N. Jon Shah1,2,3,4 and Seong Dae Yun1
1Institute of Neuroscience and Medicine 4, INM-4, Forschungszentrum Juelich, Juelich, Germany, 2Institute of Neuroscience and Medicine 11, INM-11, JARA, Forschungszentrum Juelich, Juelich, Germany, 3JARA - BRAIN - Translational Medicine, Aachen, Germany, 4Department of Neurology, RWTH Aachen University, Aachen, Germany
Synopsis
Due
to their relative insensitivity to B1 inhomogeneities, Look-Locker
methods are widely used for the quantification of T1 relaxation time.
One such Look-Locker method, TAPIR, has been demonstrated with several clinical
applications and has been shown to be faster than conventional gradient-echo
sequences. However, it still requires a considerable acquisition time for
whole-brain imaging. To overcome this limitation, a much faster method, ir-EPIK,
has been presented in our earlier work. This work aims to perform a quantitative
assessment of ir-EPIK in comparison to TAPIR using phantom data and twenty data
sets from subjects. All data were acquired at 3T.
Introduction
The
quantification of T1 relaxation time has been performed using
various methods such as variable flip angle or Look-Locker approaches.1-4
Compared to the variable flip angle scheme, the Look-Locker methods are more
robust against B1 inhomogeneity and consequently do not require
additional efforts for the correction of inhomogeneity effects.5 TAPIR
(T1-mApping-with-Partial-Inversion-Recovery) is one such Look-Locker approach
and has been demonstrated with several clinical applications.6,7 Although
the method is faster than conventional gradient-echo sequences, it still
requires a considerable acquisition time for whole-brain imaging. To overcome
this limitation, a much faster Look-Locker method, inversion recovery EPI with
keyhole (ir-EPIK), was presented in our earlier work.3,8-11 This
work aims to perform a quantitative assessment of ir-EPIK in comparison to
TAPIR using phantom data and data sets from 20 subjects. All data were gathered
at 3T. Here, the T1 difference between the male and female groups is
also considered.Methods
Figure 1 depicts the sequence diagram of TAPIR and the
corresponding schematic plot of the T1 recovery curve showing how
the temporal points are sampled. TAPIR is essentially based on multi-slice
FLASH imaging, but here its scheme was modified to sample multiple phase
encoding lines (e.g. 3) per excitation in the same way as an EPI readout. As
shown in Fig. 1b, TAPIR acquires multi-slice data within the same T1
recovery plot, which is repeated as many times as the number of required phase
encoding set. Figure 2 illustrates the case for ir-EPIK. In contrast to TAPIR,
it only acquires single-slice data per each T1 recovery plot, which
is repeated as many times as the number of slices. Phantom data were acquired with
the following imaging conditions: FOV = 240 × 240 mm2, matrix = 96 × 96 (in-plane
resolution: 2.5 ×
2.5 mm2), number of slices = 1 with 2.5 mm thickness, TR/TE/TI =
60/25/20 ms, FA = 25°
and number of time points = 63. Here, only a single slice was measured to keep
the same condition for both ir-EPIK and TAPIR. It is important to note that
ir-EPIK and TAPIR acquire single-slice and multi-slice data per each single T1
recovery plot, respectively. The same imaging parameters were applied to the
measurement of twenty subjects (10 males, 10 females; mean ± SD age, 28.20 ±
4.99 years), except that the number of slices was 49 for ir-EPIK. For TAPIR, only
seven slices were acquired with the shortest possible TE/TR (125.3/7.67 ms) to keep
the total acquisition time as short as possible. The number of time points in
TAPIR was also changed to 30 for the in vivo
measurements so as to maintain the similar total T1 recovery time
(3.76 s) as ir-EPIK (3.78 s). The above configuration was employed on a Siemens
Magnetom Trio 3T MRI scanner.Results
Figure 3 shows obtained T1 maps. For the in vivo data, all slices (7) are
presented for TAPIR, whilst only 7 slices at the identical slice positions, out
of 49 slices, are presented for ir-EPIK. For the phantom data, five ROIs were
defined and the corresponding computed mean ± SD T1 values are
listed in Fig. 4a. The same analysis was applied to the in vivo data but only two ROIs (WM and GM) were used. Here, the mean
± SD T1 over 20 subjects is listed in Fig. 4b. The phantom results
suggest that the T1 values from both methods are quite comparable to
each other. The in vivo results
reveal that the estimated T1 values from ir-EPIK were slightly higher
than those from TAPIR; they were around 6.84% and 3.67% higher for the WM and
GM regions, respectively. Figure 4c shows two-sample t-test results between the
male and female groups, where relatively high p-values were obtained. This
indicates that the T1 difference between the two groups was not
significant. Figure 5 shows the T1 maps for the entire slices (49)
from ir-EPIK, presenting that reliable T1 estimation was performed
for all slices.Discussion and conclusions
This
work demonstrates rapid, whole-brain T1 mapping using ir-EPIK. The
method acquired the entire T1 recovery data (2.5 mm2
resolution × 49 slices × 63 time-points) within 5 minutes. The estimated T1
values from ir-EPIK were comparable to those from TAPIR, but were found to be
slightly higher for the in vivo data.
This was mainly due to the use of a different TE for the measurement of in vivo data. A previous work reported
the dependence of the estimated T1 values on the TE employed; T1
increases by nearly 100ms between TE = 9.5 and TE = 25ms.12Acknowledgements
No acknowledgement found.References
- Heule R, Ganter C, Bieri O. Variable flip angle T1 mapping in the human brain with reduced T2 sensitivity using fast radiofrequency-spoiled gradient echo imaging. Magn Reson Med. 2016 Apr;75(4):1413-22.
- Shah NJ, Zaitsev M, Steinhoff S, Zilles K. A new method for fast multislice T1 mapping. Neuroimage. 2001 Nov;14(5):1175-85.
- Shah NJ, Yun S. A rapid, whole-brain look-locker method for T1 mapping using inversion recovery EPIK. In Proceedings of the 25th Annual Meeting of ISMRM, Honolulu, USA, 2017. Abstract 3719.
- Wright PJ, Mougin OE, Totman JJ, Peters AM, Brookes MJ, Coxon R, Morris PE, Clemence M, Francis ST, Bowtell RW, Gowland PA. Water proton T1 measurements in brain tissue at 7, 3, and 1.5 T using IR-EPI, IR-TSE, and MPRAGE: results and optimization. MAGMA. 2008 Mar;21(1-2):121-30.
- Li W, Griswold M, Yu X. Rapid T1 mapping of mouse myocardium with saturation recovery Look-Locker method. Magn Reson Med. 2010;64(5):1296-1303.
- Shah NJ, Neeb H, Kircheis G, et al. Quantitative cerebral water content mapping in hepatic encephalopathy. Neuroimage. 2008;41(3):706-717.
- Shah NJ, Neeb H, Zaitsev M, Steinhoff S, Kircheis G, Amunts K, Häussinger D, Zilles K. Quantitative T1 mapping of hepatic encephalopathy using magnetic resonance imaging. Hepatology. 2003 Nov;38(5):1219-26.
- Zaitsev M, Zilles K, Shah NJ. Shared k-space echo planar imaging with keyhole. Magn Reson Med. 2001;45(1):109-117.
- Zaitsev M, D'Arcy J, Collins DJ, Leach MO, Zilles K, Shah NJ. Dual-contrast echo planar imaging with keyhole: application to dynamic contrast-enhanced perfusion studies. Phys Med Biol. 2005 Oct 7;50(19):4491-505.
- Yun S, Reske M, Vahedipour K, et al. Parallel imaging acceleration of EPIK for reduced image distortions in fMRI. NeuroImage. 2013;73:135-143.
- Yun S, Shah NJ. Whole-brain high in-plane resolution fMRI using accelerated EPIK for enhanced characterisation of functional areas at 3T. PLoS One. 2017;12(9):e0184759.
- Oros-Peusquens AM, Keil F, Shah NJ. Fast two-dimensional relaxometric characterisation of brain tissue in vivo: is T1 relaxation multi-component? In: Proc. Intl. Soc. Mag. Reson. Med. 18 (2010) 4988.