Synopsis
Compared to standard two-dimensional (2D) arterial spin labeling (ASL), simultaneous multi-slice (SMS) ASL imaging techniques can reduce T1 relaxation effect of the label; improve spatial coverage and resolution. However, existing 2D SMS ASL techniques are sub-optimal for the background suppression (BS) technique since multiple SMS excitations are required. In this study, we propose a novel constrained slice-dependent BS scheme for 2D multi-slice pseudo-continuous ASL (pCASL) with SMS-EPI acquisition, to suppress background signal across a wide range of T1s. In vivo experiment showed that the BS scheme can increase temporal SNR of perfusion images 1.5-2 folds.Purpose
Two-dimensional simultaneous multi-slice (SMS)
imaging techniques have recently been applied for arterial spin labeling (ASL)
perfusion imaging. Compared to standard 2D ASL, SMS ASL can reduce T1
relaxation effect of the label through shortened acquisition time; improve
spatial coverage and resolution with little penalty in SNR
1-2. However,
existing 2D SMS ASL techniques are sub-optimal for the background suppression
(BS) technique since multiple SMS excitations are required. As a result, the
temporal SNR (tSNR) of SMS-EPI ASL is inferior to that of 3D BS GRASE ASL
1. In
this study, we propose a novel constrained slice-dependent BS scheme for 2D
multi-slice pseudo-continuous ASL (pCASL) with SMS-EPI acquisition, to suppress
background signal across a wide range of T1s.
Theory
Background suppression is generally achieved by
suppressing brain tissue signals using several inversion pulses following a pre-saturation
of the whole imaging volume
3, while leaving the target spins in blood
undisturbed. For 2D multi-slice imaging, however, the background tissue signal
recovers between multi-slice acquisitions, thereby compromising the efficiency
of BS. As shown in Fig. 1, a constrained slice-dependent BS scheme is proposed
that employs a group of slice selective pre-saturation pulses and two global
inversion pulses, the timings of which are determined based on the pCASL labeling
duration and post-labeling delay (PLD) so that the
background tissue signal of each imaging slice can be precisely
nulled/suppressed when it is excited. As shown in Fig. 2 of simulation results,
the signal of static brain tissue of 2 imaging slices can be suppressed at the
time of excitation, while the blood signal can be largely preserved under the
proposed BS scheme. This BS scheme can be expanded for 2D SMS pCASL scans using
multiband pre-saturation and excitation RF pulses. Optimization of BS timing
parameters including flip angles (
FA0) of pre-saturation pulses, saturation times (
τsat) and inversion times (
τinv) (Fig. 1) can be treated as a nonlinear optimization
problem to minimize total residual signal, expressed as the squared sum of background
signals across a range of T1 values for all imaging slices.
Methods
The proposed BS scheme was implemented for an
SMS EPI pCASL sequence on a Siemens 3T Prisma scanner with a 20-channel head
coil. The imaging parameters were: FOV=220 mm, matrix=64×64, slice thickness=5
mm, TE=14 ms, TR=3540 ms, flip angle=90°, label/control
duration=1500 ms, PLD=1000 ms, slice acquisition time=45 ms and 30 acquisitions.
6/12/18 slices were acquired using 2D EPI with single-band (SB) and multiband (MB)-2/3
slice accelerations, respectively, and aliased slices were separated with
slice-GRAPPA algorithm
4. Two hyperbolic secant pulses were implemented as
non-selective inversion pulses and pre-saturation was achieved by slice
selective RF pulses followed by gradient spoilers. Scans were performed on a phantom
and human brain. The phantom was designed as a four-layer (slice) rack with
four test tubes in each layer containing gel of different T1 values (1148,
1586, 1908 and 2197 ms from left to right in Fig. 3). Optimization of the BS
scheme was performed to suppress all 4 T1 values of the phantom, as well as T1
values of
gray and white matter
(600 to 1400 ms) for human studies. To evaluate the performance of the proposed BS
scheme, voxel-wise tSNR was calculated as the mean value divided by the
standard deviation of subtracted perfusion images.
Results and Discussions
Figure
3 shows 4 slices of phantom acquired with and without BS. The signal of the 4
test tubes can be on average suppressed to 10.68% of its original intensity
across 4 slices. For human experiment, an optimized BS scheme was designed for
six slices considering the timing constraints. The optimized BS timing
parameters were: [
FA0,1 ...
FA0,6 ] = [97, 119, 123, 161, 180, 180]°, [
τsat,1 ...
τsat,6 ] = [108, 293, 6, 6, 0, 0] ms and [
τinv,1 τinv,2 ] = [84, 795] ms.
Average background signal was 4.8±4.4% after suppression based on simulation. Figure
4 shows both background images (a, c, e, g) and corresponding perfusion maps (b,
d, f, h) from non-BS SB, BS SB, BS MB2 and BS MB3 sequences of the same scale. As
shown in Table 1, an 89% of background signal reduction was achieved with the proposed
BS scheme and the mean tSNR across the whole brain increased from 0.43 to
0.95/0.80/0.64 in BS SB/MB-2/3 data respectively, which indicates a more stable perfusion
signal in BS acquisitions.
Conclusions
The
proposed constrained
BS scheme for 2D multi-slice pCASL in conjunction with SMS acquisition offers a
promising approach for effective suppression of background signals across a
wide rang of T1s while ensuring sufficient brain volume coverage.
Acknowledgements
This research was supported by the National Institute of Health 553
(NIH) Grants R01-NS081077 and R01-EB014922.References
[1]
D. Feinberg, et al., MRM, 2013, p. 1500-1506. [2] T. Kim, et al., MRM, 2013, p.
1653-1661. [3] Mani S, et
al. MRM, 1997, 37(6): 898-905. [4] K. Setsompop, et al., MRM, 2012, p.
1210-1224.