Stephen Powell1,2,3, Stephanie Withey2,3,4, Yu Sun2,3,5, Lesley Macpherson6, Laurence Abernathy7, Barry Pizer8, Richard Grundy9, Simon Bailey10, Dipayan Mitra11, Dorothee Auer12, Shivaram Avula7, Theodoros N. Arvanitis2,3,13, and Andrew Peet2,3
1Physical Sciences for Health CDT, University of Birmingham, Birmingham, United Kingdom, 2Institute of Cancer and Genomic Sciences, University of Birmingham, Birmingham, United Kingdom, 3Department of Oncology, Birmingham Children's Hospital, Birmingham, United Kingdom, 4RRPPS, University Hospitals Birmingham NHS Foundation Trust, Birmingham, United Kingdom, 5School of Biological Sciences and Medical Engineering, Southeast University, Nanjing, China, 6Radiology, Birmingham Children's Hospital, Birmingham, United Kingdom, 7Radiology, Alder Hey Children's NHS Foundation Trust, Liverpool, United Kingdom, 8Oncology, Alder Hey Children's NHS Foundation Trust, Li, United Kingdom, 9The Children's Brain Tumour Research Centre, University Of Nottingham, Nottingham, United Kingdom, 10Sir James Spence Institute of Child Health, Royal Victoria Infirmary, Newcastle, United Kingdom, 11Neuroradiology, The Newcastle upon Tyne Hospitals NHS Foundation Trust, Newcastle, United Kingdom, 12Sir Peter Mansfield Imaging Centre, University Of Nottingham, Nottingham, United Kingdom, 13Institute of Digital Healthcare, University of Warwick, Coventry, United Kingdom
Synopsis
Dynamic
Susceptibility Contrast (DSC-) MRI estimates biomarkers, such as cerebral blood
volume (CBV). However, data quality varies between centres and quality control
(QC) is carried out by qualitative
review, which is time-consuming and
subjective. An automated QC pipeline was developed and tested on 34 patient
data sets. The pipeline analysed four slices from each patient, producing SNR,
RMSE, relative CBV (rCBV), and quality maps for each slice, which were used to
quantify QC. Average values for each
parameter were produced for each centre, protocol and field strength, showing
variability in data quality and providing a basis for multi-centre protocol optimisation.
Introduction
Dynamic
Susceptibility Contrast (DSC-) MRI uses a contrast agent to estimate perfusion
biomarkers, such as cerebral blood volume (CBV). The contrast agent causes a
decrease in signal intensity as it passes through the brain. Each pixel has a
time course, which shows changes in signal intensity1. The time
course is susceptible to changes in protocol, field strength and artefacts such
as motion and susceptibility; affecting biomarker accuracy. Different centres
use different scanners and protocols, causing variability in data quality; making
it difficult to obtain robust biomarker estimates. Some centres use a pre-bolus, which is
important in low grade tumours, as leakage correction can result in T1 effects2.
The current quality control (QC) method is based on expert qualitative review.
However, this can be subjective and time-consuming, particularly in multicentre
studies. This work focuses on developing an automated QC pipeline, which aims to
provide a quantitative measure of QC.Methods
The
QC pipeline was developed in Matlab 9.4. Anonymised
scans from patients with cerebellar tumours
were obtained from the CCLG functional
imaging of tumours database, which
contains patient scans from multiple centres with differing MR protocols. 34
patients were analysed, from four
centres. Four consecutive slices, including the corpus callosum, were manually
chosen from each data set. For data sets that didn’t include the corpus
callosum in the field of view, the top four slices were used. Figure 1 shows
the QC pipeline. Ventricle segmentation differed between the two protocols, due
to reduced contrast in sPRESTO data. Pipeline outputs included parametric maps
of signal-to-noise-ratio (SNR), root-mean-square error (RMSE), rCBV and a quality
map.Results
The
automated QC code segmented and created parametric maps for all 34 data sets.
Figure 2 shows example maps produced by the algorithm for one slice. Figure 3
shows quality maps overlaid onto brain, for data acquired at different field
strengths. Table 1 shows quality parameters and rCBV values, grouped by centre
and protocol.Discussion
Figure 1 shows larger RMSE and lower SNR in white matter
(WM) than grey matter (GM), which is expected as WM is less perfused; therefore
it is more affected by noise. From table 1, the worst-performing data is 1.5T
data from centre 1, suggesting that 1.5T acquisition with pre-bolus results in
poor SNR. When a pre-bolus was used, the total dose was split between the
pre-bolus dose and the main dose, resulting in a reduced main dose. The
best-performing data is from centre 3, giving the largest SNR and highest
percentage of pixels passing. Centre 3 is the only 3T GE-EPI data that does not
have any pre-bolus, boosting SNR, and the percentage of pixels passing. The
lack of pre-bolus may explain why sPRESTO equals or outperforms GE-EPI in
centres 1 and 4. Although sPRESTO has high SNR in these cases, this comes at
the cost of spatial resolution and image contrast. Figure 4 shows a comparison
of sPRESTO and GE-EPI data, showing
sPRESTO gives poor contrast between GM, WM and ventricles, and that the spatial
resolution of sPRESTO datasets is poor. The sPRESTO sequence is also reported
to have poorer temporal resolution than the EPI sequence7.
SNR from
table 1 shows a correlation with rCBV, as larger SNR gives a larger signal drop.
There is also an inverse correlation between SNR and RMSE. TR, TE, flip angle
and voxel volume appears to have less effect on data quality. Each centre gives
similar average coverage, other than centre 3, as the field of view is focussed
on the cerebellum. Although it has better SNR the whole brain isn’t in the
field of view, meaning the whole tumour may not be covered.
Normalised
rCBV is smaller for sPRESTO data sets, showing greater similarity between rCBV
values of passed and failed pixels. Standard deviations in rCBV for failed pixels
are large compared to average rCBV showing the importance of good SNR and how
it affects biomarker accuracy. The standard deviation for passed pixels is
large, due to the inclusion of GM and WM.
Conclusion
This
work demonstrates an automated QC pipeline, tested on healthy slices of brain
acquired using a range of different protocols. The results show that 1.5T
acquisition with a pre-bolus results in too many pixels being discarded. It
shows the importance of good SNR in achieving robust biomarkers. It also shows
some of the compromises involved in DSC-MRI acquisition. For example,
choosing between GE-EPI and sPRESTO means compromising either on signal quality
or image contrast, whilst splitting the Gd bolus reduces SNR. These factors
should be taken into account when designing protocols for multi-centre use. Acknowledgements
This work was funded by EPSRC through a studentship from
the Sci-Phy-4-Health CDT (EP/L016346/1) and the National Institute for Health Research (NIHR) via a research
professorship (13-0053) and Help Harry Help Others Cure.References
[1] Shiroishi, Mark S., et al. "Principles of
T2*‐weighted dynamic susceptibility contrast MRI technique in brain tumor
imaging." Journal of Magnetic Resonance Imaging 41.2
(2015): 296-313.
[2] Withey, Stephanie., et al. “Multicentre pediatric
brain tumour dynamic susceptibility contrast (DSC-) MRI with contrast agent
leakage correction.” Proceedings of the
Joint Annual Meeting ISMRM-ESMRMB 2018 June 16-21, Paris, Abstract No. 5418
[3] MacQueen, James. "Some methods for
classification and analysis of multivariate observations." Proceedings
of the fifth Berkeley symposium on mathematical statistics and probability.
Vol. 1. No. 14. 1967.
[4] Patel, Sakshi, Shivani Sihmar, and Aman Jatain.
"A study of hierarchical clustering algorithms." Computing
for Sustainable Global Development (INDIACom), 2015 2nd International
Conference on. IEEE, 2015.
[5] Pandit, Shraddha, and Suchita Gupta. "A
comparative study on distance measuring approaches for clustering." International
Journal of Research in Computer Science 2.1 (2011): 29-31.
[6] Kao, Yi‐Hsuan, et al. "Removal of CSF pixels on
brain MR perfusion images using first several images and Otsu's thresholding
technique." Magnetic resonance in medicine 64.3 (2010):
743-748.
[7] van Gelderen, Peter, et al. "The PRESTO
technique for fMRI." NeuroImage 62.2 (2012): 676-681.