How Accurately and Precisely Are we Measuring Coronary Endothelial Function with Radial MRI?
Jerome Yerly1,2, Danilo Gubian3, Jean-Francois Knebel2,4, Thomas Robin5, Giulia Ginami1, and Matthias Stuber1,2

1CardioVascular Magnetic Resonance (CVMR) research center, Department of Radiology, University Hospital (CHUV) and University of Lausanne (UNIL), Lausanne, Switzerland, 2Center for Biomedical Imaging (CIBM), Lausanne, Switzerland, 3University Hospital (CHUV), Lausanne, Switzerland, 4Laboratory for Investigative Neurophysiology (The LINE), Departments of Radiology and Clinical Neurosciences, University Hospital (CHUV) and University of Lausanne (UNIL), Lausanne, Switzerland, 5Transport and Mobility Laboratory (TRANSP-OR), Swiss Federal Institute of Technology of Lausanne (EPFL), Lausanne, Switzerland

Synopsis

MRI with isometric handgrip exercise was recently proposed to non-invasively assess coronary endothelial function. However, the sensitivity of this technique has not yet been fully investigated. To address this need, we have designed a phantom that simulates a physiological range of coronary cross-sectional areas. Radial cine MR images with different spatial resolutions were acquired under moving conditions. Cross-sectional areas were automatically measured and compared to the known nominal values. Statistical analysis suggests that MRI is capable of distinguishing area changes in the order of 0.2-0.3mm2, which correspond to a percentage coronary area change of 3-4% for a 3mm baseline diameter.

Purpose

To discriminate normal from abnormal coronary endothelial function, recent studies have used MRI with isometric handgrip exercise1-4 as the endothelial dependent stressor and reported excellent and reproducible results.4 However, the sensitivity of MRI to measure small changes in cross sectional area of the coronary arteries in response to stress remains to be quantitatively examined. Since the spatial resolution of MRI is limited relative to these expected area changes, it is of utmost importance to address this question. In this study, we have therefore measured the sensitivity of radial MRI for detecting small changes in coronary cross-sectional areas.

Methods

Phantom setup: A phantom was designed to simulate various cross-sectional areas of human coronary arteries5 by drilling holes of different diameters in a block of Polyacetal copolymer (POM-C) (Figure 1a). Twenty-two different diameters, ranging from 3.00mm to 3.42mm, in steps of 0.02mm, were each assigned to 5 random locations on the phantom, so as to avoid introducing potential bias due to magnetic field inhomogeneities (Figure 1b). The phantom was placed in a container filled with tap water and doped with gadolinium (concentration of 5.9mM) to simulate the time-of-flight effect observed in cine imaging. Finally, the phantom was placed on a moving tray to simulate a sinusoidal cardiac motion with a frequency of 40bpm and a maximal displacement of 2cm (Figure 2).

Data acquisition: Data were acquired on a 3T clinical scanner (MAGNETOM Prisma, Siemens Healthcare) using a conventional 2D radial retrospectively ECG-gated cine sequence with an 18-channel chest coil and a 32-channel spine coil. The imaging plane was placed perpendicular to the drilled holes and images were acquired using five different isotropic in-plane resolutions (0.5, 0.6, 0.7, 0.8 and 0.9mm). The acquisition was repeated 10 times for each resolution with the following parameters: FOV=260×260mm2, matrix=288-512, slice thickness=6.5mm, TE/TR=2.5-2.9/4.9-5.1ms, radiofrequency excitation angle=22˚, and temporal resolution=40ms.

Cross-sectional area measurements: Two cine frames with minimum motion were visually selected to measure the cross-sectional areas of the drilled holes with a fully-automated custom-written software package developed in MATLAB. The automatic segmentation followed a similar procedure as described previously6 and uses the full-width half maximum criterion (FWHM) for area measurements. Figure 3 illustrates the various stages of the segmentation.

Statistical analysis: The areas measured for each nominal diameter $$$d$$$ of the drilled holes were grouped together for statistical analysis (Figure 4a) and are denoted by $$$X_d$$$. The normality of the measurements was tested using both the Lilliefors and Jarque–Bera tests. The mean $$$\mu_{X_d}=E(X_d)$$$ and variance $$$\sigma_{X_d}^2=E\big((X_d-\mu_{X_d})^2\big)$$$ for each diameter were computed to derive a probability distribution function that describes the probability of the possible measured areas, $$$X_d\sim N(\mu_{X_d},\sigma_{X_d}^2)$$$. Two normally distributed measures $$$X_i\sim N(\mu_{X_i},\sigma_{X_i}^2)$$$ and $$$X_j\sim N(\mu_{X_j},\sigma_{X_j}^2)$$$ where $$$\mu_{X_i} > \mu_{X_j}$$$ were considered statistically different if the probability of $$$X_i - X_j \sim N(\mu_{X_i}-\mu_{X_j},\sigma_{X_i}^2+\sigma_{X_j}^2)$$$ being positive (i.e., $$$\ge0$$$) is $$$\ge0.95$$$ (Figure 4). Each pair of $$$X_i$$$ and $$$X_j$$$ was compared using this technique. The differences of diameters ($$$i-j$$$) and the results of the tests were stored for each pair. The sensitivity of MRI or smallest detectable change in cross-sectional area was defined as being greater than the highest difference of diameters that has not passed the statistical test. Bland-Altman plots and linear regression analyses were also used for statistical comparisons of the distributions.

Results

A total of 110 distributions of area measurements (22 diameters and 5 resolutions) were analyzed and tested for normality. Each diameter was measured 100 times (5 holes per image x 2 images per acquisition x 10 acquisitions). The Lilliefors and Jarque–Bera normality tests confirmed that 81.8% and 90.9% of the distributions, respectively, could be well-modeled by a normal distribution. The Bland-Altman plots in Figure 5 show a statistically significant bias in the measured areas that is inversely proportional to the image resolution; however the 95% confidence interval characterizing the spread of the data remains approximately constant at 0.4mm2 for all resolutions. The smallest detectable area change ranged between 0.21-0.31mm2 for the different resolutions (Figure 5f) and did not significantly correlate with the image resolution (R=0.13).

Discussion

We presented a moving phantom experiment to quantify the sensitivity of MRI for detecting small changes in coronary cross-sectional areas. Our results suggest that the above MRI approach is capable of distinguishing area changes in the order of 0.2-0.3mm2, which correspond to a percentage area change of 3-4% for a nominal diameter of 3mm. To put this in perspective, and for healthy subjects, the values of coronary endothelial responses reported in the literature using both MRI1-4,7 and invasive techniques8-10 range from 13.2% to 23.1%.

Acknowledgements

This work was supported by the Swiss National Science Foundation grants 320030_143923 and 326030_150828.

References

1. Hays AG, Hirsch GA, Kelle S, Gerstenblith G, Weiss RG, Stuber M. Noninvasive visualization of coronary artery endothelial function in healthy subjects and in patients with coronary artery disease. J Am Coll Cardiol 2010;56(20):1657-1665.

2. Hays AG, Kelle S, Hirsch GA, Soleimanifard S, Yu J, Agarwal HK, Gerstenblith G, Schar M, Stuber M, Weiss RG. Regional coronary endothelial function is closely related to local early coronary atherosclerosis in patients with mild coronary artery disease: pilot study. Circulation Cardiovascular imaging 2012;5(3):341-348.

3. Kelle S, Hays AG, Hirsch GA, Gerstenblith G, Miller JM, Steinberg AM, Schar M, Texter JH, Wellnhofer E, Weiss RG, Stuber M. Coronary artery distensibility assessed by 3.0 Tesla coronary magnetic resonance imaging in subjects with and without coronary artery disease. Am J Cardiol 2011;108(4):491-497.

4. Hays AG, Stuber M, Hirsch GA, Yu J, Schar M, Weiss RG, Gerstenblith G, Kelle S. Non-invasive detection of coronary endothelial response to sequential handgrip exercise in coronary artery disease patients and healthy adults. PloS one 2013;8(3):e58047.

5. Dodge JT, Jr., Brown BG, Bolson EL, Dodge HT. Lumen diameter of normal human coronary arteries. Influence of age, sex, anatomic variation, and left ventricular hypertrophy or dilation. Circulation 1992;86(1):232-246.

6. Yerly J, Ginami G, Nordio G, Coristine AJ, Coppo S, Monney P, Stuber M. Coronary Endothelial Function Assessment Using Self-Gated Cardiac Cine MRI and k-t Sparse SENSE. Magnetic Resonance in Medicine 2015(In press).

7. Terashima M, Meyer CH, Keeffe BG, Putz EJ, de la Pena-Almaguer E, Yang PC, Hu BS, Nishimura DG, McConnell MV. Noninvasive assessment of coronary vasodilation using magnetic resonance angiography. J Am Coll Cardiol 2005;45(1):104-110.

8. Schachinger V, Britten MB, Zeiher AM. Prognostic impact of coronary vasodilator dysfunction on adverse long-term outcome of coronary heart disease. Circulation 2000;101(16):1899-1906.

9. Zhang SH, Tse ZT, Dumoulin CL, Watkins RD, Wang W, Ward J, Schmidt EJ. Gradient-induced voltages on 12-lead ECGs during high-duty-cycle MRI sequences and a theoretically-based method to remove them. Journal of Cardiovascular Magnetic Resonance 2015;17(Suppl 1):P243.

10. Nabel EG, Ganz P, Gordon JB, Alexander RW, Selwyn AP. Dilation of normal and constriction of atherosclerotic coronary arteries caused by the cold pressor test. Circulation 1988;77(1):43-52.

Figures

Illustration of the phantom used in this study to simulate various physiological vasomotor responses of proximal and middle segments of human coronary arteries by drilling holes of different diameters in a block of Polyacetal copolymer (POM-C): (a) picture and (b) drilling layout of the phantom.

Cine images of the moving phantom simulating a cardiac frequency of 40bpm and a maximal displacement of 2cm.

Overview of the different computational steps of the fully-automated custom-written software used to measure the cross-sectional area: (a) a selected image with minimum motion; (b) an automatically detected hole; (c) analyzed radial profiles; (d) full-width half maximum segmentation of the radial profiles; and (e) final segmentation of the cross-sectional area.

Illustration of the proposed statistical test to quantify the cross-sectional area change that is detectable with MRI: (a) distributions of the area measurements corresponding to the various nominal diameters; (b-c) two specific distributions that do not pass the test; and (d-e) two specific distributions that do pass the test.

Linear regression analysis and Bland-Altman plots of the area measurements for every investigated image resolution (a-e). Smallest detectable area changes for the different resolution expressed in mm2, pixel, and percentage with respect to a 3mm nominal diameter (f).



Proc. Intl. Soc. Mag. Reson. Med. 24 (2016)
0454