We have used UTE sequence to obtain the subject-specific susceptibility distribution, which was then used to simulate motion-induced B0 change at two head positions. A Fourier-based dipole-approximation method was used to map susceptibility to B0. We have evaluated the simulation results against the measured B0 at the same positions and observed a good agreement between the simulated and real data.
Echo Planar Imaging(EPI) and balanced Stead-Stated Free Precession(bSSFP) are limited in their application for real-time fMRI by both subject motion and main magnetic field(B0) inhomogeneities. Motion-related B0 variations result from changing position and orientation of susceptibility interfaces relative to the B0 field; this causes ghosting and ringing artifacts in structural imaging and time-series phase instability in functional scans. Alternatively, the B0 field inhomogeneities have to be determined for each large-scale movement of the imaged object during acquisitions1. However, it is impractical to acquire the B0 field for every possible positions and orientations of the object. Previously, field probes have been used to correct for this, but they cannot fully estimate B0 in the brain as they are outside the brain2
Here, we estimate the motion-related B0 variations using a Fourier-based dipole-approximation method3-5 and combine with ultrashort echo time(UTE) imaging for computing the subject’s susceptibility model. The UTE sequence allows delineating the cortical bone and air cavities; and thus provides susceptibility models specific to each subject.
UTE sequence was applied to a healthy subject after applying the scanner’s second-order Spherical Harmonic (SH) global shimming. We took 192 slices of dual-echo UTE(1.5mm3voxel; TE1=0.05ms; TE2= 2.46ms; TR=6ms; flip-angle=8°, FOV=288mm3) on a 3T scanner(Prisma, Siemens, Erlangen, Germany). As a reference, off-resonance field map was measured using dual-echo GRE sequence(1.5mm3voxel; TE1=6.66ms; TE2=9.12ms; TR=1630ms; flip-angle=60°, FOV=288x288x192mm). The reference field map, “Measured ΔB0 (x,y,z)”, was calculated by measuring the phase accrued between two echo times at each image voxel.
A 3-classes(air, cortical bones, and soft tissues) UTE based susceptibility model was processed through 3 steps, as shown in Figure.1. First the magnitude images at the first(TE1) and second(TE2) echo times were used to calculate the air mask. An empirically determined threshold was chosen to segment the air cavities6. Then the bone and soft tissue were segmented using inverse of the transverse effective relaxation rate(R2*) estimated from TE1 and TE2, where cortical bone has high R2* values(R2*bone≥0.3ms-1) and soft tissue expected to have low R2* values(0ms-1<R2*soft-tissue<0.3ms-1)7. Finally, the air mask was multiplied back to the R2* map(R2*air=0ms-1) and corresponding magnetic susceptibilities $$$(\chi_{soft-tissue}\approx-9.2ppm, \chi_{bone}\approx-11.4ppm, \chi_{air}\approx0.36ppm)$$$8,9 were assigned to the R2* map.
The simulated off-resonance field map, “Simulated ΔB0 (x,y,z)”, was calculated using a Fourier-based method(Eq.1),
$${Estimated}\triangle{B_0^{z}}({\bf{x}})=\underbrace{FT^{-1}\left\{B_0\left[\frac{1}{3}-\frac{k_z^2}{k_x^2+k_y^2+k_z^2}\right]\cdot{\tilde{\chi}({\bf{k}})}\right\}}_{simulated\triangle{B_0^z}}+B_{in} \quad{(1)}$$
Where the tilde denotes a 3-dimensional Fourier transform of susceptibility model and k indicates k-space vector. Susceptibility is weighted by a k-space scaling factor(the terms in brackets)10. The Bin is measured background inhomogeneities, We measured Bin in a spherical phantom with identical SH shimming setting as measured ΔB0. The resulting “estimated ΔB0 (x,y,z)” is the sum of simulated ΔB0 and Bin. The standard deviation of B0(σB0) within a brain mask was used to assess the simulation performance. Image processing and simulations were performed in MATLAB(Mathwork, Natick, MA).
Two different head positions were measured. We used the first position(pos.1) as reference position. The scanner’s second-order SH shimming was calculated and then applied to the first position. For the second position(pos.2), we kept the identical SH shimming values as position one during measurements. Then second position’s field maps were registered to the first position using the FMRIB Software Library package11 of FLIRT12. We calculated the difference between pos.1 and registered pos.2 for “Estimated ΔB0” and “Measured ΔB0”, in order to demonstrate that the B0 field estimated with our method could be used to predict the motion-induced B0 variations
1 D. F. Hillenbrand K. M. L., W. F. B. Punchard , T. G. Reese, and P. M. Starewicz I.Applied Magnetic Resonance.2005; 29:39-64.
2 Gross S, Duerst Y, Vionnet LM, Barmet C and Pruessmann KP, B0-Atlas with Field-Probe Guidance: Application in Real-Time Field Control, In Proceedings of the 24th Annual Meeting of ISMRM, Melbourne, Singapore, 2016. p. 0930
3 Marques J. P. & Bowtell R.Concepts in Magnetic Resonance Part B: Magnetic Resonance Engineering.2005; 25B:65-78.
4 Boegle R., Maclaren J. & Zaitsev M.Magnetic Resonance Materials in Physics, Biology and Medicine.2010; 23:263-273.
5 Koch K. M., Papademetris X., Rothman D. L. et al.Phys Med Biol.2006; 51:6381-6402.
6 Catana C., van der Kouwe A., Benner T. et al.Journal of Nuclear Medicine.2010; 51:1431-1438.
7 Keereman V., Fierens Y., Broux T. et al.Journal of Nuclear Medicine.2010; 51:812-818.
8 Collins C. M., Yang B., Yang Q. X. et al.Magnetic Resonance Imaging.2002; 20:413-424.
9 Hopkins J. A. & Wehrli F. W.Magnetic Resonance in Medicine.1997; 37:494-500.
10 Cheng Y. C., Neelavalli J. & Haacke E. M.Phys Med Biol.2009; 54:1169-1189.
11 Smith S. M., Jenkinson M., Woolrich M. W. et al.NeuroImage.2004; 23 Suppl 1:S208-219.
12 Jenkinson M., Bannister P., Brady M. et al.NeuroImage.2002; 17:825-841.
13 Sostheim R, Maclaren J, Testud F, Zaitsev M. In Proceedings of the 20th Annual Meeting of ISMRM, Melbourne, Australia, 2012, p.3386