Yiming Wang1, Joshua S. Greer1,2, Trevor Wigal1, Marco C. Pinho1,3, Joseph A. Maldjian1,3, and Ananth J. Madhuranthakam1,3
1Radiology, UT Southwestern Medical Center, Dallas, TX, United States, 2Pediatrics, UT Southwestern Medical Center, Dallas, TX, United States, 3Advanced Imaging Research Center, UT Southwestern Medical Center, Dallas, TX, United States
Synopsis
3D fast/turbo spin echo (FSE/TSE)
acquisitions are preferred for arterial spin labeled (ASL) MRI due to their
higher SNR and compatibility with background suppression. However, 3D TSE
suffers from T2 blurring caused by the T2 decay of the
ASL signal along the prolonged echo train lengths, which may degrade image
quality. This is often more noticeable in 3D TSE with Cartesian acquisitions.
In this study, a truncated k-space filter is designed to compensate the T2
blurring of 3D TSE with Cartesian acquisitions and improve sharpness of ASL
brain perfusion images.
Introduction
3D fast/turbo spin echo (FSE/TSE)
acquisitions are preferred for arterial spin labeled (ASL) MRI due to their
higher signal to noise ratio (SNR) and compatibility with highly efficient
background suppression (BGS). Although, 3D ASL using stack-of-stars (SOS) or
GRASE has been recommended by the ISMRM perfusion study group for brain
perfusion imaging, these acquisitions are challenging in areas with increased B0
inhomogeneities (e.g. skull base). 3D TSE with Cartesian acquisitions have been
shown to be robust in such applications [1]. However, the T2 decay due
to prolonged echo train lengths (ETLs) of 3D TSE broadens local point spread
functions, and leads to image blurring. Nevertheless, T2 blurring
can be addressed through multiplicative k-space filtering [2], which is well-suited for ASL applications, since the majority of the signal is from a
single tissue component. In this study, such a truncated k-space filter was designed
to reduce image blurring while minimizing SNR loss in 3D ASL images of the
brain. Methods
Pseudo-continuous ASL (pCASL) with optimized background suppression
was used to acquire brain perfusion images [1]. With optimized background
suppression, the signals of the perfusion-weighted images are mostly from the labeled
blood, and thus the blood T2 of 275 ms (at 3T) was used for the
filter design [3]. A 3D TSE Cartesian Acquisition using Spiral Profile
Reordering (CASPR) was used for data acquisition (Fig. 1). Due to sampling of
the center of k-space at the beginning of each echo train, CASPR has been shown
to improve the robustness of ASL signal preparation similar to a spiral
trajectory, but still maintaining acquisition on a Cartesian grid, thus
increasing its robustness to B0 inhomogeneities.
The
truncated k-space filter was designed to compensate T2 blurring, such
that, \begin{equation}
I(n) =\ \left\{
\begin{array}{**lr**}
e^{\left[(n-1) \cdot \frac{\Delta t }{T_2} \right]
}, &n<n_c &\\
\frac{e^{ \left[(n_c-1)\cdot \frac{\Delta t }{T_2} \right] }-1}{e^{\left[k\cdot(n-n_c)-p\right]}+1}+1, &
otherwise\\
\end{array} , \right.
\end{equation} where Δt is the echo
spacing, n is the echo index within
an echo train, nc is the
cut-off echo index where the truncation begins, k, p are two constants
related to the truncation shape. The filter intensity I increases exponentially with n
when n < nc, and progressively decreases to 1 when n
exceeds nc, following the
Fermi-Dirac function (Fig. 2). This filter compensates the T2 signal
decay in the central k-space (up to nc), without amplifying the later echoes at the higher spatial frequencies,
potentially avoiding significant SNR loss.
The
filter was applied to 3D ASL images of the brain in 7 volunteers, acquired
using 3D pCASL with CASPR view ordering at 3T. The imaging parameters were:
TR/TE = 6800/12 ms, FOV = 200x200x155
mm3, matrix = 68x67 with 77 slices, acquired resolution =
3x3x3 mm3, reconstructed resolution = 0.69x0.69x1.50 mm3, label duration = 1.8 s, post-label delay = 1.8 s, 4 BGS pulses and total
acquisition time = 5.5 minutes. The
Just Noticeable Blur (JNB) metric, a no-reference objective image sharpness
metric, was used to evaluate the 3D pCASL image sharpness [4]. Results
The k, p were found to be optimal at 0.3 and 7 considering the filter
shape, the ETL, and the echo spacing used. The cut-off echo index, nc was found to be optimal at
66 and about 75% of the total echoes were multiplied with filter intensities
greater than 1. This optimal combination improved image sharpness, without
increasing the background noise (Fig. 3). The axial reformats from a sagittal
acquisition showed improved image sharpness across the entire brain (Fig. 4). The
JNB metric showed improvement across
all slices (Fig. 5a), with an average improvement of 17% across all subjects
(Fig. 5b). Discussion and Conclusion
The truncated k-space
filter designed in this study can be used to retrospectively improve image
sharpness of 3D pCASL images. While the improvements are shown in brain
perfusion images acquired with 3D CASPR, this retrospective filter can also be
applied to other 3D TSE acquisitions as well as to other anatomies such as
kidneys. Future studies will consider optimizing the refocusing flip angles
along with k-space filtering for prospective acquisitions [5]. Acknowledgements
This work was supported by the NIH/NCI (grant U01CA207091).References
[1] Greer, JS et al.
ISMRM 2017:3628.
[2] Zhou, X et al. JMRI
1993; 3(5):803-7.
[3] Stanisz, GJ et al.
MRM 2005; 54:507-512.
[4] Ferzli, R et al. IEEE
TIP 2009; 18(4):717-728.
[5] Zhao, L et al. MRM 2018; 80(4):1391-1401.