Solenn Toupin1,2, Matthieu Lepetit-Coiffe2, Pierre Bour1, Valery Ozenne1, Baudouin Denis de Senneville3, Rainer Schneider4, Kimble Jenkins5, Arnaud Chaumeil1, Pierre Jais1, and Bruno Quesson1
1IHU-LIRYC, Bordeaux, France, 2Siemens Healthcare, Saint Denis, France, 3Mathematical Institute of Bordeaux, Bordeaux, France, 4Siemens Healthcare, Erlangen, Germany, 5MRI Interventions, Irvine, CA, United States
Synopsis
The visualization of lesion formation in real time is
one potential benefit of carrying out radiofrequency ablation (RFA) under
magnetic resonance (MR) guidance in the treatment of ventricular arrhythmia. In
this study, we propose a real-time MR thermometry method to visualize online
the temperature distribution in the myocardium during catheter-based RFA. An
echo-navigated sequence is used with slice tracking to compensate respiratory-induced through-plane
motion and allow all image orientation. The
method was evaluated during free breathing in 5 healthy volunteers and during RF
delivery on the left ventricle (LV) of a sheep in vivo.Purpose
Catheter-based
radiofrequency ablation (RFA) has become a reference curative therapy for the
treatment of cardiac arrhythmias. Magnetic Resonance (MR) Thermometry based on the water proton resonance frequency
shift (PRFS) may provide real-time visualization of temperature distribution
in the myocardium during RFA
1 for improved safety and online assessment of
the therapy outcome. However, to allow accurate temperature mapping for any
slice orientation, real-time slice tracking must be employed. Moreover, rapid
and robust motion correction algorithms are mandatory to compensate
susceptibility related phase (and therefore temperature) variations. In this
study, we propose to combine an echo-navigated EPI sequence with slice tracking
to allow arbitrary
positioning of the images
2,3 and optical flow algorithms to correct
residual in-plane motion and susceptibility artefacts in the heart. This
approach was evaluated on the left ventricle (LV) in five healthy volunteers
without heating and in vivo in a large animal model during RFA, with a spatial
resolution of 1.6x1.6x3 mm
3 and 3 to 5 temperature slices acquired
at each cardiac cycle.
Methods
MR protocol: MR
Temperature imaging was performed on a 1.5T MR system (Avanto, Siemens Healthcare, Germany) using two 16-channel cardiac coils. A fat-saturated, single-shot
Echo Planar Imaging (EPI) sequence was combined with GRAPPA (acceleration
factor of 2) to achieve a 1.6x1.6x3 mm3 spatial resolution, TE/TR/FA=16-18ms/100ms/60°, bandwidth = 1576 Hz/px, FOV = 180x180x3 mm3. Up
to five slices (slice gap = 0.3 mm) were acquired at each heart beat during 250
consecutive cardiac cycles with ECG triggering. A crossed-pair navigator was
located on the diaphragm to adjust the slice position in real-time. Saturation
slabs were positioned along the FOV in the phase encoding direction to avoid
aliasing and two additional saturation bands were set parallel to the image
slices to reduce signal of blood (Fig.1-A).
MR Thermometry: Residual in-plane motion were compensated online using a Principal
Component Analysis (PCA) based optical flow algorithm4. The
overall magnetic field variations with respiration were approximated as the sum
of linear phase changes of each motion displacement on a pixel-by-pixel basis
giving a parameterized magnetic field model5. During the interventional
procedure, the current flow field was used to reconstruct magnetic field distribution
from the parameterized model. Spatio-temporal phase drift was corrected and a
temporal low-pass Butterworth filter was added at the end of the thermometry
pipeline. Image reconstruction and all correction algorithms were implemented
in C++/GPU and integrated into the Gadgetron reconstruction framework6, with
a total processing time below 120 ms/image.
Volunteer study: MR-guided
thermometry was evaluated in short axis orientation on volunteers (N=5) under
free-breathing conditions. Pixel-wise temporal standard deviation of
temperature σT was calculated to assess the precision of the
thermometry.
Radiofrequency ablation on a sheep: Two
MR-compatible catheters were inserted into the sheep heart (Fig.2-A) under
fluoroscopic guidance (Toshiba InfiniX, Toshiba Medical, Japan). The first
catheter (MRI Intervention, U.S.A.) was located into the left ventricle (LV)
and connected to a RF generator outside
the Faraday cage. The second catheter (BiosenseWebster, Israel) was positioned
into the right ventricle (RV) for heart pacing (140 bpm). RFA was run for 85s
at 40 W simultaneously to MR thermometry. The temperature evolution was
visualized in real-time using Thermoguide software (Image Guided Therapy, France).
Results
On each volunteer, σ
T remained
below 2°C over the whole LV (Fig.1-A), except for a limited number of pixels
located in area where residual phase wrapping were not properly corrected
(Fig.1-C). On the animal, σT was found below 1°C due to a lower and
more regular respiratory amplitude (1.25 cm, see Fig.2-B, mechanically
controlled ventilation) and after excluding 2% of the pixels in the LV
corresponding to the uncorrected phase jump. During the ablation, a temperature
increase of approximately 30°C was observed in real-time in the vicinity of the
catheter tip (Fig.2-C,D).
Discussion/Conclusion
Real-time PRFS-based cardiac MR Thermometry
using an echo-navigated GRE-EPI sequence can achieve average temperature
stability of 2°C or better in the myocardium, combining cardiac triggering,
parallel imaging and PCA-based motion and susceptibility artefacts correction
algorithms. The spatial (1.6x1.6x3 mm
3) and temporal resolution allow direct
visualization of temperature evolution in the myocardium during RFA at each heart beat, with full
flexibility on image orientation provided by online, navigator-based, slice
tracking.
Acknowledgements
ANR TACITReferences
1. B.D. de Senneville et al.
NMR Biomed.,2012,25:556-562.
2. M. Ries et al. MRM,2010,64(6):1704-12
3. Z. Celicanin et al. MRM 2014,72(4):1087-95
4. B.D. de Senneville et al. IEEE Trans Med Imaging,2015;34(4):974-982.
5. G. Maclair et al. MICCAI,2007,411-419
6. M. Hansen et al, MRM,2013,69,1768-1776.