Christopher W Roy1, Jerome Yerly1,2, Jessica AM Bastiaansen1, Nemanja Masala1, Lorenzo Di Sopra1, Jens Wetzl3, Christoph Forman3, Davide Piccini1,4,5, and Matthias Stuber1,2
1Radiology, Lausanne University Hospital (CHUV) and University of Lausanne (UNIL), Lausanne, Switzerland, 2Center for Biomedical Imaging (CIBM), Lausanne, Switzerland, 3Siemens Healthcare GmbH, Erlangen, Germany, 4Advanced Clinical Imaging Technology (ACIT), Siemens Healthcare AG, Lausanne, Switzerland, 5LTS5, École Polytechnique Fédérale de Lausanne (EPFL), Lausanne, Switzerland
Synopsis
Recent
advances have enabled high resolution cardiac imaging using continuous
acquisitions that do not require external gating devices and can be
reconstructed in arbitrary dimensions. Here, we extend the use of this
Free-running framework to a fully self-gated free-breathing 3D Cartesian
trajectory with spiral profile ordering for cardiac and respiratory motion
resolved 5D imaging. We demonstrate the feasibility of this Cartesian approach
by reconstructing and comparing images from both radial and Cartesian sequences
with matching scan parameters in healthy volunteers. Overall, Cartesian images
demonstrated comparable cardiac and respiratory motion albeit with more
residual artifacts present in the Cartesian images.
Introduction
Whole-heart
MRI requires acquisitions and reconstructions that are robust to cardiac and
respiratory motion. In conventional examinations, a combination of ECG-gating
and breath-holding or respiratory navigation is used to freeze physiological
motion and produce sharp images. However, external gating devices add to
patient setup time, and respiratory control can be difficult in some patients.
Recent advances have enabled high-resolution imaging using continuous radial
phyllotaxis acquisitions that do not require external devices and can be
reconstructed exploiting the intrinsic similarities of arbitrary dimensions
(Free-running framework) [1-6]. This is accomplished using a repeated readout
along the superior-inferior (SI) direction for retrospective extraction of
respiratory and cardiac signals, which informs the compressed sensing
reconstruction to suppress undersampling artifact. In this work, we investigate
the feasibility of applying the Free-running framework to a Cartesian
trajectory with spiral profile reordering for cardiac and respiratory resolved
5D imaging [7-9]. Cartesian sampling is a
potentially useful alternative to radial as it is less sensitive to trajectory
errors, the imaging volume can more easily be tailored to specific applications
allowing for a trade-off between scan time, contrast, and resolution, and the
reconstruction can be performed without computationally demanding re-gridding.
We demonstrate the feasibility of Free-running Cartesian 5D imaging in healthy
volunteers and both qualitatively and quantitatively compare image
reconstructions of radial and Cartesian acquisitions.Methods
Data
were acquired in 8 healthy volunteers (5 male, ages 26-33 years) on a 1.5T
clinical MRI scanner (MAGNETOM Aera, Siemens Healthcare, Erlangen, Germany).
Two prototype slab-selective 3D bSSFP sequences with radial (Fig. 1b) and
Cartesian (Fib. 1c) sampling were acquired using golden angle phyllotaxis [6] and spiral profile
reordering [8], respectively. To match the self-navigation and flexible binning
of the radial acquisition, the Cartesian sequence was modified to acquire data
independent of the ECG, and one readout oriented along the SI direction was
repeated per interleave for subsequent extraction of physiological signals. All
imaging parameters, listed in Fig. 1, were consistent between the two sequences
except for the field-of-view which was isotropic (220 mm3) for the
radial sequence and rectangular for Cartesian (220 x 380 x 165 mm3).
Additionally, the RF excitation angle was lowered in some volunteers due to SAR
limitations. Data was retrospectively sorted into 16-25 cardiac phases of 50ms
window-width and 4 respiratory motion states [3]. The resulting 5D datasets
were reconstructed offline using a k-t SPARSE SENSE algorithm with
regularization along both temporal dimensions [1, 3]. Qualitative comparison of
images was performed by visual inspection across all reconstructed slices and
temporal dimensions. Quantitative comparison of contrasts measurements was performed
using the ratio of manually selected regions of interest containing the blood
pool and myocardium. Measurements of image sharpness were performed using the
signal gradient along the blood-myocardium, heart-lung, and heart-liver
boundaries [10]. All measurements were taken during end-systole at
end-expiration and statistical analysis was performed using paired t-tests.Results
Cardiac
and respiratory signals were successfully extracted (Fig. 2) from all 8 volunteers
and used for reconstructions of radial and Cartesian acquisitions. In general,
Cartesian reconstructions took less time (3-5 hours) using a CPU Fourier
transform than their radial counterparts (5-8 hours) which used a GPU
non-uniform Fourier Transform. Figs. 3 and 4 show two example reconstructions.
Visual comparison of radial and Cartesian images shows that dynamic cardiac and
respiratory motion is correctly resolved. However, Cartesian images contain
noticeably more residual artifact, leading to an apparent decrease in contrast
between the myocardium and blood pool, which was confirmed to be statistically
higher (p<0.05) for radial images (Fig. 5a). Quantitative comparison of
sharpness also yielded statistically higher (p<0.05) measurements along the
blood-myocardium and heart-lung boundaries for radial images, but there was no
significant difference in sharpness at the heart-liver boundary.Discussion and Conclusions
5D imaging using a
Free-running framework is a promising technique for easily assessing the
structure and function of the whole-heart in one single scan. It can be
performed without ECG during free-breathing and without extensive image
planning relative to the standard clinical 2D CINE approach. Here, we show that
Cartesian sampling may be used as an alternative to radial phyllotaxis sampling
within the Free-running framework. This study is to our knowledge, the first
comparison of 5D imaging between radial and Cartesian acquisitions. To provide
a baseline comparison, acquisition and reconstruction parameters were nearly
identical. Overall the Cartesian sequence produced similar images to the radial
sequence. Still, additional optimization of the Cartesian approach is required
to improve image quality. In particular, the use of ESPIRiT for coil
sensitivities may help reduce residual fold-over artifacts in the Cartesian 5D
images [11].Acknowledgements
No acknowledgement found.References
[1] S. Coppo, D. Piccini, G. Bonanno, J. Chaptinel, G. Vincenti, H. Feliciano, R.B. van Heeswijk, J. Schwitter, M. Stuber. Free-Running 4D Whole-Heart Self-Navigated Golden Angle MRI: Initial Results. Magn. Reson. Med. 2015; 74:1306–1316.
[2] L. Feng, L. Axel, H. Chandarana, K. T. Block, D. K. Sodickson, and R. Otazo, “XD-GRASP: Golden-angle radial MRI with reconstruction of extra motion-state dimensions using compressed sensing,” Magn. Reson. Med. 2016; 75:775–788
[3] L. Feng, S. Coppo, D. Piccini, J. Yerly, R.P. Lim, P.G. Masci, M. Stuber, D.K. Sodickson, R. Otazo. 5D whole-heart sparse MRI. Magn. Reson. Med. 2018; 79:826–838
[4] J. Yerly, D. Piccini, L. Di Sopra, J. AM Bastiannsen, S. Coppo, and M. Stuber “Improved Time Efficiency and Workflow for Fully Self-Gated Non-Contrast 5D Imaging of the Heart,” in Proc. Intl. Soc. Mag. Reson. Med. 2017, 2874
[5] L. Di Sopra, D. Piccini, S. Coppo, J. A.M. Bastiaansen, M. Stuber, and J. Yerly, “Motion-resolved 5d imaging of the heart: time to get rid of the ECG?,” in Proc. Intl. Soc. Mag. Reson. Med. 2017, 3148
[6] D. Piccini, A. Littmann, S. Nielles-Vallespin, and M. O. Zenge, “Spiral phyllotaxis: the natural way to construct a 3D radial trajectory in MRI.,” Magn. Reson. Med. 2011; 66: 1049–56
[7] C. Forman, D. Piccini, R. Grimm, J. Hutter, J. Hornegger, and M. O. Zenge, “High-resolution 3D whole-heart coronary MRA: a study on the combination of data acquisition in multiple breath-holds and 1D residual respiratory motion compensation,” Magn. Reson. Mater. Physics, Biol. Med. 2014 27; 435–443
[8] J. Wetzl, F. Lugauer, M. Schmidt, A. Maier, J. Hornegger, and C. Forman, “Free-Breathing, Self-Navigated Isotropic 3-D CINE Imaging of the Whole Heart Using Cartesian Sampling,” Proc. Intl. Soc. Mag. Reson. Med. 2016, 411
[9] T. Correia, G. Ginami, G. Cruz, R. Neji, R. Rashid, R.M. Botnar, C. Prieto. Optimized respiratory-resolved motion-compensated 3D Cartesian coronary MR angiography. Magn Reson Med. 2018;80(6):2618–2629
[10] R. Ahmad, Y. Ding, O.P. Simonetti. Edge sharpness assessment by parametric modeling: Application to magnetic resonance imaging. Concepts in Magnetic Resonance. 2015; 44:138–149.
[11] M. Uecker et al., “ESPIRiT--an eigenvalue approach to autocalibrating parallel MRI: where SENSE meets GRAPPA.,” Magn. Reson. Med. 2014; 71:990–1001