Concomitant gradient effects on chemical shift encoded imaging
Timothy J Colgan1,2, Diego Hernando1, Samir D Sharma1, Ann Shimakawa3, and Scott B Reeder1,2,4,5,6

1Radiology, University of Wisconsin, Madison, WI, United States, 2Medical Physics, University of Wisconsin, Madison, WI, United States, 3Global Applied Science Lab, GE Healthcare, Menlo Park, CA, United States, 4Biomedical Engineering, University of Wisconsin, Madison, WI, United States, 5Medicine, University of Wisconsin, Madison, WI, United States, 6Emergency Medicine, University of Wisconsin, Madison, WI, United States

Synopsis

Quantitative chemical shift-encoded (CSE) MRI techniques acquire complex-valued (magnitude and phase) images at multiple echo times (TE), enabling simultaneous mapping of fat-fraction, R2* (=1/T2*) and B0 field. Applications of CSE-MRI include tissue fat quantification, iron quantification and quantitative susceptibility mapping (QSM). Recently, phase shifts due to concomitant gradients (CG) have been identified as a source of error for quantitative CSE techniques, so their effects on fat-fraction, R2* and B0 maps are characterized in this study. CG correction of experimental data demonstrates that the detrimental effects of CG phase shifts can be removed before reconstruction to produce more accurate estimates of the fat-fraction, R2*, and field map measurements.

Purpose

Quantitative chemical shift-encoded (CSE) MRI techniques acquire complex-valued (magnitude and phase) images at multiple echo times (TE), enabling simultaneous mapping of fat-fraction, R2* (=1/T2*) and B0 field. Applications of CSE-MRI include tissue fat quantification, iron quantification and quantitative susceptibility mapping (QSM). However, these techniques are vulnerable to unanticipated phase shifts in the acquired CSE images. Recently, phase shifts due to concomitant gradients (CG) [1] have been identified as a source of error for quantitative CSE techniques [2], however their effects on fat-fraction, R2* and B0 maps are not well understood. Therefore, the purpose of this study was to characterize the effect of phase shifts due to CGs on CSE-based fat-fraction, R2* and B0 measurements, at both 1.5T and 3T

Methods

In this study the phase shifts due to CGs were analyzed for different versions of a multi-echo gradient-echo pulse sequence: a monopolar flyback single echo train, monopolar flyback with two interleaved echo trains, and a bipolar single echo train, as shown in Fig. 1. The phase shifts due to CGs can be calculated from the known gradients [1]. Simulations were used to characterize the CG phase shifts and experimental measurements were used to demonstrate the correction of these shifts before parameter estimation.

Simulations

In our 3T simulations, complex data from a homogeneous water signal (fat-fraction=0%) were synthesized, and CG phase shifts were added. Representative CG phase shifts were calculated using gradients for a 3D six-echo CSE gradient echo sequence with an axial acquisition and a 48 cm z-axis field of view (FOV) for the three waveforms shown in Fig. 1. Subsequently, a complex fitting algorithm [3] was used to estimate the fat-fraction, R2*, and B0 field map.

MRI Measurements

Measurements were made on a 1.5T Signa HDxt GE scanner and a 3T Discovery MR 750 GE scanner using an 8-channel and 32-channel torso coil, respectively. A NiCl2 and NaCl doped-water phantom with 0% fat-fraction and homogenous R2* was scanned. At 3T, the scan was performed with FOV=210 x 210 x 480 cm (X x Y x Z), 6 echoes, TE1=1.2ms, and ∆TE=0.97ms. At 1.5T two acquisitions with interleaved echo trains were made sequentially, one with FOV=210 x 210 x 480 cm, 6 echoes, TE1=1.2ms, and ∆TE=0.98ms and a second with FOV=350 x 350 x 480 cm, 6 echoes, TE1=0.9ms, and ∆TE=0.70ms. The strategy behind these two scans was to maintain the same B0 field since the true field map inhomogeneity is unknown. This was accomplished by keeping the pre-scans and shims constant. Changing the FOV with the back to back scans results in different CG phase shifts and will produce different estimates of the field maps from the raw data.

Next, the CG phase shifts were calculated numerically using the known gradient waveforms. Then the fat-fraction and R2* maps were calculated from the raw data and data with the calculated CG phase shifts removed. At 1.5T, field maps were estimated from both sets of raw data and after the corresponding CG phase shifts were removed.

Results

Fig. 2a demonstrates that two interleaved echo trains leads to a nonlinear accumulation of phase error, whereas the other sequences accumulate error linearly with respect to echo number. These simulations demonstrate that the fat-fraction in Fig. 2a and R2* in Fig. 2b are only affected with a two shot interleaved acquisition, but significant errors accumulate in the B0 field map estimate in Fig. 1d for all acquisition types. The phantom experiments demonstrate that CG correction also results in improved estimates of the fat-fraction (Figs. 3a and 4), and R2* (Fig. 3b) for this phantom. Fig. 3c also demonstrates that CGs distort the field map estimate differently depending on the FOV, but removal of CG phase shifts in the sequential scans results in similar field maps with only a small constant offset.

Discussion & Conclusions

Nonlinear phase errors that result from interleaved echo train acquisitions will corrupt the simultaneous estimation of fat-fraction and R2*. These changes can be relatively large at off-isocenter locations (eg. >10% fat-fraction as shown in Fig. 4). CG correction of experimental data demonstrates that the detrimental effects of CG phase shifts can be removed before reconstruction to produce more accurate estimates of the fat-fraction, R2*, and field map measurements.

Acknowledgements

The authors wish to acknowledge support form the NIH (UL1TR00427, R01 DK083380, R01 DK088925, R01 DK100651, and K24 DK102595), GE Healthcare, and the National Cancer Institute of the National Institutes of Health under Award Number T32CA009206.

References

[1] Bernstein M, Zhou X, Polzin J, King K, Ganin A, Pelc N, Glover G, Concomitant gradient terms in phase contrast MR: analysis and correction, Magn Reson Med 1998;39:300-8.

[2] Ruschke S, Eggers H, Kooijman H, Baum T, Settles M, Haase A, Rummeny E, Karampinos D, Addressing phase errors in quantitative water-fat imaging at 3 T using a time-interleaved multi-echo gradient-echo acquisition, In Proceedings of the 23rd Annual Meeting of ISMRM, Toronto, Canada, 2015. Abstract 3657.

[3] Hernando D, Kellman P, Haldar J, Liang Z, Robust water/fat separation in the presence of large field inhomogeneities using a graph cut algorithm, Magn Reson Med 2010;63:79-90.

Figures

Figure 1: Readout gradients for six-echo CSE gradients for the three simulated acquisition types. Echo times are marked with a black diamond.

Figure 2: The nonlinear CG phase shift as a function of echo number causes errors in all three estimated parameters for the two interleaved echo trains, whereas the linear phase errors only cause errors in the B0 field map.

Figure 3: Experimental correction of CG shifts improve the fat-fraction and R2* for a homogenous doped-water phantoms. Corrected fat-fractions are closer to 0% (left) and corrected R2* estimates are more homogeneous (middle), and acquisitions with different CGs result in similar CG corrected field maps (right). Plots are along the z-axis through isocenter.

Figure 4: Experimental correction for concomitant gradients reduces errors in the estimated fat-fraction when using interleaved echo train acquisitions. The fat-fraction estimate of the raw data (a) versus the fat-fraction estimate after concomitant gradient correction (b), with the SI direction shown on the vertical axis.



Proc. Intl. Soc. Mag. Reson. Med. 24 (2016)
0044