Nadège Corbin1, Jonathan Vappou1, Pramod Rao1, Benoit Wach1, Laurent Barbé1, Pierre Renaud1, Michel de Mathelin1, and Elodie Breton1
1ICube-University of Strasbourg, Strasbourg, France
Synopsis
MR-guided
percutaneous thermal ablations are currently monitored by MR thermometry.
However, no information related to intrinsic property changes of the tissue is
available during the procedure. The feasibility of monitoring in vivo thermal
ablations by simultaneous Magnetic Resonance Elastography (MRE) and
MR-thermometry is demonstrated in this work. The interventional MRE system
includes a needle MRE driver, a respiratory triggered gradient-echo sequence with
motion encoding and an online reconstruction method that provides elasticity
and temperature measurements in real-time.
Changes in elasticity and temperature occurring during laser thermal
ablation are successfully measured in vivo over 20 minutes thanks to this
interventional MRE system.Purpose
Percutaneous hyperthermal ablations are intended to destroy
cancerous tissues by locally increasing the temperature over 60°C. The MRI guidance of these minimally-invasive
procedures offers the possibility of measuring the temperature in real-time. Nevertheless, no information related to structural
properties of the tissue is available during the procedure. Measuring elasticity during the procedure has
been proposed and liver stiffness has been shown to increase after a thermal
ablation
1,2. Magnetic Resonance
Elastography (MRE) allows measuring the mechanical properties of tissue in vivo
from MR phase images by estimating the wave propagation speed generated by a
mechanical exciter
3. However,
current MRE systems do not meet the demanding conditions of interventional MRI
in terms of bulk and update rate. An all-in-one interventional MRE system has been
proposed, and feasibility of monitoring elasticity changes in real-time using
this method has been demonstrated on phantoms
4. The objective of this work is to demonstrate the
feasibility of monitoring thermal ablations in vivo in real-time using interventional
MRE combined to MR thermometry.
Method
The interventional MRE system is
composed of: 1/ a vibrating needle MRE driver that generates waves
directly within the region of interest, 2/ a fast and interactive spoiled
gradient-echo MR pulse sequence that encodes the motion on MR phase images, and
3/ an online LFE-based inverse problem solver that reconstructs elasticity maps
in real-time. The MRE pulse sequence has been adapted to allow for respiratory
triggering during the long in vivo thermal ablation procedures, making continuous
monitoring of the procedure possible without any need for breath-holding. Each
respiratory trigger is followed by a mechanical excitation trigger, dummy cycles that allow the steady-state to be reached and finally, two phase
images with opposite gradient polarities. On the one hand, these two images are
subtracted to enhance the wave image and remove the background phase. On the other hand, these images are also used for temperature
measurement by adding them together 5, allowing simultaneous MRE and MR
thermometry. The shift between the respiratory
trigger and the mechanical excitation trigger is incremented so that three
phase-offsets, covering one mechanical period, are acquired every three
respiratory cycles (Fig.1). Thanks to a sliding window scheme, one elastogram
is reconstructed with every respiratory cycle, i.e. with every new phase
difference image.The number of MRE phase offsets is reduced to three in order
to optimize elasticity measurements while minimizing temporal smoothing effects4.
First, in order to assess the stability of the system, 127 elastograms are acquired during 10 minutes with the described
protocol in a swine liver in vivo.
Second, a thermal laser ablation
(DIOMED 25) is performed in another swine liver in vivo and is monitored in real-time over 20 minutes by
simultaneous MRE and MR thermometry. Four optical fibers are inserted in the
liver by an experimented interventional radiologist (Fig.2). A total power of 12 W is delivered through
the fibers during the first 10 minutes of the procedure. The needle MRE driver is positioned so that mechanical waves are
propagating through the planned ablation zone, i.e. at the optical fiber tips.
The animals are
continuously ventilated with a respiratory rate of 12 breaths per minute. A
respiratory belt is placed on their abdomen for triggering during the expiration
phase. Relevant MRE parameters include:
excitation frequency 60 Hz, encoding frequency 90 Hz, acquisition matrix
102×128, GRAPPA ×2, MEG amplitude 20 mT/m, TE/TR 9.34/16.67ms, flip
angle 13°, through slice encoding, one slice orthogonal to the needle MRE
driver. Total acquisition time per respiratory cycle is 2.76s corresponding to
the dummy cycles and the acquisition of the two images.
Results
A standard deviation inferior to 5 % of the average shear
modulus is obtained during the first experiment (Fig.3). Wave images acquired during the ablation are shown in Fig.4
The simultaneous evolutions of shear modulus and temperature in the lesion are
plotted in Fig.5. When the laser is turned on, the wavelength continuously
increases in the fiber tip zone corresponding to the increase of the shear
modulus in the elastograms. As expected, the target ablation temperature of 60°C
is reached after 10 minutes. When the laser is turned off, the temperature
decreases while the shear modulus stabilizes in the lesion.
Conclusion
For the first time, in vivo
thermal ablation is monitored by simultaneous MR elastography and MR
thermometry with a refresh rate of one elastogram with every breathing cycle,
i.e. every 5 seconds. This experiment highlights the complementarity of simultaneous
temperature and elasticity monitoring during thermal ablations.
Acknowledgements
This work was supported by French state funds managed by the ANR within the Investissements d'Avenir programme for the Labex CAMI (ANR-11-LABX-0004) and the IHU Strasbourg (ANR-10-IAHU-02).References
1.
Arnal, B. Pernot, M. et Tanter, M. 2011. Monitoring of thermal therapy based on shear modulus changes: II. Shear
wave imaging of thermal lesions. IEEE Transactions on Ultrasonics,
Ferroelectrics, and Frequency Control 58 (8): 1603-11
2.
Chen, J. Woodrum, D.A. , Glaser, K.J. Murphy, M.C. Gorny, K. and Ehman,R. 2013. Assessment of in Vivo Laser
Ablation Using MR Elastography with an Inertial Driver. Magnetic
Resonance in Medicine 72(1):59-67
3.
Muthupillai, R., Lomas, D.J. Rossman, P.J. Greenleaf, J.F. Manduca, A.
and Ehman. R.L. 1995. Magnetic resonance elastography by direct
visualization of propagating acoustic strain waves. Science 269
(5232): 1854-57.
4.
Corbin, N. Vappou, J. Breton, E. Boehler, Q. Barbé, L. Renaud, P. and de Mathelin, M. 2015. Interventional MR Elastography for MRI-Guided
Percutaneous Procedures. Magnetic Resonance in Medicine, doi:10.1002/mrm.25694.
5. Yuan, L. Glaser, K. Rouviere, O. Ehman, R. et Felmlee, J.P. 2006. Feasibility of Simultaneous Temperature and Tissue
Stiffness Detection by MRE. Magnetic Resonance in Medicine 55
(3): 700-705.