Dinil Sasi S1, Anup Singh1,2, Rupsa Bhattacharjee1,3, Ayan Debnath1,4, Snekha Sehrawat1, Rakesh K Gupta5, Indrajit Saha3, and Marc Van Cauteren6
1Indian Institute of Technology Delhi, New Delhi, India, 2AIIMS, New Delhi, India, 3Philips India Limited, Gurugram, India, 4University of Pennsylvania, Philadelphia, PA, United States, 5Fortis Memorial Research Institute, Gurugram, India, 6Philips Health Tech Asia Pacific, Tokyo, Japan
Synopsis
Parallel-imaging and compressed-sensing based approaches are playing
crucial role in accelerating MRI data acquisition. Objective of the study was
to accelerate the data acquisition of T1, T2 and PD-weighted
TSE images and to evaluate the accuracy of T1 and T2
mapping in the human brain. Data was acquired using SENSE parallel-imaging and
Compressed-SENSE technique for different factors as well as without any
acceleration. T1 and T2 values obtained using data with SENSE
(upto factor of 3) and CSENSE (upto factor of 6) were comparable to those
acquired without any acceleration. Errors in T1 and T2 increased
with increase in acceleration factor.
Introduction
Spin lattice relaxation time constant (T1) and spin-spin
relaxation time constant (T2) are two important parameters required
for quantitative MRI and several methods have been developed for estimating
these parameters individually1,2,3. To estimate these parameters
together, a novel method was previously reported4 that require three
independent T1, T2 and PD-weighted(W) images acquired
using SE/TSE sequence, and is less sensitive to the effect of B1
field inhomogeneity on the relaxivity maps. Conventional images such as T1, T2 and PD-W, based upon TSE sequence, are routinely acquired during
human brain exam; however, takes a long scan time depending upon resolution. Several
methods including parallel imaging6, Compressed Sensing (CS)7
and Compressed Sense (CSENSE)5 are being developed to accelerate
data acquisition. However, the effect of these acceleration techniques on the
accuracy of T1 and T2 maps needs to be evaluated. In this
study, we have evaluated the effect of SENSE parallel imaging and a novel scan
acceleration method by combining compressed sensing technology with the SENSE
infrastructure called Compressed-SENSE (CSENSE) on T1, T2
and PD-W images as well as on T1 and T2 maps.Methods
MRI data of nine human subjects
were acquired at 3.0 T (Ingenia, Philips, The Netherlands) with a 15 channel
head coil. Nine sets of T1
and dual PD-T2 W images were acquired for each subject using TSE
sequence. First data set was acquired without any SENSE and CSENSE (considered
as reference for comparison). Four data sets were acquired with SENSE factors
2,3,4,5 and the other four were acquired with CSENSE factors 2,4,6,8
respectively. Other MRI parameters were:
FOV = 240×240 mm2, number of slices= 20, slice thickness=5mm and
acquisition matrix= 240×240. TE/TR values for dual PD-T2 and T1
W images were 7.2ms/90ms/3500ms and 10ms/360ms respectively. Three subjects
were scanned twice to check for reproducibility.
Data was processed using in-house developed programs in MATLAB. T1
and T2 maps were generated using previously described procedure4
and the white matter (WM) part was segmented out. Normalized mean squared error
(NMSE) was calculated in WM ROI for each combinations to analyze the nature of
error propagation.Results
Data acquisitions of both T1 -W and PD- T2 W with
both SENSE and CSENSE gives similar amount of acceleration in time, which
follows a reciprocal like pattern as shown in Figure-1. T1 and T2
maps obtained using data from SENSE (acceleration factor upto 3) as well as from
CSENSE (acceleration factor upto 6) looked similar to those obtained using
conventional approach as shown in Figure-2 and Figure-3. While going for higher
factors, SENSE factor introduces more noise compared to CSENSE. Visually T1
and T2 maps looks stable till CSENSE factor 6, while noise breakthrough
is clearly visible for maps with SENSE factor 3. Figure-4 shows the mean and
standard deviation of NMER calculated for WM ROI for each subjects. The error propagation
in both T1 and T2 maps for different SENSE and CSENSE
factor follow an exponential pattern. On increasing the CSENSE factor, the
error stay below 1% for both T1 and T2 map until an
acceleration factor 6. While for SENSE acquisition, the error stay below 1%
till an acceleration factor 3. The error propagation is observed to be more for
SENSE acquisition when compared to CSENSE. Results were also reproducible.Discussion
Use of SENSE and CSENSE approaches
in data acquisition results in a substantial reduction in data acquisition
time. Results of current study show that T1 and T2 maps
of human brain obtained using SENSE factor upto 3 and CSENSE factor upto 6 are
comparable to those obtained using conventional data acquisition approach. Moreover,
maps corresponding to SENSE or CSENSE approaches are also reproducible.
Therefore, acceleration gained in data acquisition using these factors can be
used either to reduce data acquisition time or to further increase spatial
resolution at same scan time. In the current study, T1 values were
comparable to reported literature values; however, T2 values show
slightly higher values compared to reported T2 values using other
methods. However, both T1 and T2 values were
reproducible. From the analysis, error propagation is more for SENSE based data
acquisition and more acceleration factor is observed to be stable for CSENSE
acquisition. Conclusion
In conclusion, quality of T1, T2 and PD W TSE images as well as accuracy of T1 and T2 maps derived from these images is not affected by the use of SENSE(upto factor of 3) and CSENSE approach(upto a factor of 6). Therefore, data required for T1 and T2 mapping can be accelerated significantly using CSENSE approach without loss of accuracy compared to conventional approach.Acknowledgements
The Authors acknowledge
technical support of Philips India Limited and Fortis Memorial Research
Institute Gurugram for MRI data acquisition. This work was supported by Science and
Engineering Research Board (IN) (YSS/2014/000092).References
[1]
Crawley, Adrian P., and R. Mark
Henkelman. "A comparison of one‐shot and recovery methods in T1
imaging." Magnetic resonance in medicine 7.1 (1988):
23-34.
[2]
Evans, Avery J., et al.
"Evaluation of steady and pulsatile flow with dynamic MRI using limited
flip angles and gradient refocused echoes." Magnetic resonance
imaging 5.6 (1987): 475-482.
[3]
Kay, I., and R. M. Henkelman.
"Practical Implementation and Optimization of One‐shot T1 imaging." Magnetic
resonance in medicine 22.2 (1991): 414-424.
[4]
Singh A, Singh A, Haris M, Rathore D, Purwar A, Sarma M, Bayu G, Husain N,
Rathore RK, Gupta RK. Quantification of physiological and hemodynamic indices
using T(1) dynamic contrast-enhanced MRI in intracranial mass lesions. J
Magn Reson Imaging. 2007; 26: 871-880.
[5] Geerts-Ossevoort, L., de Weerdt, E.,
Duijndam, A., van IJperen, G., Peeters, H., Doneva, M., Nijenhuis, M. and
Huang, A., Speed done right. Every time. Philips Healthcare; 2018 Jan.
Report No: 4522 991 31821. https://www.philips.de/content/dam/b2bhc/de/resourcecatalog/landingpages/ingeniaelition/White_Paper_Compressed_SENSE-opt.pdf
[6]
Pruessmann, K. P., Weiger, M.,
Scheidegger, M. B., & Boesiger, P. (1999). SENSE: sensitivity encoding for
fast MRI. Magnetic resonance in medicine, 42(5),
952-962.
[7]
Michael Lustig, David Donoho, and John M. Pauly. Sparse MRI: The Application of
Compressed Sensing for Rapid MR Imaging. Magnetic Resonance in Medicine. 2017;
58:1182–1195 (2007)
[8]
Heidemann, R.M., Özsarlak, Ö.,
Parizel, P.M., Michiels, J., Kiefer, B., Jellus, V., Müller, M., Breuer, F.,
Blaimer, M., Griswold, M.A. and Jakob, P.M., 2003. A brief review of parallel
magnetic resonance imaging. European radiology, 13(10),
pp.2323-2337
[9]David
S. Smith, E. Brian Welch, Xia Li, Lori R. Arlinghaus, Mary E. Loveless, Tatsuki
Koyama, John C. Gore and Thomas E. Yankeelov. Quantitative effects of using
compressed sensing in dynamic contrast enhanced MRI. Phys Med Biol., 2011;
56(15): 4933–4946
[10]
SoHyun Hana, Jeffrey L. Paulsenb, Gang Zhuc, Youngkyu Songd, SongI Chund,
Gyunggoo, Chod, Ellen Ackerstaffe, Jason A. Koutchere, and HyungJoon Choa.
Temporal/spatial resolution improvement of in vivo DCE-MRI with compressed
sensing-optimized FLASH. Magn Reson Imaging, 2012; 30(6): 741–752