Tom Hilbert1,2,3, Jennifer Schulz4, Lauren J. Bains4, José P. Marques4, Reto Meuli2, Jean-Philippe Thiran2,3, Gunnar Krueger2,3,5, David G. Norris4, and Tobias Kober1,2,3
1Advanced Clinical Imaging Technology (HC CMEA SUI DI BM PI), Siemens Healthcare AG, Lausanne, Switzerland, 2Department of Radiology, University Hospital (CHUV), Lausanne, Switzerland, 3LTS5, École Polytechnique Fédérale de Lausanne, Lausanne, Switzerland, 4Radboud University Nijmegen, Donders Institute for Brain, Cognition and Behaviour, Nijmegen, Netherlands, 5Siemens Medical Solutions USA, Inc., Boston, MA, United States
Synopsis
Long acquisition times of quantitative magnetic
resonance imaging (qMRI) are one obstacle that prevents qMRI to be used in
clinical routine. Acceleration methods, such as simultaneous-multi-slice and
model-based iterative reconstruction proved in the past to allow high
acceleration factors in MRI. Here we suggest combining these two methods to
allow fast quantitative T2 mapping, yielding a high-resolution (0.7x0.7x3mm³)
whole brain (40 slices) acquisition within a clinically acceptable acquisition
time of less than 3 minutes. T2 values of the proposed method are similar to
the values of the standard method as it is shown on phantom experiments.Intoduction
Despite providing better comparability,
the long acquisition times of classical
quantitative magnetic resonance imaging sequences have impeded their widespread
application in clinical routine and research. Accelerating these acquisitions
is important for their broader acceptance. Here, we investigate the combination
of two techniques to accelerate T2 mapping: Simultaneous Multi Slice (SMS) together
with a model-based undersampling approach. SMS speeds up acquisitions by
simultaneously exciting several slices, subsequently disentangling the
superimposed signals by using parallel imaging techniques
1,5,6. We use
the PINS (“Power independent of number of slices”)
2 SMS pulses due
to their advantageous specific absorption rate (SAR) behaviour. The application
of MARTINI (“Model-based Accelerated RelaxomeTry by Iterative Nonlinear
Inversion”)
3 allows for k-space undersampling, which further reduces
the acquisition time. Using this combination, our technique enables the
sampling of high-resolution whole-brain T2 maps in less than 3 minutes.
Materials & Methods
A multi-echo spin-echo sequence (MESE) was
modified to acquire SMS data. Since PINS pulses show “excitation aliasing”,
i.e. infinitely repeat in slice direction, the excitation of the MESE sequence
was performed using a multiband SMS pulse4. The refocusing pulses
were however replaced by PINS pulses, which reduced the SAR load. Gradient blips
were added in order to shift slices depending on their z-position (“blipped
CAIPI”)5, improving the subsequent parallel imaging reconstruction.
After obtaining written consent from two healthy volunteers, fully
sampled k-space data were acquired using the described prototype sequence with
four simultaneous slices (∆TE/TR/TA 12ms/4s/13:37min, resolution 0.7x0.7x3mm3,
slice gap 0.3mm, 40 slices, 33mm distance between simultaneous slices,
fat-saturation) on a 3T scanner (MAGNETOM Skyra, Siemens Healthcare, Germany)
using a 64-channel head/neck coil. Note that the long TR was chosen to avoid
stimulated-echo effects. The obtained data were first artificially undersampled
(5x acceleration) using the MARTINI block-sampling scheme3. The superimposed
and undersampled k-spaces of the different slices were separated with Split-Slice-GRAPPA4,
yielding undersampled k-space datasets for each single slice. Subsequently,
MARTINI3 was applied on the k-space of each slice, thus obtaining quantitative
M0 and T2 maps. A flowchart of the algorithm is shown in Fig. 1. As
reference, a standard MESE sequence (same parameter as prototype sequence but
only 10 slices, TA=13:37min) was acquired. Using the MESE data, a non-linear
fitting was conducted to obtain T2 values. In an additional experiment, a
multi-purpose phantom (five
compartments with different concentrations of MnCl2∙4H2O, Siemens E-38-19-195-K2130) was scanned with both the
prototype sequence and the reference MESE sequence using the same sequence
parameters as in the in-vivo experiments.
Results & Discussion
The phantom T2 values obtained by PINS-MARTINI
are in good agreement with the reference MESE acquisition (c.f. Table 1),
except for the compartment with very short T2 (~19ms). The observed
overestimation is probably due to a long ∆TE=12ms, rendering the ill-posed fit of such short T2 values even more sensitive
to noise. The sequence protocol could be optimized for shorter T2 values by
shortening the ∆TE; however, we target T2 values of white and grey matter,
which are typically in the range of 70ms to 120ms.
A juxtaposition of T2 maps estimated using the reference method, fully
sampled and 5-fold undersampled PINS-MARTINI is shown in Fig. 2. The T2 map of PINS-MARTINI resembles the reference.
However, an increase in noise is visible which can be explained by the 5-fold
undersampling and noise amplifications caused by the slice-GRAPPA reconstruction.
It should be noted, however, that the increase in noise is acceptable
considering the high acceleration (13:37min for 10 slices versus 2:42min for 40
slices).
Fig. 3 shows four M0 and T2 maps that were simultaneously acquired
using 4x5-fold accelerated PINS-MARTINI, corresponding to an acquisition time
of 2:42min (note that undersampling was still artificial). No inter-slice
leakage or undersampling artifacts are visible within the brain.
Using a PINS instead of multi-band pulse has the advantage of decreasing
the power deposition of the sequence, facilitating the acquisition of whole
brain datasets within the SAR limits. For example, an acquisition with the
proposed sequence parameters reaches ~50% of the SAR limit whereas it may be
exceeded using a standard MESE sequence to acquire the same amount of slices. Furthermore,
PINS pulses do not cause magnetization-transfer, thus provide a better SNR and
the M0 map becomes more proton density weighted.
Conclusion
Using the complementary acceleration techniques
PINS and MARTINI, high-resolution (0.7x0.7x3mm³) whole-brain (40 slices) T2
mapping can be performed in clinically acceptable acquisition times of less
than 3 minutes. In the value range expected in brain tissue, the obtained T2s
are in good agreement with a gold-standard reference fitting.
Acknowledgements
No acknowledgement found.References
1Breuer, Felix A., et al. Controlled aliasing in parallel imaging
results in higher acceleration (CAIPIRINHA) for multi-slice imaging. Magnetic resonance in medicine 53.3 (2005): 684-691.
2Norris, David G., et al. Power independent of number of slices
(PINS) radiofrequency pulses for low-power simultaneous multislice excitation.
Magnetic resonance in medicine 66.5 (2011): 1234-1240.
3Sumpf, Tilman J., et al. Model-based nonlinear inverse
reconstruction for T2 mapping using highly undersampled spin-echo MRI. Journal
of Magnetic Resonance Imaging 34.2 (2011): 420-428.
4Larkman, David J., et al. Use of multicoil arrays for separation of
signal from multiple slices simultaneously excited. Journal of Magnetic
Resonance Imaging 13.2 (2001): 313-317.
5Setsompop, Kawin, et al. Blipped-controlled aliasing in parallel
imaging for simultaneous multislice echo planar imaging with reduced g-factor
penalty. Magnetic Resonance in Medicine 67.5 (2012): 1210-1224.
6Cauley, Stephen F., et al. Interslice leakage artifact reduction
technique for simultaneous multislice acquisitions. Magnetic Resonance in
Medicine 72.1 (2014): 93-102.