Ana Beatriz Solana1, Anne Menini1, and Florian Wiesinger1
1GE Global Research, Garching bei Muenchen, Germany
Synopsis
Zero TE is an extremely efficient 3D pulse sequence which
also has the advantages of low geometrical distortion, reduced acoustic noise
and the capacity of imaging short T2 structure. However, its native contrast is proton density. Here we present a novel method that allows gradient
refocusing at echo times suitable for fMRI or susceptibility weighted imaging. As
a proof of concept, this new imaging strategy is tested in phantom experiments.Purpose
After its invention more than two decades ago [1,2,3], Zero
TE has experienced a recent revival [4,5] with many interesting clinical applications
[6,7,8,9]. Here, we describe a novel Zero TE method that allows
multi-gradient-echo refocusing and its implementation and testing in phantom
experiments.
Methods
Zero
TE is a 3D pulse sequence consisting of 1) short non-selective RF excitation (typically
below 5deg) and 2) a constant readout gradient with only directional updates in
between repetitions. Image encoding happens in a 3D radial manner and starts
immediately at the moment of RF excitation (TE=0). The Zero TE pulse sequence native
contrast is proton density; however, different contrast can be added contrast
preparation modules before the readout using e.g. inversion-recovery, or T2
preparation modules.
A novel, self-refocusing zero TE imaging pulse sequence
(termed Looping star) is implemented as illustrated in Fig. Without loss of generality, we consider a
single segment of 2*32=64 3D radial spokes (upper subplot) arranged in a 2D
circle (lower left subplot) with a constant phase increment of 2*pi/32 (black
arrows). Hence after 32 repetitions each spoke gets refocused and forms a
gradient echo of the original FID signal (red connected arrows). In this way gradient refocusing can be
achieved with minimal gradient switching and hence negligible acoustic noise;
an important characteristic of zero TE imaging especially for silent imaging
applications. In the pulse sequence (top
plot) 32 spokes are first excited and encoded along the 3D spokes. Afterwards each of the early FID echoes gets sequentially
refocused and forms a gradient echo.
Hence for the latter 32 spokes, RF excitation is turned off. The refocusing can be repeated to form more
echoes. For the following segments, the
3D spokes are rotated along the z-axis to obtain 3D coverage as illustrated in
the lower right plot.
Results
The Looping star pulse sequence was implemented on a GE
MR750w 3T scanner (GE Healthcare, Waukesha, WI) and tested in phantom
experiments. Each segment consisted of
128 spokes with the first 32 spokes encoding FID echoes and the latter ones
encoding three gradient echoes. The ZTE parameters were chosen to be FA=3deg,
TRspoke=1ms, FoV=19.2cm, 3x3x3 mm resolution, total_spokes=8192, acquisition BW=±15.6kHz. Accordingly, the echo time amounted to
TE=32*TR=32ms and the total scan time was 12 seconds.
Figure 2 shows 3 orthogonal planes of the spherical phantom
for the FID image (top) and the first gradient echo image (middle). The obtained quantitative T2* = 45ms was
uniform and consistent with the value obtained from a conventional 3D Cartesian
spoiled gradient echo measurement.
Discussion
The presented Looping star method describes a novel scheme
for gradient refocusing which differs from the conventional multi gradient echo
readout following each excitation.
Instead, the refocusing happens over multiple excitations; in a way
similar to the PRESTO concept [10]. That
way, the high efficiency of the Zero TE sequence and its silent imaging
characteristic can be maintained. Also, its multiecho capability shows the
potential of this imaging strategy for T2* quantification methods. As a next
step, this method will be investigated in vivo for quantitative susceptibility mapping
(QSM) and fMRI.
Acknowledgements
No acknowledgement found.References
[1] Hafner S (1994) Fast imaging in liquids and solids with
the backprojection Low Angle ShoT (BLAST) technique. Magn Reson Imaging
12(7):1047–1051
[2] Madio DP, Lowe IJ (1995) Ultra-fast imaging using low
flip angles and fids. Magn Reson Med 34:525–529
[3] Wu Y, Ackerman JL, Chesler DA, Graham L, Wang Y,
Glimcher MJ (2003) Density of organic matrix of native mineralized bone
measured by water-and fat-suppressed proton projection MRI. Magn Reson Med
50:59–68
[4] Weiger M, Pruessmann KP, Hennel F (2011) MRI with zero
echo time: hard versus sweep pulse excitation. Magn Reson Med 66(2):379–389
[5] Grodzki DM, Jakob PM, Heismann B (2012) Ultrashort echo
time imaging using pointwise encoding time reduction with radial acquisition
(PETRA). Magn Reson Med 67:510–518
[6] Weiger M, Brunner DO, Dietrich BE, Müller CF,
Pruessmann KP (2013) ZTE imaging in humans. Magn Reson Med 70(2): 328–332
[7] Gibiino F, Sacolick L, Menini A, Landini L, Wiesinger F
(2014) Free-breathing, zero-TE MR lung imaging. Magnetic Resonance Materials
in Physics, Biology and Medicine, 28(3):207–215.
[8] Solana AB, Menini A,
Sacolick LI, Hehn N, Wiesinger F (2015) Quiet and distortion-free,
whole brain BOLD fMRI
using T(2) -prepared RUFIS. Magn ResonMed.
doi: 10.1002/mrm.25658
[9] Wiesinger F, Sacolick
LI, Menini A, Kaushik SS, Ahn S, Veit-Haibach P, Delso
G, Shanbhag DD (2015)
Zero TE MR bone imaging in the head. Magn Reson Med. doi: 10.1002/mrm.25545.
[10] Liu G, Sobering G, Duyn J, Moonen CT. A
functional MRI technique combining principles of echo-shifting with a train of
observations (PRESTO) (1993) Magn Reson Med.;30:764–768