Synopsis
Achieving sufficiently good quality co-registration
between the anatomical and functional images is currently a large stumbling
block for laminar fMRI. Here, we present a distortion-matched T1weighted/T1-estimation
mapping approach using two 3D-EPI readouts per inversion, following the MP2RAGE
signal combination. 0.7mm isotropic T1 data with matching
distortions to a 0.7mm isotropic fMRI protocol can be acquired in less than two
minutes.Target audience
Anyone working on laminar fMRI
Purpose
To
generate resolution and distortion-matched T1-maps and bias-field
corrected T1w-images for use as anatomical reference to
submillimetre fMRI.
The
co-registration of high resolution functional data is problematic because of
(small) local and global distortions which complicate alignment with the
structural scans. The consequences of misalignment are further enhanced when trying
to disentangle functional processes as a function of cortical depth. Cortical
depth surfaces, laminae, are usually defined in the anatomical dataset and
subsequently overlaid on the functional data from which the depth-dependent
signals are derived. Hence, co-registration needs to be correct, locally, with
a precision of around 0.1mm.
Here, the two GRE readouts of the MP2RAGE were replaced with two 3D-EPI blocks, with equal
length readouts as the functional data, to obtain T1 estimation maps
and bias-field corrected T1w images with distortions matched to
those of the functional data.
Methods
A 0.7mm isotropic 3D-EPI protocol(1) for submillimetre fMRI was adapted to
generate the T1-estimation maps and bias-field corrected T1w-images
using the Multiple Interleaved Scanning Sequences,
MISS, environment on the Philips platform. The 3D-EPI parameters were:
FOV 120 x131 x 24 mm, matrix size 132 x 182 x 34, TR/TE = 57/28ms, EPI factor =
27, SENSE undersampling factor 3.5(LR)*1.3(AP), 72 EPI readouts
(segments) per volume, volume acquisition time 4.0s.
The
MP2RAGE sequence(2) contains two gradient echo readout blocks following a single
inversion. Signal from both images is combined following: $$$ MP2RAGE = \frac{GRE_{TI1} GRE_{TI2}}{GRE_{TI1}^{2}
+GRE_{TI2}^{2}} $$$. In the 3D-EPI equivalent, T1-imaging with 2
3D-EPI’s, or T123DEPI, following a single inversion pulse a much
larger k-space section is acquired (up to half or a quarter for this 3D-EPI
protocol), drastically reducing acquisition time. A sequence diagram is shown
in Figure 1. Simulations following(2) were used to derive protocol
parameters for the MP2RAGE and T123DEPI
protocols.
Four volunteers were scanned at 7T (Philips, Netherlands) with a
32-channel surface coil (MRcoils, Netherlands, n=2) or a 32-channel volume coil
(Nova Medical, USA, n=2). Two T123DEPI
protocols were compared, both matching the FOV and matrix sizes of the fMRI
acquisitions:
1)
36 segments per readout (2 blocks), TRT123DEPI
=8.25s, TI1/TI2=1200/3800ms, α1/ α2
= 14/10, 4 averages, total acquisition time 80s.
2)
18 segments per readout (4 blocks), TRT123DEPI
=10s, TI1/TI2=1000/2700ms, α1/ α2
= 20/16, 4 averages, total acquisition time 128s.
MP2RAGE anatomical data were acquired for 3 volunteers (2x volume coil,
1x surface coil): voxel size 0.64 mm isotropic, FOV 205
x 205 x 164 mm, TRMP2RAGE/TE/TI1/TI2 = 8s/6.2/800/2700
ms, α1/ α2 = 7/5, total acquisition time 11min.
For one
volume-coil volunteer, 30 volumes of resting 3D-EPI were also
acquired. For the fourth volunteer, a 6-min functional run with the
corresponding 3D-EPI acquisition was acquired with a visual stimulus and alternating ON=4s,
OFF=10s blocks. Functional data were analysed with SPM12 (GLM, 1mm
smoothing).
Results and Discussion
The division
of the 3DEPI readout in 4 clustered blocks means an optimal TI1 can
be used (Figure 1), whereas in the 2-block protocol the minimum possible TI1
was 1200ms. Both protocols yielded homogeneous T1-weighting, comparable to the MP2RAGE,
throughout the acquisition slab (Figure 2). Although data was combined over a large
period of time (TRT123DEPI*Nblocks = 8.25*4 s instead of 10*2), 4-block data did not demonstrate
more motion sensitivity than the 2-block data. Generally, SNR was higher in the
4block T123DEPI (see Figure 2), in agreement with the simulations results.
The 11-minute MP2RAGE acquisition yielded much superior SNR to the
T123DEPI’s, but did not coregister with the same local fidelity as
the T123DEPI to the mean EPI (Figure 3). Overall
coregistration was successful for all 3 T1w-images (left panels Figure
3), but when the gray matter boundaries of the MP2RAGE are shown overlaid on
the 3D-EPI data, small mis-registrations appear due to local distortions;
around CSF/veins in de sulci (light blue arrow); on the cortical surface (mid
blue arrow) and on the gray-white matter boundary (dark blue arrow). Distortions in the T123DEPI match those of the mean 3D-EPI in the top right panel.
Gray matter T1-values were highly reproducible between
MP2RAGE and
the two T123DEPI protocols (Figure 4).
Functional
activation perfectly aligned with cortical gray matter and/or pial veins
(Figure 5). Generally,
SNR could easily be improved by acquiring more averages given the very short
acquisition times used here.
Conclusion
With the
increased interest in laminar fMRI, high confidence in lamina definition is
becoming more and more important. The iso-resolution, distortion-matched T
1w-images and T
1–maps can provide anatomical reference data which
is straightforward to co-register to the functional 3D-EPI data and can be
acquired in less than two minutes.
Acknowledgements
The authors would like to thank Dr Kamil Ugurbil
for emphasizing the need for a distortion-matched T1w-EPI at the
laminar fMRI meeting in Nijmegen.References
1)
Petridou
et al, 2013, NMR Biomedicine
2)
Marques
et al, 2010, Neuroimage