Looping star: A novel, self-refocusing zero TE imaging strategy
Ana Beatriz Solana1, Anne Menini1, and Florian Wiesinger1

1GE Global Research, Garching bei Muenchen, Germany

Synopsis

Zero TE is an extremely efficient 3D pulse sequence which also has the advantages of low geometrical distortion, reduced acoustic noise and the capacity of imaging short T2 structure. However, its native contrast is proton density. Here we present a novel method that allows gradient refocusing at echo times suitable for fMRI or susceptibility weighted imaging. As a proof of concept, this new imaging strategy is tested in phantom experiments.

Purpose

After its invention more than two decades ago [1,2,3], Zero TE has experienced a recent revival [4,5] with many interesting clinical applications [6,7,8,9]. Here, we describe a novel Zero TE method that allows multi-gradient-echo refocusing and its implementation and testing in phantom experiments.

Methods

Zero TE is a 3D pulse sequence consisting of 1) short non-selective RF excitation (typically below 5deg) and 2) a constant readout gradient with only directional updates in between repetitions. Image encoding happens in a 3D radial manner and starts immediately at the moment of RF excitation (TE=0). The Zero TE pulse sequence native contrast is proton density; however, different contrast can be added contrast preparation modules before the readout using e.g. inversion-recovery, or T2 preparation modules.

A novel, self-refocusing zero TE imaging pulse sequence (termed Looping star) is implemented as illustrated in Fig. Without loss of generality, we consider a single segment of 2*32=64 3D radial spokes (upper subplot) arranged in a 2D circle (lower left subplot) with a constant phase increment of 2*pi/32 (black arrows). Hence after 32 repetitions each spoke gets refocused and forms a gradient echo of the original FID signal (red connected arrows). In this way gradient refocusing can be achieved with minimal gradient switching and hence negligible acoustic noise; an important characteristic of zero TE imaging especially for silent imaging applications. In the pulse sequence (top plot) 32 spokes are first excited and encoded along the 3D spokes. Afterwards each of the early FID echoes gets sequentially refocused and forms a gradient echo. Hence for the latter 32 spokes, RF excitation is turned off. The refocusing can be repeated to form more echoes. For the following segments, the 3D spokes are rotated along the z-axis to obtain 3D coverage as illustrated in the lower right plot.

Results

The Looping star pulse sequence was implemented on a GE MR750w 3T scanner (GE Healthcare, Waukesha, WI) and tested in phantom experiments. Each segment consisted of 128 spokes with the first 32 spokes encoding FID echoes and the latter ones encoding three gradient echoes. The ZTE parameters were chosen to be FA=3deg, TRspoke=1ms, FoV=19.2cm, 3x3x3 mm resolution, total_spokes=8192, acquisition BW=±15.6kHz. Accordingly, the echo time amounted to TE=32*TR=32ms and the total scan time was 12 seconds.

Figure 2 shows 3 orthogonal planes of the spherical phantom for the FID image (top) and the first gradient echo image (middle). The obtained quantitative T2* = 45ms was uniform and consistent with the value obtained from a conventional 3D Cartesian spoiled gradient echo measurement.

Discussion

The presented Looping star method describes a novel scheme for gradient refocusing which differs from the conventional multi gradient echo readout following each excitation. Instead, the refocusing happens over multiple excitations; in a way similar to the PRESTO concept [10]. That way, the high efficiency of the Zero TE sequence and its silent imaging characteristic can be maintained. Also, its multiecho capability shows the potential of this imaging strategy for T2* quantification methods. As a next step, this method will be investigated in vivo for quantitative susceptibility mapping (QSM) and fMRI.

Acknowledgements

No acknowledgement found.

References

[1] Hafner S (1994) Fast imaging in liquids and solids with the backprojection Low Angle ShoT (BLAST) technique. Magn Reson Imaging 12(7):1047–1051

[2] Madio DP, Lowe IJ (1995) Ultra-fast imaging using low flip angles and fids. Magn Reson Med 34:525–529

[3] Wu Y, Ackerman JL, Chesler DA, Graham L, Wang Y, Glimcher MJ (2003) Density of organic matrix of native mineralized bone measured by water-and fat-suppressed proton projection MRI. Magn Reson Med 50:59–68

[4] Weiger M, Pruessmann KP, Hennel F (2011) MRI with zero echo time: hard versus sweep pulse excitation. Magn Reson Med 66(2):379–389

[5] Grodzki DM, Jakob PM, Heismann B (2012) Ultrashort echo time imaging using pointwise encoding time reduction with radial acquisition (PETRA). Magn Reson Med 67:510–518

[6] Weiger M, Brunner DO, Dietrich BE, Müller CF, Pruessmann KP (2013) ZTE imaging in humans. Magn Reson Med 70(2): 328–332

[7] Gibiino F, Sacolick L, Menini A, Landini L, Wiesinger F (2014) Free-breathing, zero-TE MR lung imaging. Magnetic Resonance Materials in Physics, Biology and Medicine, 28(3):207–215.

[8] Solana AB, Menini A, Sacolick LI, Hehn N, Wiesinger F (2015) Quiet and distortion-free, whole brain BOLD fMRI using T(2) -prepared RUFIS. Magn ResonMed. doi: 10.1002/mrm.25658

[9] Wiesinger F, Sacolick LI, Menini A, Kaushik SS, Ahn S, Veit-Haibach P, Delso G, Shanbhag DD (2015) Zero TE MR bone imaging in the head. Magn Reson Med. doi: 10.1002/mrm.25545.

[10] Liu G, Sobering G, Duyn J, Moonen CT. A functional MRI technique combining principles of echo-shifting with a train of observations (PRESTO) (1993) Magn Reson Med.;30:764–768

Figures

Fig. 1: Looping star simplified pulse sequence (top) and spoke arrangement (bottom left) is resulting in this example in an FID train and one train of gradient refocused echo. Via rotation of 2D planes along the z-axis, full 3D k-space coverage is obtained (bottom right).

Fig. 2: Looping star FID echo (top) and 1st gradient echo (middle) images of a uniform spherical phantom. The obtained T2* map (bottom) is uniform and matches the expected T2*=45ms.



Proc. Intl. Soc. Mag. Reson. Med. 24 (2016)
0104