Alex S Hong1, Emilie Bollache1, Pim van Ooij1, James C Carr1, Alex J Barker1, Jeremy D Collins1, and Michael Markl2
1Department of Radiology, Northwestern University, Chicago, IL, United States, 2Department of Radiology, Department of Biomedical Engineering, Northwestern University, Chicago, IL, United States
Synopsis
Aortic valve replacement (AVR) is an effective surgical approach to
treating aortic valvular disease, but it is unclear if and what type of
prosthesis can fully reproduce physiologically normal flow characteristic of a
native aortic valve. We utilized 4D
flow MRI to systematically compare blood flow in the thoracic aorta in post-AVR
(bioprosthetic vs. mechanical) patients and healthy controls. Both
bioprosthetic and mechanical valves were found to produce higher peak systolic flow
velocities and peak wall shear stress in the ascending aorta than native valves,
demonstrating the presence of significant changes in aortic blood flow in AVR
patients.Purpose
Recent studies have demonstrated the potential
diagnostic value of 4D flow MRI for evaluating altered aortic hemodynamics in
patients with aortopathy and after aortic valve replacement (AVR).
1,2
In this study, 4D flow MRI was used to characterize aortic blood flow in
patients who have undergone AVR with
bioprosthetic vs. mechanical valves.
Methods
ECG and respiratory gated 4D flow MRI with full
volumetric coverage of the thoracic aorta was acquired in 19 patients who
received AVR at Northwestern Memorial Hospital (Chicago, IL, USA) between 2012
and 2015 (n = 10 bioprosthetic valves, n = 9 mechanical valves), as well as 6
healthy controls matched for age and blood pressure. All acquisitions were made
on 1.5T MRI systems (Siemens, Erlangen, Germany) during free breathing with
average spatial resolution of approx. 2.5 mm3 and temporal
resolution of 38-42 ms. 4D flow data analysis included corrections for Maxwell
terms, eddy currents and velocity aliasing (Matlab, MathWorks Inc., MA, USA) and
3D segmentation of the aortic volume (Mimics, Materialize NV, Leuven, Belgium).
Blood flow in the thoracic aorta was visualized in
EnSight (CEI, Apex, NC, USA) by way of 3D pathlines showing the time-resolved
evolution of velocities throughout one cardiac cycle. Nine analysis planes were
placed along the aorta, perpendicular to the aortic axis, at the following anatomical
landmarks as shown in Figure 1: 1) at the aortic root 1-2 cm above the valve, 2)
at the proximal ascending aorta (AAo) 1-2 cm above the sinotubular junction, 4)
at the distal AAo just before the brachiocephalic trunk, 3) at the mid-AAo
halfway between planes 2 and 4, 5) at the mid-arch just before the common
carotid artery, 6) at the distal arch just before the left subclavian artery, 7)
at the proximal descending aorta (DAo), 8) at the DAo at the level of plane 2,
and 9) at the distal DAo. In all analysis planes, quantification of peak
systolic velocities was performed.
In addition, peak systolic 3D wall shear stress (WSS)
was calculated along the entire segmented 3D aorta wall based on a previously
reported method.3,4 As shown in Figure 2, the aorta was partitioned
into 10 distinct regions of interest (ROI), corresponding to the inner (along
the lesser curvature) and outer (along the greater curvature) segments of the
proximal AAo, distal AAo, arch, proximal DAo, and distal DAo, respectively. Mean
peak systolic WSS was computed inside each ROI.
Inter-group differences in peak systolic velocity and
peak WSS were studied using Welch’s t-test with a p-value less than 0.05
denoting statistical significance.
Results
Bioprosthetic valves produced significantly higher
peak systolic velocities than controls in the root (2.24 vs. 1.25 m/s, p <
0.01) and all three AAo planes (1.93 vs. 1.21 m/s, p < 0.01; 1.58 vs. 1.02
m/s, p < 0.01; 1.23 vs. 0.89 m/s, p = 0.01), while AVR with mechanical
valves resulted in higher velocities than controls in only the root (2.09 vs.
1.25 m/s, p < 0.01) and proximal AAo (1.58 vs. 1.21 m/s, p = 0.03). When
comparing the two AVR groups, bioprosthetic valves produced greater velocities
than mechanical valves in the proximal and mid-AAo. Both bioprosthetic and
mechanical valves had greater peak WSS than controls in the inner proximal AAo (2.70
and 2.22 vs. 1.40 N/m2, both p < 0.01) and outer proximal AAo
(2.52 and 2.02 vs. 1.31 N/m2, both p < 0.01). There were no
significant differences in peak WSS in the AAo between the two AVR groups.
Additionally, mechanical valves had greater WSS than bioprosthetic valves in
the inner proximal DAo, as well as in the inner and outer distal DAo. Tables 1 and 2 provide a full tabulation of results.
Conclusion
The results of our study show that there are
significant changes to aortic blood flow in AVR patients compared to healthy
controls, which include elevated peak systolic velocities and WSS in the
ascending aorta. Although increased WSS in the AAo has been implicated in the
development of aortic pathology
5, the long-term clinical outcomes
of AVR patients with elevated systolic velocities and WSS on 4D flow MRI are
not well-understood and present opportunities for future research. Subsequent
analyses will investigate the independent influence of aortic stiffness related
to variable extents of graft replacement of the aorta on local blood flow. Accurate
characterization of post-operative blood flow by 4D flow MRI might
differentiate the efficacies of various surgical approaches to AVR in
reproducing physiologic flow patterns and reducing further aortopathy that may
result from derangements to aortic hemodynamics over time.
Acknowledgements
Grant support by the National Institutes of Health [grant numbers R01HL115828 and K25HL119608].References
1. von
Knobelsdorff-Brenkenhoff F, Trauzeddel RF, Barker AJ, Gruettner H, Markl M,
Schulz-Menger J. Blood flow characteristics in the ascending aorta after aortic
valve replacement-a pilot study using 4d-flow mri. Int J Cardiol.
2014;170:426-433
2. Semaan E, Markl
M, Chris Malaisrie S, Barker A, Allen B, McCarthy P, Carr JC, Collins JD.
Haemodynamic outcome at four-dimensional flow magnetic resonance imaging
following valve-sparing aortic root replacement with tricuspid and bicuspid
valve morphology. Eur J Cardiothorac Surg. 2014;45:818-825
3. van Ooij P,
Potters WV, Collins J, Carr M, Carr J, Malaisrie SC, Fedak PW, McCarthy PM,
Markl M, Barker AJ. Characterization of abnormal wall shear stress using 4d
flow mri in human bicuspid aortopathy. Ann Biomed Eng. 2015;43:1385-1397
4. Potters WV, van Ooij
P, Marquering H, Vanbavel E, Nederveen AJ. Volumetric arterial wall shear
stress calculation based on cine phase contrast MRI. J Magn Reson Imaging. 2015;41(2):505-16.
5. Barker AJ, Markl M,
Burk J, Lorenz R, Bock J, Bauer S, Shulz-Menger J, von Knobelsdorff-Brenkenhoff
F. Bicuspid aortic valve is associated with altered wall shear stress in the
ascending aorta. Circ Cardiovasc Imaging.
2012;5(4):457-66.