Junyu Guo1, Zoltan Patay1, and Wilbrun E. Reddick1
1St Jude Children's Research Hospital, Memphis, TN, United States
Synopsis
We present a novel,
simple and fast MR spectroscopic imaging technique and show its conceptual validation
with simulations and demonstrate proof-of-principle with phantom and human studies.
First, compared to the conventional spectroscopic imaging in the time-domain,
our method acquires data in the frequency domain, allowing flexible non-uniform
sampling to speed up the acquisition. Second, using ultra-small RF pulses
offers intrinsic water and fat suppression, greatly simplifying the scanning
procedures. Third, this new technique has hundreds of times lower energy deposition
than conventional MRI scans. We believe our method could allow spectroscopic
imaging to play a larger role in clinical applications.Purpose
MR
spectroscopic imaging (MRSI) is an important tool in clinical applications. As
the spectrum is measured in either the frequency or the time domain, MRS is
divided into the frequency-resolved and the time-resolved techniques (1, 2, 3).
The early frequency-resolved techniques have been supplanted by the
time-resolved technique with faster acquisition and higher sensitivity. We
introduce a fast frequency-resolved MRSI technique, termed phase-cycled
spectroscopic imaging (PCSI). PCSI uses an ultra-low flip-angle steady state to
achieve high acquisition efficiency and faster frequency sweeping by changing
cycled RF phase and using flexible non-uniform sampling, and greatly reduces RF
energy deposition. With its intrinsic water and fat suppression, performing PCSI
more closely resembles routine clinical scans by eliminating outer volume suppression
steps. We demonstrate its feasibility using simulations, phantom and human
studies.
Method
PCSI is the
first pulsed frequency-resolved MRSI method. Using a series of ultra-small RF
pulses to excite a target frequency, PCSI acquires an image at that frequency. By
sweeping through discrete frequency points using RF phase-cycling, PCSI
acquires a series of images at a range of frequencies and generates a spectrum
in each voxel. PCSI makes frequency sweeping more flexible, allowing
non-uniform frequency sampling of selected frequencies to speed up acquisition.
We
performed simulations with multiple small flip angles to validate its
feasibility and investigate the importance of the choice of flip angle. In simulations, a spectrum was created with
four peaks: water, Cho, Cr, and NAA, with magnitudes of 10000, 0.6, 0.8, and
1.2, respectively. The profiles of the transverse magnetization were computed for
different flip angles (0.24, 1, 3 degree) using TR=2.4 ms, T1=1300 ms, T2=250
ms.
All imaging studies were performed on a Siemens 3T scanner. PCSI acquisition uses a
modified bSSFP sequence with the cycled RF phase changing for each measurement.
A single axial slice was selected, and the advanced shimming mode was
used. The system frequency was decreased
by 200 Hz for phantom, 190 Hz for human. The protocol was as follows: ms; ms; flip angle for phantom, for human;
acquisition matrix 32×32; resolution 6.25 ×
6.25×15 mm3;
23 averages; 143 measurements with non-uniform sampling; total time 4:28
minutes.
For comparison, single-voxel
spectra were acquired using a single voxel spectroscopy (SVS) sequence with TE =135
ms, voxel size = 20×20×20 mm3, and the total time 4:24 minutes. Chemical shift
imaging (CSI) data were acquired from a patient with TE=135 ms; voxel size 10×10×15 mm3; acquisition matrix 10×10; and total time 8:07 minutes.
Results
Figure
1a shows the original generated spectrum. Fig. 1b shows the simulated spectrum
with an optimal a=0.24°. All metabolic peaks are observed
(Fig.2b), with heights proportional to the original values. Fig. 1c shows
metabolic peaks are barely detectable in the simulated spectrum with α=1°. Those peaks are not appreciable in
Fig. 1d, with α=3°.
Figure
2a shows PCSI signals of the phantom at the center (ROI size: 18.75×18.75×15 mm3). “Zooming in” inset
demonstrates three metabolite peaks on the real signal, which was converted to
the spectrum (Fig. 2b). For comparison, Fig. 2c shows a single-voxel spectrum
using SVS. PCSI and SVS spectra (Fig. 2b, 2c) were similar and had consistent
peak positions.
Figure
3a shows a PCSI spectrum from a volunteer (ROI size: 18.75×18.75×15 mm3), in which three
metabolite peaks can be identified and fitted. The corresponding SVS spectrum
is shown in Fig. 3b, with inset showing ROI’s location similar to that in Fig.
3a. Both spectra show consistent peak positions and relative peak heights after
alignment. Figure 4 shows metabolic parametric maps were generated for two
repeated measurements to demonstrate a good robustness of PCSI. Two sets of
maps were interpolated and overlaid on a T2-weighted image for comparison.
Figure
5 shows CSI parametric maps and PCSI maps from a patient with a multifocal
anaplastic astrocytoma. There were four lesion foci (1-4) shown in one T2w
image, in which lesion 1 corresponds to a surgical cavity and lesions 2-4
represent tumor foci. Fig. 5 shows the importance of full coverage of the whole
slice due to the heterogeneity of tumors.
Conclusion
We
demonstrated PCSI, a new flexible way, to perform MRSI in the frequency domain.
PCSI uses a new frequency-sweep technique and a non-uniform sampling scheme
that speed up acquisition. PCSI also offers intrinsic water and fat
suppression, reducing operator dependence and greatly simplifying scanning
procedures. Further, PCSI substantially reduce SAR using
ultra-small RF pulses. PCSI paves the way to faster, simpler, and broader
clinical applications of MRSI.
Acknowledgements
The authors wish to acknowledge the valuable
contributions of Angela Edwards, MR Technician, for her efforts in acquiring
and processing of the CSI data. We acknowledge the efforts of Sharon Naron for her
editing assistance. We also thanks Dr. Samuel Brady for letting us borrow the
phantom.References
1 Arnold, J. T., Dharmatti, S. S. & Packard, M. E. Chemical
Effects on Nuclear Induction Signals from Organic Compounds. J Chem Phys 19, 507-507, doi:Doi 10.1063/1.1748264 (1951).
2 Meyer, L. H., Saika, A. & Gutowsky, H. S. Electron
Distribution in Molecules .
3. The Proton Magnetic Spectra of Simple Organic
Groups. J Am Chem Soc 75, 4567-4573, doi:Doi
10.1021/Ja01114a053 (1953).
3 Ernst, R. R. & Anderson, W. A. Application of Fourier
Transform Spectroscopy to Magnetic Resonance. Rev Sci Instrum 37,
93-+, doi:Doi 10.1063/1.1719961 (1966).