Diffusion-weighted MRI using undersampled radial STEAM with iterative image reconstruction
Andreas Merrem1, Jakob Klosowski1, Sabine Hofer1, Klaus-Dietmar Merboldt1, and Jens Frahm1

1Biomedizinische NMR Forschungs GmbH, Max-Planck-Institut für Biophysikalische Chemie, Göttingen, Germany

Synopsis

Single-shot STEAM MRI is a method for black-blood diffusion-weighted imaging where the use of radiofrequency-refocussed echoes leads to no image distortions, no susceptibility artifacts, and no violations of the Carr-Purcell-Meiboom-Gill condition. Despite these favorable properties, clinical applications have been limited by a low signal-to-noise ratio. Here, we demonstrate the development of highly undersampled radial diffusion-weighted single-shot STEAM MRI with iterative reconstruction to achieve acceptable signal-to-noise for studies of the human brain.

Introduction

Susceptibility artifacts and image distortions are a major problem in diffusion-weighted EPI for various clinical applications. The problem may be overcome by the use of single-shot STEAM (ssSTEAM) MRI [1] which allows for anatomically correct imaging without sensitivity to magnetic field inhomogeneities due to the acquisition of radiofrequency-refocused echoes. Because multiple 180˚-pulses as in spin-echo MRI are not required, exceedingly high energy absorption rates as well as artifacts due to violations of the Carr-Purcell-Meiboom-Gill condition are also avoided. Additionally, the black-blood property of the sequence [2] suppresses unwanted signal contributions to the diffusion-weighted signal. So far, a poor signal-to-noise ratio (SNR) has been the main limitation for a more extended clinical use of diffusion-weighted ssSTEAM MRI, although high undersampling factors with correspondingly increased readout flip angles should improve the SNR. Here, we demonstrate the feasibility of undersampled radial diffusion-weighted ssSTEAM MRI of the human brain in combination with image reconstruction by the iteratively regularized Gauss-Newton method [3].

Methods

The diffusion-weighted ssSTEAM sequence (Fig. 1) prepares the magnetization with a diffusion-encoded spin echo, followed by a 90˚ pulse. A series of small angle pulses generates stimulated echoes, one for each radial spoke of the sampling trajectory. A fat saturation pulse is applied before each slice. Multiple slices without gaps are acquired in an interleaved manner using a sufficiently long TR. The procedure is repeated for each b-value and each diffusion direction. All studies were performed at 3 T (Magnetom Prisma, Siemens Healthcare) using the standard 64-channel head coil. Validation measurements were performed for the diffusion coefficient of distilled water using b-values of 0, 100, …, 1000. The imaging protocol for healthy volunteers contained the following parameters: FOV: 192 × 192 mm2, in-plane resolution: 1.5 mm, slice thickness: 5 mm, sampling: 25 spokes, TE: 7.44 ms, spin-echo-TE: 36.1 ms, TRS: 6.89 ms, TR: 12 s, TM: 10 ms. With these parameters, one image with b = 0 and 6 diffusion-weighted images in different directions with a b-value up to 1000 can be acquired in 84 s for 40 slices covering the entire brain. Image reconstruction was based on the iteratively regularized Gauss-Newton method [3] modified to jointly estimate the image content and coil sensitivities of the entire volume, i.e. all slices. The original image regularization, penalizing the L2-norm of the estimated data, is replaced by a regularization which penalizes high values of $$$h(ρ,c)=‖ρ‖_2^2+∑_{i=2}^{s-1}‖c_i-0.5β(c_{i+1}+c_{i-1})‖_2^2 $$$ with ρ the image content and ci the coil sensitivity profile in slice i, β is a regularization parameter. The procedure takes advantage of the spatial smoothness of coil sensitivities perpendicular to the image section. Post-processing involved a novel image denoising filter using a modified non-local means algorithm [4].

Results and discussion

The diffusion coefficients measured in distilled water were in agreement with literature values [5], see Table 1. In vivo experiments yielded anatomically correct diffusion-weighted images of the brain with acceptable SNR and no averaging (Fig. 2). No further post processing (e.g., for removal of artifacts) was necessary. The present results demonstrate that undersampled radial ssSTEAM MRI, in combination with IRGNM-based image reconstruction, is a potentially useful strategy for clinical diffusion-weighted MRI, e.g. for tumor and stroke diagnosis. Its performance will be further tested in studies of the brain as well as other organs such as the liver to explore the range of suitable clinical applications.

Acknowledgements

No acknowledgement found.

References

[1] Nolte et al., MRM 2000, 44:731-6

[2] Karaus et al., JMRI 2007, 26:1666-71

[3] Uecker et al., MRM 2008, 60:674-82

[4] Buades, CVPR 2005, 2:60-5

[5] Holz et al., PCCP 2000, 2:4740-72.

Figures

Figure 1: Diffusion-weighted single-shot STEAM sequence

Figure 2: Two sections of a human brain acquired with diffusion-weighted ssSTEAM MRI. Top: Mean diffusion-weighted images averaged across 6 directions with b = 500 s mm-2, bottom: b = 0 images

Table 1: Directional diffusion coefficients of distilled water in units of 10-3 mm2 s-1 using diffusion-weighted ssSTEAM



Proc. Intl. Soc. Mag. Reson. Med. 24 (2016)
0216