.1002/mrm.25841.

(3) Campbell-Washburn AE, Xue H, Lederman RJ, Faranesh AZ, Hansen MS. Real-Time Distortion Correction of Spiral and Echo Planar Images Using the Gradient System Impulse Response Function. Magn. Reson. Med. 2015,DOI10.1002/mrm.25788.

(4) Cavusoglu M, Dietrich B, Brunner D, Pruessmann K. Optimization of Parallel RF Transmission Enabled by Concurrent Recording of RF and Gradient Fields. Proc. 23th Annu. Meet. ISMRM Tor. Can. 2015 p.:1819.

(5) Busch J, Vannesjo SJ, Barmet C, Pruessmann KP, Kozerke S. Analysis of Temperature Dependence of Background Phase Errors in Phase-Contrast Cardiovascular Magnetic Resonance. J. Cardiovasc. Magn. Reson. 2014;16,DOI10.1186/s12968-014-0097-6.

(6) Kasper L, Bollmann S, Vannesjo SJ, Gross S, Haeberlin M, Dietrich BE, Pruessmann KP. Monitoring, Analysis, and Correction of Magnetic Field Fluctuations in Echo Planar Imaging Time Series. Magn. Reson. Med. 2014

(7) Wilm BJ, Nagy Z, Barmet C, Vannesjo SJ, Kasper L, Haeberlin M, Gross S, Dietrich BE, Brunner DO, Schmid T, Pruessmann KP. Diffusion MRI with Concurrent Magnetic Field Monitoring. Magn. Reson. Med. 2015

(8) De Zanche N, Barmet C, Nordmeyer-Massner JA, Pruessmann KP. NMR Probes for Measuring Magnetic Fields and Field Dynamics in MR Systems. Magn. Reson. Med. Off. J. Soc. Magn. Reson. Med. Soc. Magn. Reson. Med. 2008;60:176–186.

(9) Barmet C, De Zanche N, Pruessmann KP. Spatiotemporal Magnetic Field Monitoring for MR. Magn. Reson. Med. Off. J. Soc. Magn. Reson. Med. Soc. Magn. Reson. Med. 2008;60:187–197.

(10) Dietrich BE, Brunner DO, Wilm BJ, Barmet C, Gross S, Kasper L, Haeberlin M, Schmid T, Vannesjo SJ, Pruessmann KP. A Field Camera for MR Sequence Monitoring and System Analysis. Magn. Reson. Med. 2015

(11) Duerst Y, Wilm BJ, Dietrich BE, Vannesjo SJ, Barmet C, Schmid T, Brunner DO, Pruessmann KP. Real-Time Feedback for Spatiotemporal Field Stabilization in MR Systems. Magn. Reson. Med. 2015;73:884–893.

Figures

Figure 1: Left: Hardware setup: Field probes mounted to imaging coil. Right: Sequence schematic: Field probe data is acquired repeatedly during the entire MR sequence.

Figure 2: Schematic visualization of the linear system $$$o=Ic$$$ that is used to calculate the unknown gradient response $$$c$$$ (Mx1). The vector $$$o$$$ (Nx1) holds the measured gradient samples. Each row of $$$I$$$ (NxM) holds a timeframe of the input gradients, each row being shifted by the time relating to the corresponding sample in $$$o$$$.

Figure 3: Upper: Excerpt of the given gradient input $$$i_{cont}$$$ (black) for the EPI read gradient axis, the obtained continuous gradient output $$$o_{cont}$$$ (blue) and the measured gradients $$$o$$$ (red). Lower: Zoomed regions of the sequence. $$$o$$$ and $$$o_{cont}$$$ closely match in all regions of the sequence.

Figure 4: Left: Gradient response obtained from the begin ($$$c^{(first)}$$$) and the end ($$$c^{(last)}$$$) of the EPI sequence (read gradient axis). Right: Obtained gradient output from the begin $$$o_{cont}^{(first)}$$$ and the end $$$o_{cont}^{(last)}$$$ of the EPI sequence (top). Zoomed regions show scaling of the gradients (middle) and changed oscillatory terms (bottom).

Figure 5: Gradient echo imaging experiment. Left: Excerpt of the continuously obtained gradient evolution $$$o_{cont}$$$ and measured gradients $$$o$$$ for the read ($$$G_m$$$) and phase encoding ($$$G_p$$$) axis. Middle: Resulting k-space trajectory. Right: Gridding reconstruction of spherical phantom based on k-space trajectory calculated from $$$o_{cont}$$$.



Proc. Intl. Soc. Mag. Reson. Med. 24 (2016)
0544