Neerav Dixit1, Pascal Stang2, John Pauly1, and Greig Scott1
1Electrical Engineering, Stanford University, Stanford, CA, United States, 2Procyon Engineering, San Jose, CA, United States
Synopsis
Thermo-acoustic ultrasound uses pressure waves generated by
thermoelastic expansion to measure heating. This technique can be used to
detect excessive local SAR and RF tip heating in implanted or interventional devices.
We compare the signal quality and inherent properties of several modulation
schemes for thermo-acoustic ultrasound. We then interface a system for
thermo-acoustic detection of heating with an MRI scanner and demonstrate the
first use of thermo-acoustic ultrasound to detect RF tip heating from an MRI scanner.Purpose & Introduction
Patients with implanted devices, including deep brain
stimulators and pacemakers, often have restricted access to MRI [1,2] due to RF
tip heating concerns. RF tip current must also be limited when visualizing
electrophysiology (EP) catheters or dipole micro-transceivers in interventional
MRI [3]. Thermo-acoustic ultrasound, in which thermoelastic expansion caused by
heating creates measureable pressure waves, is a promising method for detection
of local heating. Here, we compare sinusoidal and square wave transmit
modulation schemes to pulse trains for thermo-acoustic ultrasound. We then
demonstrate the first thermo-acoustic ultrasound detection of tip heating in an
MRI scanner.
Theory & Methods
For initial tests comparing alternative transmit methods, we
generated thermo-acoustic signals by driving current directly into a coaxial
cable immersed in saline solution with its center conductor exposed at the tip
(Figure 1e). All methods used the same average transmit power level and a total
transmit duration of 4ms.
Continuous modulation of RF at ultrasound transducer
frequencies generates continuous wave (CW) thermo-acoustic ultrasound [4,5]. Linear
frequency modulated CW (FMCW) results in frequency-encoding of depth. To create
a thermo-acoustic signal, a 64MHz carrier was amplitude-modulated by a 2ms RF
chirp with a 375kHz bandwidth centered at 1MHz using the FMCW technique (Figure
1a).
The RF power of a sinusoidal amplitude-modulated signal has
an envelope similar to a square wave. Because thermo-acoustic pressure depends
on transmit power, we tested the ability of square wave modulation to create
thermo-acoustic signals. A 64MHz carrier was square-wave modulated at a 50%
duty cycle, but with pulse repetition frequency modulated using FMCW with a
375kHz bandwidth and 1MHz center frequency (Figure 1b). A similar method, with the square wave on-time
fixed at 500ns but with FMCW-modulated pulse repetition frequency, was tested
as well (Figure 1c).
Classic ultrasound pulse trains were also tested. A pulse
length of 500ns was chosen so that the power profile of the transmitted signal
had significant spectral content at 1MHz - the ultrasound transducer frequency.
Repetitions of 64MHz carrier pulses were used (Figure 1d).
To detect the thermo-acoustic signal from tip heating
induced in MRI, we interfaced our thermo-acoustic detection setup with the body
coil transmitter of a GE Signa Excite 1.5T scanner (Figure 3). For 250 averages,
a Medusa console [6] configured in slave mode transmitted
a 3ms FMCW signal centered at 1MHz with a 375kHz bandwidth when triggered by
the body coil transmit gate. The signal amplitude
modulated a 63.855MHz carrier frequency, which was applied to the body coil
amplifier through an RF SPDT switch. The transmit signal waveform (Figure 3c) was verified
using a pickup loop inside the magnet. A long wire was placed in
the scanner bore, and one end of the wire with an exposed tip was immersed in
saline solution. An Olympus V303 immersion transducer was placed in the solution
and attached to an RF-shielded box containing a preamplifier and LED driver
powered by a non-magnetic battery. The LED optically coupled the amplified
transducer signal to a photo-receiver outside the MRI scan room to minimize RF
interference from the body coil. This signal
was then filtered and sent to the Medusa RF receiver.
Results
Figure 2 shows the depth profiles of the thermo-acoustic
signals from the different transmit and receive schemes when directly driving
the stripped coaxial cable as a signal source. We also found that signal
sensitivity dropped by ~4x when the driven coax tests were repeated inside the magnet
bore. Figure 4 shows the thermo-acoustic signal resulting from coupling between
the body coil’s electric field and a long wire in the bore. System
nonlinearities and RF interference created an observable feedthrough signal at
0cm. In both figures, the signal source was present at a depth of ~6cm.
Discussion & Conclusions
For the chosen transmit parameters and using a 1MHz transducer,
the SNR of the different transmit and receive methods are comparable. For the modulated
CW techniques, resolution of the reconstructed signal was ~5mm for the 375kHz
bandwidth transmit signal. For the pulse train transmit, the resolution was
limited to ~3mm by the 500kHz bandwidth of the Medusa receiver. The required
scan time was significantly longer for the pulse train because unlike the
modulated CW methods, transmission and reception could not occur simultaneously.
Our first demonstration of thermo-acoustic detection of RF
tip heating from an MRI body coil achieved > 100 SNR. Conceptually, a
thermo-acoustic modulation system could augment any MRI scanner through a SPDT
RF routing switch to the body coil transmitter. This demonstration bodes well
for the prospects of using thermo-acoustic ultrasound to assess RF heating
risks of implanted and interventional devices.
Acknowledgements
NIH Grant Support: R01EB008108, P01CA159992.
DARPA MEDS program
References
1. Rezai AR, Phillips M, Baker KB, et al. Neurostimulation
System Used for Deep Brain Stimulation (DBS): MR Safety Issues and Implications
of Failing to Follow Safety Recommendations. Invest Radiol. 2004; 39:300-303.
2. Nyenhius JA, Park S, Kamondetdacha R, et al. MRI and
Implanted Medical Devices: Basic Interactions With an Emphasis on Heating. IEEE
Trans Dev Mat Rel. 2005; 5:467-480.
3. Etezadi-Amoli M,
Stang P, Kerr A, et al. Interventional device visualization with toroidal
transceiver and optically coupled current sensor for radiofrequency safety
monitoring. MRM. 2015; 73:1315-1327.
4. Scott G, Etezadi-Amoli M, Stang P, et al. Thermo-Acoustic
Ultrasound Detection of RF Coil and Tip SAR. Proc. Int. Soc. Magn. Res. Med. 2015;
23:0377
5. Nan H, Arbabian A. Stepped-frequency continuous-wave
microwave-induced thermoacoustic imaging. Appl Phys Lett. 2014; 104:224104
6. Stang PP, Conolly SM, Santos JM, et al. Medusa: A
Scalable MR Console Using USB. IEEE TMI.
2012; 31:370-739