Michael Wyss1, Andrei Manoliu2, Georg Spinner1, Magda Marcon2, Roger Luechinger1, Daniel Nanz2, Klaas P. Pruessmann1, and Gustav Andreisek2
1Institute for Biomedical Engineering, University of Zurich and ETH Zurich, Zurich, Switzerland, 2Institute of Diagnostic and Interventional Radiology, University Hospital Zurich and University of Zurich, Zurich, Switzerland
Synopsis
Reliable fat suppression is
challenging but mandatory for clinical 7.0T imaging. Purpose of this study was to
evaluate different fat suppression techniques for clinical 7.0T knee MRI. Eight
volunteers were imaged at 7.0T (Achieva, Philips) using a dedicated 28-channel
TX-knee coil (QED) and axial PDw-TSE sequences without fat suppression, with
SPIR, with SPAIR, with SSGR and with the combination of SSGR+SPIR. Purpose
Recently, dedicated knee coils have become available for clinical
7.0T imaging. For the latter, reliable fat suppression is mandatory which is,
however, challenging due to increased magnetic field inhomogeneity,
inhomogeneous RF transmit field and higher specific absorption rate (SAR). Standard
techniques such as spectral pre-saturation with inversion recovery (SPIR) or
spectral adiabatic inversion recovery (SPAIR) increase SAR and lengthen the acquisition
time. An alternative is slice selective gradient reversal (SSGR) which uses gradients of opposing polarity for the
excitation and refocusing pulse (1). SSGR is expected to perform well at 7.0T
without introducing additional scan time or increasing SAR (2,3). Purpose
of this study was to evaluate different fat suppression techniques for clinical
7.0T knee MRI.
Methods
Local ethics board approved study
with written informed consent from eight healthy volunteers (mean age 31±4
years, six males). Volunteers were imaged at 7.0T (Achieva, Philips Healthcare,
Cleveland, OH) using a dedicated 28-channel TX-knee coil (QED, Quality
Electrodynamics, Mayfield Village, OH) and at 3.0T (Skyra, Siemens Healthcare,
Erlangen, Germany) using a 15-channel TX-knee coil (QED). At 3.0T, an axial
proton-density-weighted turbo spin echo (PDw-TSE) sequence with SPAIR from the
hospital`s clinical standard protocol, and at 7.0T, a series of five PDw-TSE
sequences were acquired: a) without fat suppression, b) with SPIR, c) with
SPAIR, d) with SSGR and e) with the combination of SSGR+SPIR. Except for the fat
suppression technique, all other imaging parameters were kept identical for all
acquisitions (field-of-view: 160x160mm
2, TR: 3800ms, TE: 35ms, voxel
size: 0.35x0.45x2.50 mm
3, 15 slices, 1 signal average, bandwidth:
216 Hz/pixel, acquisition time: 3:40 min). Each 7.0T sequence contained an
additional noise scan without gradients and RF to allow pixel-wise calculation of
signal-to-noise ratio (SNR) maps (4). SNR
values were extracted from regions-of-interest (ROIs) within the cartilage,
bone marrow, joint fluid, muscle and fat. Contrast-to-noise ratio (CNR) values
were calculated between these tissue types. All MR measurements were repeated
on a dedicated water-fat phantom (ultrasonic gel and swine fat). Additionally,
the image quality and fat suppression was analyzed by two independent radiologists
using 5-point Likert scales. Kappa statistics were calculated to evaluate
inter-reader agreement. Quantitative and qualitative results from the different
sequences were compared using paired sample t-tests and Wilcoxon signed rank
tests. A p-value of <0.05 was considered statistically significant.
Results
Quantitative analysis: In the phantom
experiment, fat was only partially suppressed using SPIR and SPAIR. SSGR and SSGR+SPIR
suppressed the fat signal much more compared to the original fat signal without
any suppression (below 15% of the original signal) (Fig. 1). Figure 2 shows
examples of acquired images in vivo at both field strengths and corresponding
SNR maps. At 7.0T, SSGR and SSGR+SPIR yielded significantly lower SNR in fat
compared to SPIR and SPAIR (p<0.001, Fig. 3). SSGR and SSGR+SPIR yielded
significantly lower SNR in cartilage compared to SPIR (p≤0.002) but similar SNR
compared to SPAIR (SSGR: p=0.105, SSGR+SPIR: p=0.132). Relative SNRs of fat
demonstrated that the SPIR technique reduced the fat signal to 48±6%, SPAIR:
22±1.6%, SSGR: 12±0.9% and SSGR+SPIR: 10±0.25% (Fig. 1 and 3). Evaluation of CNR
showed superior contrast between muscle-fat, cartilage-fat, fluid-fat, and
fluid-cartilage for the SSGR method (Fig. 4).
Qualitative analysis: In-vivo images showed improved image quality for all 7.0T methods
compared to the clinical reference 3.0T SPAIR image (all p≤0.001) (Tbl. 1c,d).
Compared to 3.0T, the grade of fat suppression was rated lower for SPIR
(mean±SD, 2.0±0.00) and SPAIR (3.0±0.00) but higher for SSGR (4.0±0.25) and
SSGR+SPIR (5.0±0.00). The homogeneity of the fat suppression was rated lower
for 7.0T SPIR (3.69±0.48) compared to the clinical reference 3.0T SPAIR image
(3.94±0.44). The homogeneity was better than the clinical reference with 7.0T
SPAIR (4.0±0.0), SSGR (4.19±0.40) and 7.0T SSGR+SPIR (4.31±0.48). However, only
for SSGR+SPIR the homogeneity improvement over the clinical reference image was
finally statistically significant (Tbl. 1c,d).
Discussion
The SSGR method provided a strong, stable and homogenous fat suppression
for clinical knee imaging at 7.0T.
Quantitatively no significant differences in the grade of fat
suppression was found between SSGR and SSGR+SPIR, but SSGR+SPIR yielded qualitatively
better images in terms of grade of fat suppression and homogeneity of fat
suppression than SSGR as assessed by two independent radiologist. Compared to
standard spectrally-selective suppression methods at 3.0T, the SSGR technique can
provide stand-alone or in combination with SPIR fat suppression in better
quality, a shorter scan time and without a SAR elevation.
Conclusion
At 7.0T, fat saturation for clinical knee imaging using SSGR and the
combination of SSGR and SPIR was superior compared to methods based on
spectrally selective RF pulses.
Acknowledgements
The authors gratefully acknowledge Zoltan Nagy for his advises regarding the SSGR technique.References
1. Park HW, Kim DJ, Cho ZH. Gradient reversal
technique and its applications to chemical-shift-related NMR imaging. Magnetic
Resonance in Medicine. 1987;4(6):526–536.
2. Nagy Z, Weiskopf N.
Efficient fat suppression by slice-selection gradient reversal in
twice-refocused diffusion encoding. Magnetic Resonance in Medicine.
2008;60(5):1256–1260.
3. T.Takahara, J.
Zwanenburg, F. Fisser, T. Ogino, D. Klomp, H. Hoogduin, J. Hendrikse, T. Kwee,
W. Mali, P. Luijten. Fat suppression with Slice-Selection Gradient Reversal
(SSGR) revisited. In: Proc. Intl. Soc. Mag. Reson. Med.
17 (2009). Vol. 2642. Honolulu, Hawaii, USA; 2009.
4.
Nordmeyer-Massner JA, De Zanche N, Pruessmann KP. Mechanically adjustable coil
array for wrist MRI. Magnetic Resonance in Medicine. 2009;61(2):429–438.