Christoph Stefan Aigner1, Armin Rund2, Christina Graf1, Karl Kunisch2, and Rudolf Stollberger1
1Institute of Medical Engineering, Graz University of Technology, Graz, Austria, 2Institute for Mathematics and Scientific Computing, University of Graz, Graz, Austria
Synopsis
This work demonstrates a constrained joint
design of minimum duration RF pulse and slice selective gradient waveforms for combined
SMS excitation, refocusing and inversion scenarios. A hybrid trust-region
semismooth Newton/quasi-Newton method with exact derivatives via adjoint
calculus is used to solve the time optimal problem on fine spatial and temporal
grids. Specific hardware and safety constraints, including maximal RF, slice
selective gradient, slew rate amplitudes as well as global SAR estimates,
guarantee practical applicability. High-resolution GRE, crushed SE and
inversion recovery GRE slice profile measurements on a 3T MR system validate
the numerical results.
Target audience:
Researchers working in the field of RF pulse
design and SMS imaging.Purpose:
Simultaneous Multislice (SMS) imaging is often
limited by long RF pulse durations which result from hardware limitations and
the demand for many accurate slices1. The overall pulse duration and
RF energy requirements can be addressed with various methods1,2,3
including time optimal joint design of RF pulse and slice selective gradient
shape4,5. Since the resulting Gs amplitudes are allowed to vary
heavily, a refined gradient system model based on the gradient impulse response
function was introduced to correct for potential gradient system imperfections6.
In this work, we extend the optimization framework5 to the design of
combined time optimal SMS excitation, refocusing and inversion scenarios.Theory:
The
time optimal control framework for the constraint joint design of RF and slice
selective gradient shape is formulated in the Bloch spin domain in terms of Cayley-Klein parameters for different RF pulse types7 including
excitation $$$(2ab^*)$$$, inversion $$$(1-2|b|^2)$$$ and crushed refocusing
$$$(|b|^2)$$$. In addition to the magnitude slice profile evaluation, the phase
angle of the slice profile can be added for excitation or refocusing scenarios.
Maximal amplitude values for RF and Gs slew rate are included as pointwise hard
constraints. The time optimal problem is solved with a hybrid trust-region
semismooth Newton/quasi-Newton method with exact derivatives via adjoint
calculus4,5. Methods:
The proposed algorithm
is implemented in MATLAB (The MathWorks, Inc, Natick, USA) and the calculations
are done in parallel (OpenMP). The iterative optimization starts with a PINS8
RF and slice selective gradient shape based on a SLR7 sub-pulse with
a flip angle of 90° for excitation and 180° for refocusing and inversion. The
optimization framework is tested on 39 different examples covering a wide range
of parameters (multiband-factor MB=2-8, slice thickness THK=1-5mm and time-bandwidth-product
TBWP=2-8). The spatial
domain is fixed to 120mm with varying spatial discretization points (961-4801) and
a fixed temporal discretization of 10µs with varying initial pulse durations (6.82ms-32.8ms).
The amplitude constraints were fixed to RFmax=12.5µT, Gsmax=30mT/m
and slew ratemax=180T/m/s, respectively. The slice profile
constraints are set to a maximum magnitude error of 1% and an optional maximum
phase deviation of 0.025rad. The optimized RF and slice selective gradient
shapes are implemented on a 3T MR scanner (Magnetom Skyra, Siemens Healthcare,
Erlangen, Germany). High-resolution GRE, SE and inversion recovery GRE scans (TR/TE=100-500ms/10-20ms,
FOV=125x125mm, matrix=640×640) of a homogeneous cylinder (T1/T2≈102/77ms) with
phase encoding in slice-direction validate the numerical results.Results and Discussion:
Figure
1 shows the pulse durations of the used PINS initializations (row 1) together
with 39 time optimal results (row 2) for varying MB factor (3-9), TBWP (2-8)
and THK (1-5mm). The average durations are heavily reduced across all optimized
results from initial 11.1ms to 4.23ms (excitation), 17.56ms to 3.99ms
(refocusing) and to 3.81ms (inversion). The achieved reduction therein results
from an exploitation of prescribed inequality constraints such as maximal RF,
slice selective gradient (Gs) and slew rate amplitudes as well as slice profile
magnitude and phase constraints and guarantee practical applicability of the
optimized waveforms. Contrary to the initial expectations, the pulse duration
of the optimized excitation pulses is close to the optimized 180° examples where
a more sensitive out-of-slice error limited a further time reduction. Figure 2 shows
one representative optimized excitation example (MB=3, THK=2mm, TBWP=4).
Depicted are the simulated excitation and the measured spoiled GRE slice
profile. A standard trapezoidal refocusing gradient refocused the linear phase
angle of the excitation profile (not shown). Figure 3 shows one optimized
refocusing example (MB=5, THK=2mm, TBWP=4) with a matching excitation pulse and
compares the Bloch simulation to the corresponding crushed SE measurement. Figure
4 shows one optimized inversion example (MB=9, THK=2mm, TBWP=4) with a matching
excitation pulse and compares the numerical with the experimental results. All
measurements reproduce the Bloch simulations and demonstrate the applicability
of the optimized waveforms. Figure 5 shows the simulated slice profiles for a B1
variation of 85-125% and a B0 offset range of ±200Hz for the previously
shown results. In particular, the results show an increased robustness of the
optimized excitation pulse (row 1) compared to the similar refocusing and
inversion efficiency.Conclusion:
The
proposed framework designs minimum-duration RF and Gs shapes
for the most commonly used RF pulse types in the presence of physical and technical
constraints. The optimized results can be combined for
time optimal excitation, refocusing or inversion purposes. The proposed framework serves as a basis for a
direct design of pulse series such as steady state or magnetization preparation
sequences to further reduce effective echo times.Acknowledgements
Supported
by the Austrian Science Fund (FWF): SFB F32‐N18 and
BioTechMed-GrazReferences
1) Barth M, Breuer F, Koopmans PJ, Norris
DG and Poser BA. Simultaneous multislice (SMS) imaging techniques. Magn. Reson.
Med. 75(1):63–81, 2016.
2) Grissom WA, Setsompop K, Hurley SA,
Tsao J, Velikina JV and Samsonov AA. Advancing RF pulse design using an
open-competition format: Report from the 2015 ISMRM challenge. Magn. Reson.
Med. 78(4):1352–1361, 2016.
3) S. Abo Seada, A.N. Price, T. Schneider, J.V. Hajnal and S.J. Malik.
Multiband RF pulse design for realistic gradient performance. Magn.
Reson. Med. Early View, 2018.
4) A. Rund, C.S. Aigner, K. Kunisch, and R. Stollberger. Magnetic
Resonance RF pulse design by optimal control with physical constraints. IEEE
Trans. Med. Imaging, 37(2):461-472, 2018.
5) A. Rund, C.S. Aigner, K. Kunisch, and
R. Stollberger. Simultaneous multislice refocusing via time optimal control.
Magn. Reson. Med. 80(4):1416-1428, 2018.
6) C.S. Aigner, A. Rund, S. Abo Seada, S.
Malik, J.V. Hajnal, K. Kunisch and R. Stollberger. Time-optimal control based
RF pulse design under gradient imperfections. Proc. Intl. Soc. Mag. Reson. Med.
26, 0209, Paris, France, 2018.
7) Pauly J, Le Roux P, Nishimura D, and
Macovski A. Parameter relations for the Shinnar-Le Roux selective excitation
pulse design algorithm. IEEE Trans. Med. Imaging 10(1), 1991.
8) Norris DG, Koopmans
PJ, Boyacioglu R and Barth M. Power Independent of Number of
Slices (PINS) radiofrequency pulses for low-power simultaneous multislice
excitation. Magn. Reson. Med. 66(5):1234–1240, 2011.