Andrew C Yung1, Reza Nickmanesh2, Piotr Kozlowski1,3, and David R Wilson2,4
1UBC MRI Research Centre, University of British Columbia, Vancouver, BC, Canada, 2Centre for Hip Health and Mobility, University of British Columbia, Vancouver, BC, Canada, 3Radiology, University of British Columbia, Vancouver, BC, Canada, 4Department of Orthopaedics, University of British Columbia, Vancouver, BC, Canada
Synopsis
With the use of an upright open MR scanner, we demonstrate knee
cartilage T2 mapping using DESS in a true standing position for the first time,
and have shown preliminary evidence that there may be differences between
loading the joint in the standing position versus the supine loaded and unloaded case. The volumetric DESS T2 maps were acquired
with short acquisition time which is critical for imaging weightbearing
postures, while maintaining a range of T2 values that were similar to gold-standard
T2 maps generated by a multi-spin-echo sequence. Introduction
There is
strong interest in measuring cartilage load to assess the causes and treatments
of joint disorders such as osteoarthritis.
One approach is to map T2 in
loaded joints as a surrogate for cartilage deformation due to load
1-3. Open, vertical gap scanners offer
the potential to image weightbearing joints, but rapid T2 mapping sequences are
required because of participant movement.
The effect of loading on T2 in cartilage has not been assessed in
weightbearing joints.
Purpose
To assess
feasibility of T2 mapping using DESS (double-echo steady state) as a surrogate
for deformation of knee cartilage
1 in an open upright MR scanner using the DESS sequence
(double-echo steady state).
Methods
Experiments were performed on a 0.5
Tesla upright MR scanner (MR Open, Paramed, Italy) which allows a full range of
poses from supine to sitting to standing (see Figure 1). A normal female volunteer was scanned with a
dual-channel knee coil in a supine position (unloaded, or loaded on a single
knee with 0.15x body weight using a hanging-weight loading rig), or a standing
position. A 3D DESS T2 mapping sequence with global T1 estimator
4 was implemented to provide full sagittal
coverage of the knee joint (matrix=256x256x24, FOV=22x22x12 cm, TE/TR = 6/16 ms,
NA=1, scan time= 1min38sec, T1 setpoint for fitting procedure = 1.2 sec). A MSE (multi-spin-echo) sequence
5 was used to collect gold-standard
T2 data in a single slice parallel to a DESS slice passing through a tibial
plateau (nonselective composite refocusing pulse, descending/alternating
gradient spoilers, 16 echoes, matrix=256x128, FOV=36x18 cm, slice
thickness=5mm, NA=2, TR=1sec, echo spacing = 15ms, scan time = 4min21sec,
monoexponential fit). All images were postprocessed
with a non-local means filter denoising
6 algorithm.
Results and Discussion:
In regions where cartilage surfaces were in contact,
median T2 in both tibial and femoral cartilage was less for standing than for
both the unloaded and loaded conditions in the supine posture (Figure 2). Figure 3 shows that the standing-posture T2
distributions (femoral/tibial) in contact areas skews lower than the supine
cases; a two-sample Kolmogorov-Smirnov test (threshold p < 0.05) indeed shows
that the standing-posture distributions are significantly different than the
supine-posture distributions. The
decrease in T2 may be due to the consolidation of the cartilage (more immobile
water molecules following loading) and resulting efflux of water from the
tissue.
The differences in T2 distribution between supine loaded and
unloaded were slighter. A 0.15x
body-weight load on one leg without preloading the knee was chosen to minimize
the chance for participant movement, but may not have applied a large enough
load to expect significant changes in T2 values. The supine-posture T2 distributions in the
femoral cartilage contact areas were tested to be significantly different from
each other, which was not the case in the tibial cartilage.
Visual comparison
of the DESS T2 maps and histograms versus the MSE gold-standard T2 data (Figure
4) show similar spatial patterns, with a similar range in T2 values mostly
between 10 and 50 ms. However, the
correspondence between techniques is poorer in certain regions especially for
the standing position in femoral cartilage.
This may be due to increased B0 inhomogeneity at the posterior surface
of the knee, which may increase off-resonance effects in the non-selective
refocusing and thereby increase the MSE signal decay near this region. Another potential source of difference is the
higher partial volume effect in the MSE data, which has a 2.7x larger voxel
size than the DESS data. The DESS/MSE comparison
was performed in a single slice only, since MSE operates in a single slice at a
time (due to nonselective refocusing) and is therefore impractical for whole
volume coverage (4min21sec per MSE slice, whereas DESS provides full coverage
of 24 3D sections in 1min21sec).
Conclusions
To our knowledge, we have demonstrated knee cartilage T2 mapping in a
true standing position for the first time, and have shown preliminary evidence
that there may be differences between loading the joint in the standing
position versus the supine loaded case which is the only option for more
conventional closed-bore higher-field MR scanners. A larger decrease in T2 is expected in true
weightbearing than in supine loading, due to the higher loads
1. The volumetric DESS T2 maps were acquired
with short acquisition time which is critical for imaging weightbearing
postures, while maintaining a range of T2 values that were similar to
gold-standard MSE T2 results. This
initial work paves the way for further cartilage T2 mapping studies in
symptomatic subjects in true physiological load-bearing conditions.
Acknowledgements
The authors thank Erin Macri for the use of the supine loading rig and to Shannon J. Patterson for technologist assistance. This work was supported by an operating grant funded by the Canadian Institutes of Health Research.References
1. Mosher
TJ, Dardzinski BJ. Cartilage MRI T2 Relaxation Time Mapping: Overview and
Applications. Semin Musculoskelet Radiol. 2004 Nov;08(04):355–68.
2. Heule R, Ganter C, Bieri O.
Rapid estimation of cartilage T2 with reduced T1 sensitivity using double echo
steady state imaging. Magn Reson Med. 2014 Mar 1;71(3):1137–43.
3. Poon CS, Henkelman RM.
Practical T2 quantitation for clinical applications. J Magn Reson Imaging JMRI.
1992 Oct;2(5):541–53.
4. Coupe P, Yger P, Prima S,
Hellier P, Kervrann C, Barillot C. An Optimized Blockwise Nonlocal Means
Denoising Filter for 3-D Magnetic Resonance Images. IEEE Trans Med Imaging.
2008 Apr;27(4):425–41.