Eddy B Boskamp1, Saikat Saha1, Ricardo Becerra1, and Michael Edwards1
1Engineering, GE Healthcare, Waukesha, WI, United States
Synopsis
In this study we are comparing 16 channel TEM, 8 channel TEM
and 8 loop array pTx body coils based on experimental data obtained from full
scale whole body prototypes as opposed to only simulation. Besides SAR,
efficiency and uniformity, there are additional criteria to include when
selecting a body coil for parallel transmit. Examples are star intensity
artifact, E fields that heat up cables and baluns, VSWR, and perturbation
sensitivity, which may make it impossible to build a certain design given
obtainable tolerances.Introduction
Degenerate
birdcages were discussed by Nistler et al(1) and Alapappan et al(2). Loop
arrays have been discussed by Alagappan et al(3) and Guerin et al(4), and TEM
arrays by Ryu et al(5) and Tian et al(6). Instead of just doing comparative
simulations, in our study we are comparing data from actual prototypes of full
scale body coils and examine the real engineering issues with implementation.
Methods
A
16 ch TEM array was built with a patient bore of 68 cm (fig1). The rungs were
26 mm wide, 420 mm long and had 4 capacitive junctions of about 36 pF. Distance
to the RF shield was 20 mm. The 8 channel version is the 16 ch TEM with half of
the rungs removed. An 8 loop array was built with a 70 cm patient bore (fig2).
The loops were 525 mm long, and 340 mm wide, traces were 20 mm wide (45cm FOV
at -6dB). The distance to the RF shield was 20 mm. The loops overlapped by 24
mm to minimize mutual inductance between direct neighbors. There were 9 capacitive
junctions per loop. The measurements on these 3 coils were compared to a 16
rung high pass birdcage coil with a diameter of 70 cm, rung width and length of
7 and 39 cm, and endring width of 10 cm. Distance to RF shield 20 mm (45 cm FOV
at -6dB). A scale model degenerate birdcage was also built with a diameter of
242mm, length 192mm. The distance to the RF shield was 14 mm. The parallel
transmit coils were connected to an 8 channel transmit chain with amplifiers
mounted close to the magnet (VSWR requirement<2). HFSS ( Ansys inc) finite
element simulations were performed in addition to the measurements.
Results
The
degenerate birdcage was found to be extremely sensitive to perturbation of
symmetry e.g. roundness of coil, patient or shield, as well as alignment. The
best isolation between channels that could be obtained was -9 dB, insufficient
for a VSWR<2. For this reason the degenerate birdcage was not considered for
full scale duplication. The following data is for 16ch TEM, 8ch TEM and 8 loop
array, in that order. Measured worst case isolation between neighbors was -14, -18 and -18 dB. The -14 dB isolation was
not enough to obtain a VSWR <2. The measured star artifact levels due to
aliasing of signal from non linear areas of the gradient was 3.1, 3.1 and 1
(Birdcage =1). The measured efficiency in an empty coil driven in quadrature
was B1+ = 0.157, 0.120 and 0.122 μT/√W (birdcage= 0.243). When the coils were
loaded with a load phantom equivalent to a 75 kg patient (590mm long cylinder,
295mm ID, 345mm OD, eps=80, sigma=0.7 S/m) the efficiencies were B1+= 0.14,
0.10, and 0.10 μT/√W (Birdcage = 0.154) when driven in quadrature. TEM E fields
are primarily in the Z direction. These E fields (empty) are shown in fig 3 and
compared to the 8 loop array and the birdcage. Fig 4 shows the Z directed E
fields for the loaded coils (45 cm diameter solid cylinder, 50 cm long of
muscle tissue eps=68, sigma=0.7S/m).
Discussion
The
8 loop array performs the best in VSWR, B1+ drop off (star artifact) and E
field. The TEM performs worst for Z directed E field and Star artifact, where
the 16 channel TEM also has the worst VSWR performance. Outside the imaging
volume the TEM B1+ only drops gradually, and not fast enough to prevent star
artifact from the non linear areas of the gradient. TEMs can be built that have a faster drop off
of B1+ as was done in (7). The drop off can even be made birdcage like by
adding spoilers to the TEM, but at the expense of efficiency. The Z directed E
field of the TEM impacts the common mode current and balun heating in system
cables that run through the bore, as well as blocking network heating for receive
array elements. Previous work has concentrated on uniformity and SAR as the
criteria of choice, there are practical considerations like perturbation
sensitivity, efficiency, star artifact, VSWR, and E fields that should be
included when making a choice of body coil.
Acknowledgements
No acknowledgement found.References
1.Nistler et
al, A degenerate bandpass birdcage as antenna for a 3T whole body transmit
array, Proceedings ISMRM 2006, 2566.
2.Alagappan
et al, A degenerate birdcage coil for parallel excitation, Proceedings ISMRM,
2007, 1028.
3.Alagappan
et al, Comparison of 3 transmit arrays for parallel transmit, Proceedings ISMRM
2007, 165.
4.Guerin et
al, Simulation study of parallel transmit arrays for 3T whole body imaging
under global and local SAR constraints, Proceedings ISMRM 2012, 2612.
5.Ryu et al,
Comparison of element geometries in B1 shimming with a 16 channel whole body
transmit array at 3T, Proceedings ISMRM 2009, 3046.
6.Tian et
al, Searching for the optimal body coil design for 3T MRI, Proceedings ISMRM
2013, 2746.
7.Leussler
et al, U shaped ladder TEM coil structure with truncated sensitivity profile in
z direction for high field MRI, Proceedings ISMRM 2012, 2805