Yuxin Zhang1 and William A Grissom2
1Biomedical Engineering, Tsinghua University, Beijing, China, People's Republic of, 2Biomedical Engineering, Vanderbilt University Institute of Imaging Science, Nashville, TN, United States
Synopsis
Signal loss induced by ablation probe prevents accurate
temperature monitoring where the thermal dose is highest. To address this problem, a dual echo sequence with z-shimming is proposed to recover the signal and an associated penalized likelihood approach is applied to estimate a single temperature map from both echoes. Phantom experiments were conducted to validate the effect of the proposed sequence. Evident signal recovery is shown in the magnitude images and temperature maps with heating. Standard deviation maps with no heating are presented to reflect the large reduction
in uncertainty over time with dual-echo z-shimmed thermometry.PURPOSE
Proton
resonance frequency-shift MR thermometry has been widely applied to monitor
temperature changes during thermal ablation. However, in RF, microwave and
laser ablations, the ablation probe may induce signal loss in tissue
surrounding it due to susceptibility-induced
B0 field gradients, which prevents accurate temperature
monitoring where the thermal dose is highest. To address this problem, we
present a dual echo sequence with z-shimming
1 that recovers signal around metallic
ablation probes, and an associated penalized likelihood approach to estimate a
single temperature map from both echoes.
METHOD
Sequence Figure 1 shows the dual
echo z-shimming sequence. To recover signal around the probe, the first echo is
partially refocused in the slice dimension by reducing the slice rephasing lobe
immediately preceeding it to p% of its full-refocusing area. Signals away from
the probe are fully refocused by a second slice refocusing gradient placed
between the two echoes with q% = 100% - p% of the full refocusing gradient area,
so the second echo images the rest of the tissue away from the probe, and has a
typical PRF thermometry echo time.
Temperature estimation The multi-echo
hybrid algorithm 2,3 was adapted to jointly estimate temperature changes from the two echoes.
The algorithm works by jointly fitting the following model to the echo images:$${\widetilde{y}_j}=(\sum\limits^{N_{b}}_{l=1}b_{j,l}w_{l})e^{i(\left\{Ac\right\}_{j}+f_{j})T_{E}}+\epsilon_{j},$$where $$$j$$$ indexes pixels, $$$N_{b}$$$ is the number of baseline images, $$$w_{l}$$$ is the weighting of baseline image $$$b_{l}$$$. $$$A$$$ is a polynomial matrix
with coefficient vector $$$c$$$ to
model background field drifts, $$$f$$$ is
the temperature-induced frequency shift, $$$T_{E}$$$ is echo time, and $$$\epsilon$$$ is
complex Gaussian noise. The model is fit using an iterative gradient-based
algorithm, while regularizing for a sparse temperature map since heating is
focal.
Phantom
experiments
The sequence was
implemented on a 3T scanner (Philips Achieva, Philips Healthcare). Two sets of
axial images of an agar phantom with a nitinol wire (diameter=3mm) inserted in
it parallel to the B0
field were acquired
using an 8-channel head coil (z-shim p=50%, TE1=10ms, TE2=16ms, TR=30ms,
FOV=150mm×150mm, slice thickness=3mm). The p value used in these experiments
was equivalent to 0.81 cycles of phase through the slice. The first data set
comprised 100 dynamic images without heating to compare temperature standard
deviation (STD) maps of dual-echo z-shim and single-echo temperature maps. In
the second data set, heating was performed with a microwave generator to validate
recovery of temperature mapping precision near the probe. An additional spin
echo image of the same slice was also acquired (TE=10ms, TR=600ms) to illustrate the size of the wire and the extent of the signal loss it
created.
RESULTS
Fig.
3 shows the magnitude of the first and second echo images, and complex combined
images from the dual echo z-shim sequence. The diameter of the signal loss
region in the second echo GRE image was approximately 12.8 mm. The diameter of
the signal loss region combined dual echo image was more similar to the spin
echo image, approximately 3.8 mm. Compared to single (second) echo temperature
maps, the temperature maps estimated from both echoes using the dual-echo
hybrid algorithm showed significant temperature closer to the wire. Fig.4 shows
that the dual echo z-shim temperature maps have lower STD than single-echo maps
around the metallic wire. Temperature curves of a pixel near the wire also
reflect a large reduction in uncertainty over time with dual-echo z-shimmed
thermometry.
DISCUSSION
A
dual echo sequence with z-shimming and an associated temperature reconstruction
algorithm were proposed, which can recover signal for precise temperature
mapping near metallic wires. The z-shim value can be selected manually before
heating or automatically chosen using a self-adaptive method as in [4].
Acknowledgements
This work was
supported by NIH Grant R21 NS 091735 and Tsinghua University
Initiative Scientific Research Program (20141081231).References
[1] JH. Seppenwoold, MA.
Viergever, C. Bakker. Passive tracking exploiting local signal conservation:
the white marker phenomenon. Magnetic Resonance in Medicine, 2003, 50(4):
784-790.
[2] WA. Grissom, V.
Rieke, AB. Holbrook, Y. Medan, M. Lustig, J. Santos, et al. Hybrid referenceless
and multibaseline subtraction MR thermometry for monitoring thermal therapies
in moving organs. Medical Physics. 2010, 37(9): 5014-5026.
[3] P. Gaur and WA. Grissom.
Comparison of single- and multi-echo PRF-shift thermometry and method for
penalized-likelihood multi-echo temperature reconstruction. ISMRM 2014, p.
2351.
[4] K. Heberlein, X.
Hu. Auto-Calibrating the Z-shim for Slice Specific Recovery of Susceptibility
Losses. ISMRM 2005, p. 2271