Improved spiral chemical shift imaging at 3 Tesla using a 32-channel integrated RF-shim coil array
Eren Kizildag1, Jason P Stockmann2, Borjan Gagoski2,3,4, Bastien Guerin2,4, P. Ellen Grant2,3,4, Lawrence L. Wald2,4, and Elfar Adalsteinsson1,5,6

1Department of Electrical Engineering and Computer Science, Massachusetts Institute of Technology, Cambridge, MA, United States, 2A. A. Martinos Center for Biomedical Imaging, Massachusetts General Hospital, Charlestown, MA, United States, 3Boston Children’s Hospital, Boston, MA, United States, 4Harvard Medical School, Boston, MA, United States, 5Harvard-MIT Health Sciences and Technology, Cambridge, MA, United States, 6Institute for Medical Engineering and Science, Cambridge, MA, United States

Synopsis

Severe B0 inhomogeneity manifests itself in the in vivo brain Chemical Shift Imaging (CSI) by broadening the lineshapes and diminishing the quality of the observed spectra. We mitigate this problem by employing a 32-channel integrated RF-shim coil array which uses an optimal combination of local B0 fields from each coil to cancel higher order local field inhomogeneities in the CSI volume. We observed 50% reduction in ΔσB0 over the slab as compared with 2nd order shimming, corresponding to pronounced improvements in the linewidths of 13 out of 24 CSI voxels while modestly worsening in only 3 voxels.

Purpose

Chemical shift imaging (CSI) holds promise for studying brain metabolism and function in healthy and disease states by providing spatial localization of metabolites. However, widespread clinical use is impeded by local B0 field variations arising from tissue susceptibility differences (esp. near sinus cavities) that are inadequately compensated by standard 2nd-order spherical harmonic (SH) shim coils, resulting in diminished spectral quality and poor water and lipid suppression [1]. Multi-coil (MC) local shim arrays [2] provide an efficient alternative to higher-order SH shim coils for nulling high-order local B0 variations without suffering from the low efficiency, high eddy currents and expensive current drivers associated with SH coils; however, the MC shim arrays compete with RF receive arrays for space near the head. The recent introduction of integrated RF-shim coils [3,4] provides the advantages of MC shim arrays and RF receive arrays without compromising the performance of either system. They have previously been used to reduce geometric distortion in EPI images (Fig. 1) [3] and here we demonstrate their use for improving spectral quality in CSI.

Methods

Experiments were performed using an anthropomorphic head phantom on a Siemens Skyra 3 Tesla scanner (Siemens Healthcare, Erlangen, Germany) to assess improvements in CSI data provided by a previously-described 32ch RF-shim array (Fig. 1) [3] over standard 2nd-order SH shims. Phantom: A realistic head model [5] was 3-D printed in ABS plastic shells forming brain and muscle compartments (Fig. 2). The phantom includes a sinus air cavity that mimics ΔB0 patterns in the frontal lobes in vivo. The brain compartment was filled with “Braino” metabolite solution [6] containing GABA, glutamate, choline, creatine, and NAA at 5X typical in vivo concentrations to reduce the need for signal averaging. Thus the phantom reflects realistic field maps (B1+ and B0) and spectral information encountered in vivo.

Acquisition: Constant density spiral-based k-space trajectories were appended to conventional PRESS-box excitation for single-slice, time-efficient spectral-spatial encoding [7]. Three TRs encoded a Cartesian matrix of (x,y,f) = (12,12,460) points (zero-padded to 16x16x512) over a 24cm FOV and spectral bandwidth of 1450Hz, for an overall voxel size of 16cc (2x2cm in-plane with 4cm slice). The PRESS-box excited VOI=[80x140x40]mm is inscribed wholly within the brain compartment. The spiral k-space data are 2X-gridded using a Kaiser-Bessel kernel. Ten averages yield NAA SNR ~ 25 for a total imaging time of 90s (TE=30ms and TR=2s). Within the shim adjust volume covering the PRESS-box excitation VOI, there were 24 target CSI voxels. Shimming: Scanner 2nd-order shims were applied over the PRESS-box VOI and spiral CSI data were acquired, followed by gradient-echo B0 field maps (2mm slice, 2.4mm in-plane). A previously mapped basis set of unit-current B0 field maps for the shim coils were used to minimize a least squares objective on residual B0 field subject to maximum current per loop (2.5A) and total current in the array (35A) constraints. After the application of optimal MC shim currents, the field mapping and CSI acquisitions were repeated. The high-resolution field maps were used to quantify shim performance using the standard deviation of ΔB0 over the whole excited volume (σB0GLOBAL) as well as over each individual CSI voxel (σB0LOCAL) [1].

Results

Multi-coil shims provide a 50% reduction in σB0GLOBAL (Fig. 3) and similar improvements in σB0LOCAL within the majority of CSI voxels (Fig. 4). Good agreement was obtained between calculated and acquired B0 field maps with the MC shims applied (using 9A of total current for the prescribed shim). As compared with 2nd-order shim field maps, the MC shims reduce peak ΔB0 and produce a more even distribution of ΔB0 over the whole CSI volume. Spectral linewidths were improved in 13 out of the 24 encoded voxels, remained comparable in 8 voxels, and modestly worsened in 3 voxels (Fig. 4).

Discussion

DISCUSSION:


MC shimming substantially improved spectra in the posterior area of the PRESS-box where the 2nd-order shim had compromised ΔB0 while attempting to shim the anterior frontal lobe B0 hotspot. Moreover, the large reduction in σB0GLOBAL enabled by MC shims is expected to benefit the performance of frequency selective RF pulses used for VOI excitation, water & lipid suppression, and spectral editing [1]. However, mismatch exists in parts of the frontal lobe (e.g. voxel A) between σB0LOCAL and the linewidth, possibly due to shifts in the acquired voxel location caused by ΔB0. Since B0 inhomogeneity in the “no-care” region is much worse in the MC shim case, the proposed methods would benefit from swapping the PRESS-box RF pulses to an excitation module that has sharper spatial localization and less sensitivity to chemical shift (e.g. LASER [8]).

Acknowledgements

The authors thank Trina Kok for help preparing brain metabolite solution and Jon Polimeni for sharing his image acquisition and analysis scripts. Support by NIH R21 EB017338, P41 EB015896, BRP NIH R01EB017337.

References

[1] Pan JW, MRM 2012. [2] Juchem C, JMR 2011. [3] Stockmann JP, MRM 2015. [4] Truong TK, NeuroImage 2014. [5] Guérin B, MRM 2015. [6] Lattanzi R, NMR Biomed. 2010. [7] Adalsteinsson E, MRM 1998. [8] Slotboom J, J Magn Reson 1991.

Figures

Figure 1. (left) A 32-channel integrated RF-shim coil uses the same close-fitting array of loops for RF signal detection and high-order B0 shimming, enabling reduced geometric distortion in high resolution echo planar imaging scans (1mm in-plane, 1.11 ms echo spacing, GRAPPA R=1).

Figure 2. A 3D-printed anthropomorphic head phantom (a.) is used for evaluating the MC shim performance with the spiral CSI acquisition. The brain compartment (b.) is filled with a “Braino” solution containing brain metabolites to provide realistic spectra. The phantom includes a sinus cavity (a.) that creates pronounced B0 deviations in the frontal lobe region (c.) as shown in the axial ΔB0 field map at the level of the red line.

Figure 3. B0 field maps for four representative slices within the 4cm-thick CSI slab demonstrate improved field homogeneity with MC shims as compared with the standard 2nd order SH shims within the excited PRESS-box (black rectangle). The MC shim reduces σB0GLOBAL over the entire slab by 50%, and also reduces σB0 within the ROI in field map slice as indicated.

Figure 4: Twelve representative voxels (each 16cc) are shown comparing performance of 2nd order and MC shimming. Spectra showing clear improvement with MC shims are highlighted. Reductions in σB0LOCAL within each voxel correspond to improved B0 homogeneity in the field maps shown in Fig. 3 and also to reduced linewidths. However, in voxel A there is a disparity between linewidth and σB0LOCAL that could arise due to B0-induced distortions in the shape of the PRESS-box excitation volume.



Proc. Intl. Soc. Mag. Reson. Med. 24 (2016)
0380