Real-Time Field Control Using Full 3rd-Order Matrix Pre-Emphasis
Yolanda Duerst1, Bertram J. Wilm1, Benjamin E. Dietrich1, Simon Gross1, Thomas Schmid1, David O. Brunner1, and Klaas P. Pruessmann1

1ETH Zurich, Zurich, Switzerland

Synopsis

Update steps of real-time field control suffer from imperfect shim responses which degrade control quality. By including full 3rd-order matrix pre-emphasis as an additional filter in the control loop, all self-term responses are shaped to be equal and all cross-term responses are directly suppressed. This leads to disturbances being rejected faster and less noise amplification. Thus enables better field control in demanding situations such as caused by disturbance of high spatial and temporal variability.

Introduction

Spatiotemporal magnetic field fluctuations induced by physiological motion or scanner imperfections lead to artifacts in MR imaging1–4. It has recently been shown that real-time field control5 enables stabilization of the field and reduction of related image artifacts. The system is based on multiple NMR field probes to measure spatiotemporal field changes6,7 and a proportional-integral (PI) controller actuating a full 3rd-order spherical harmonics shim system to counteract deviations from the desired field configuration.

Including shim pre-emphasis for self-terms and selected cross-terms to stabilize the controller output was shown to further enhance field control8. However, this implementation neglected the influence of several weaker cross-terms as well as cross-coupling induced by the applied cross-term correction itself.

To address these limitations, we include full 3rd-order matrix pre-emphasis as an additional filter in the control loop, which directly corrects for all self- and cross-term distortions.

Methods

The ultimate goal of the feedback field control system (Fig. 1) is to produce a correction field ($$$b_c$$$) which counteracts disturbances ($$$b_d$$$) such that the total field ($$$b$$$) equals the reference field ($$$b_r$$$). This is achieved by a controller ($$$C$$$) that tries to minimize the error ($$$e$$$) between the reference and the measured total field, and a plant ($$$P$$$, representing the frequency response of a full 3rd-order shim system9) which produces the correction field ($$$b_c$$$).

To improve control performance, the controller was separated into a pure PI controller ($$$C$$$), and a filter for shaping and decoupling of the control channels by matrix pre-emphasis ($$$H$$$).
The controller was tuned for maximum control bandwidth while keeping noise amplification limited.

The pre-emphasis filter ($$$H(\omega)$$$) was designed as a combination of the desired target response

$$H_T(\omega)=e^{-(\omega/1300)^2} Id$$

and the inverse of the known plant response matrix $$$P(\omega)$$$ 10

$$H(\omega)=P(\omega)^{-1}\cdot{H_T(\omega)}$$

Sensitivity function
To evaluate the response of the control system to external disturbances, the sensitivity function11

$$S(\omega)=(Id+P(\omega)H(\omega)C(\omega))^{-1}$$

$$b(\omega)=S(\omega)b_d(\omega)$$

was obtained by applying multisine13 input disturbances ($$$u_d$$$) and measuring the response with and without active field feedback:

$$b^{(feedback)}(\omega)=S(\omega)P(\omega)u_d(\omega)$$

$$b^{(no feedback)}(\omega)=P(\omega)u_d(\omega)$$

In order to calculate the full 16x16 sensitivity function matrix, 16 measurements of $$$b^{(feedback)}$$$ and $$$b^{(no feedback)}$$$ were combined into matrices $$$B^{(feedback)}$$$ and $$$B^{(no feedback)}$$$, respectively,:

$$S(\omega)=B^{(feedback)}(\omega)\cdot B^{(no feedback)}(\omega)^{-1}=[S(\omega)P(\omega)U_d(\omega)][P(\omega)U_d(\omega)]^{-1}$$

The measurement was repeated once with and once without active pre-emphasis in the control loop.

Step Response
To evaluate disturbance rejection efficiency, time domain responses to step disturbances in single channels were observed and the speed of the control was determined by the time needed to reject the disturbance to ≤ 3% of the original step amplitude.

Imaging
During a T2*-weighted gradient echo scan (TR=1sec, TE=25ms, resolution=(0.6mm)2, slice thickness=1.5mm, 10 slices) a volunteer was instructed to move his arm up to his chin and down again every 20 seconds. Imaging results obtained with and without field control, as well as with and without active pre-emphasis in the control loop were evaluated.

Results

Sensitivity
Measured sensitivity functions without pre-emphasis (Fig. 2, blue lines) showed varying self-term responses (diagonal elements) and some strongly coupling 3rd-order channels amplifying the input disturbances by up to a factor of 5 (gray background). When using pre-emphasis in the control loop (red lines), the self-term responses were all the same and cross-terms were suppressed to below 0.12.

Step response
Without pre-emphasis, self-term responses exhibited undershoot and were generally slower than with pre-emphasis (Fig. 3 a&b). Applying pre-emphasis resulted in the actually applied correction steps ($$$\bf\it{b_c}$$$) matching the desired control field ($$$\bf\it{u_c}$$$) thus preventing undershoot. Strong cross-coupling induced fields of up to 220% the original disturbance amplitude (Fig. 3c) that led to disturbance rejection times of ≥ 2.5 seconds (Fig. 4). When pre-emphasis was applied, all cross-terms were suppressed to below 6.5% and disturbance rejection times were reduced to ≤ 1.2 seconds.

Imaging
Field control without pre-emphasis counteracted field distortions from hand motion but induced additional oscillations due to cross-coupling. Including pre-emphasis in the control loop successfully suppressed these oscillations and strongly improved the quality of the obtained T2*-weighted images (Fig. 5).

Discussion & Conclusion

We showed that including pre-emphasis in the field control loop increases the achievable control bandwidth. This is due to all self-term responses being equal, hence optimal tuning of all channels can be achieved simultaneously. Additionally, counteracting coupling between shim channels directly decreases disturbance rejection times and noise amplification.

The proposed method enables stable field control also in demanding situations with disturbances of high spatial and temporal variability such as caused by limb motion of the subject.

Acknowledgements

No acknowledgement found.

References

1. Van de Moortele PF, Pfeuffer J, Glover GH, Ugurbil K, Hu X. Respiration-induced B0 fluctuations and their spatial distribution in the human brain at 7 Tesla. Magn. Reson. Med. 2002;47:888–95. doi: 10.1002/mrm.10145.

2. van Gelderen P, de Zwart JA, Starewicz P, Hinks RS, Duyn JH. Real-time shimming to compensate for respiration-induced B0 fluctuations. Magn. Reson. Med. 2007;57:362–8. doi: 10.1002/mrm.21136.

3. Versluis MJ, Peeters JM, van Rooden S, van der Grond J, van Buchem MA, Webb AG, van Osch MJ. Origin and reduction of motion and f0 artifacts in high resolution T2*-weighted magnetic resonance imaging: application in Alzheimer’s disease patients. NeuroImage 2010;51:1082–8. doi: 10.1016/j.neuroimage.2010.03.048.

4. Versluis MJ, Sutton BP, de Bruin PW, Börnert P, Webb AG, J. OM. Retrospective Image Correction in the Presence of Nonlinear Temporal Magnetic Field Changes Using Multichannel Navigator Echoes. Magn. Reson. Med. 2012;68:1836–1845.

5. Duerst Y, Wilm BJ, Dietrich BE, Vannesjo SJ, Barmet C, Schmid T, Brunner DO, Pruessmann KP. Real-time feedback for spatiotemporal field stabilization in MR systems. Magn. Reson. Med. 2015;73:884–893. doi: 10.1002/mrm.25167.

6. De Zanche N, Barmet C, Nordmeyer-Massner JA, Pruessmann KP. NMR probes for measuring magnetic fields and field dynamics in MR systems. Magn. Reson. Med. 2008;60:176–86. doi: 10.1002/mrm.21624.

7. Barmet C, De Zanche N, Wilm BJ, Pruessmann KP. A transmit/receive system for magnetic field monitoring of in vivo MRI. Magn. Reson. Med. 2009;62:269–76. doi: 10.1002/mrm.21996.

8. Duerst Y, Wilm BJ, Vannesjo SJ, Dietrich BE, Gross S, Brunner DO, Schmid T, Pruessmann KP. Faster Feedback Field Control using Shim Pre-Emphasis. In: Proceedings of the 22st Annual Meeting of ISMRM. Milano; 2014. p. 1387.

9. Vannesjo SJ, Dietrich BE, Pavan M, Brunner DO, Wilm BJ, Barmet C, Pruessmann KP. Field camera measurements of gradient and shim impulse responses using frequency sweeps. Magn. Reson. Med. 2014;72:570–583. doi: 10.1002/mrm.24934.

10. Vannesjo SJ, Dietrich BE, Pavan M, Barmet C, Pruessmann KP. Digital cross-term pre-emphasis for higher-order dynamic shimming. In: Proceedings of the 21st Annual Meeting of ISMRM. Salt Lake City; 2013. p. 0671.

11. Skogestad S, Postlethwaite I. Multivariable feedback control?: analysis and design. 2nd Edition. Chichester: John Wiley & Sons; 2005.

Figures

Block diagram of the control loop. The goal of the control system is to reject disturbances (bd) such that the output (b) follows the reference (br) as close as possible. If there is no pre-emphasis: H = 𝟙, hence upe = uc.

Measured Sensitivity function matrix without pre-emphasis (blue) and with pre-emphasis (red) from input channel (indicated on top) to output channel (indicated on the left).

Frequency range: -20Hz to +20Hz.
Amplitude ranges: 0-1.5 for self-terms (diagonal), 0-0.25 for cross-terms (off-diagonals), and 0-5.5 for three strongly coupling terms (gray background).


Response of the field control system to a step disturbance in ZY (a) and Z2Y (b&c). Self-term responses (a&b) differ from cross-term responses (c).
Two separate measurements are shown: Using field control without pre-emphasis (dashed lines) and with pre-emphasis (solid lines).


Time needed to reject a step disturbance (bd) to 3% of the initial disturbance amplitude. The colors indicate in which channel the disturbance was introduced. The different tunings are indicated by their corresponding maximum of the sensitivity function MS.

In-vivo images acquired while the volunteer repeatedly moved his hand. Without field control (a) strong artifacts are induced that are reduced with field control without pre-emphasis (b) where cross-coupling induces strong oscilatory field distortions after moving the hand. Including pre-emphasis (c) further stabilizes the field and improves image quality.



Proc. Intl. Soc. Mag. Reson. Med. 24 (2016)
0045