Kosuke Morita1, Masami Yoneyama2, Masanobu Nakamura2, Takeshi Nakaura3, Seitaro Oda3, Masahiro Hatemura1, and Yasuyuki Yamashita3
1Radiology, Kumamoto University, Kumamoto-shi, Japan, 2Philips Japan, Tokyo, Japan, 3Diagnostic Radiology, Kumamoto University, Kumamoto-shi, Japan
Synopsis
The purpose of our
study was to improve 3D-STIR for TRANCE non-contrast MR angiography in clinical
3.0T MR system using modified hyperbolic secant (HS4) pulses. the higher
field strength poses additional challenges to 3D STIR, including wider offset
frequency between water and fat combined with larger B0 and B1
inhomogeneities, which reduce the reliability of fat suppression. 3D STIR (TRANCE) with HS4 pulse has clearly improved fat suppression due
to B0/B1 inhomogeneous compared with conventional HS
pulse in clinical.
INTRODUCTION
ECG-Triggered
Non-Contrast-Enhanced MR Angiography (TRANCE1) typically applies 3D
short tau inversion recovery (STIR) sequence to effectively suppress background
fat signals. 3D STIR basically applies nonselective adiabatic IR pulses such as
hyperbolic secant (HS). In fact, the higher field strength poses additional
challenges to 3D STIR, including wider offset frequency between water and fat
combined with larger B0 and B1 inhomogeneities, which
reduce the reliability of fat suppression. The crucial factors of an adiabatic
IR pulse to achieve robust fat suppression in 3D TSE-STIR are a broader
bandwidth for B0 insensitivity and a lower adiabatic threshold for B1
insensitivity, which are challenging to achieve simultaneously with the HS
pulse2 [Fig.1]. To improve the reliability of HS pulse, several
alternatives including a derivative of the HS pulse, called a stretched HS
pulse, have been proposed for T1rho studies3,4. Furthermore, broadband
STIR pulse with increased robustness to B0 and B1
inhomogeneities at 3T for brachial plexus imaging has also been demonstrated
recently2. Therefore, we attempt to use stretched HS pulse for 3D
STIR on 3.0T. The purpose of our study was to improve fat suppression in TRANCE
on clinical 3.0T MR using stretched HS pulses.
MATERIALS AND METHODS
Experimental
data was collected from 5 healthy volunteers. Written informed consent was
obtained from each volunteer and the protocol was approved by the ethics
committee. All studies were performed with clinical 3.0T MR scanner (Philips, Ingenia
3.0T CX) and 32-channel dS-torso coil. Scan parameters of 3D STIR using HS
pulses were as follows: TR / TE / TI = 2000 / 100 / 220 ms, slices thickness = 2.0
mm, number of slices = 50, field-of-view = 40 × 40 cm2, acquisition
matrix = 320 × 320 (reconstruction matrix = 512× 512), NSA = 1, TSE factor =
30, echo space = 5.4 ms, Compressed Sense (CS) factor = 8.0, Acquisition time =
4:40, Coronal plane acquisition. Mean measured
values were measured in the regions of interest (ROI) on 3D STIR images with
default IR pulse (conventional HS) and stretched HS pulse (broadband HS). We
determined the HSn pulse using Eq.(1). The amplitude and
orientation are time dependent and these in turn are functions of the amplitude
modulations and frequency modulations used for the pulse. One way to alter the
modulations is to change the stretching factor, n. Accordingly, the amplitude modulation function of the HSn pulse becomes flatter as the factor n increases as given by
ω1(t) = ω1maxsec h(βτn) … (1)
where ω1max is the maximum pulse
amplitude in rad per sec and β is a constant. When n=1, the pulse is the original hyperbolic secant pulse5,
whereas when n > 1, a stretched
version of the pulse is produced6. In the experiments performed
here, HSn pulse parameters was held
constant except for the stretching factor (n
=4). Fig.2 shows actual
RF perfomance for conventional HS and broadband HS4 pulse. RESULTS
Figure 3 shows the representative
images for upper extremity and lower extremity. Fat suppression failure due to B1
inhomogeneous appeared on conventional HS images (arrows). In
contrast, broadband HS4 uniformly suppressed background fat sugnals. Figure 4 shows
clinical case for elbow soft tissue tumor. TRANCE MIP image demonstrated more
detailed blood vessel information around the tumor with clear fat suppression.
CONCLUSION
3D STIR (TRANCE) with HS4
pulse has clearly improved fat suppression due to B0/B1
inhomogeneous compared with conventional HS pulse on 3.0T MR system. It may facilitate the detection of small vessel lesion existing in the
extremities even with B0/B1 inhomogeneity.
Acknowledgements
No acknowledgement found.References
[1] Gutzeit a, et al.
Eur Radiol 21:1979–1987, 2011.
[2] Wang X, et al. J
Magn Reson Imaging 48:1104-1111, 2018.
[3] Jokivarsi KT, et al., Journal of Cerebral Blood Flow & Metabolism
29:206-216; 2009.
[4] Okuaki T et al., Proc. Intl. Soc. Mag.
Reson. Med. 21:4093, 2013.
[5]
Silver MS, et al. Magn Reson Med 29:688-691; 1993.
[6] Tannus A et al. J Magn Reson A
120:133-137; 1996.