Willy Gsell1, Cesar Molinos2, Carlos Correcher3, Sven Junge3, Michael Heidenreich3, Uwe Himmelreich4, Christophe M Deroose5, and Antonio J Gonzalez6
1KU Leuven, Leuven, Belgium, 2Nuclear Molecular Imaging, Bruker BioSpin Preclinical Imaging Division, Bruker Biospin, Ettlingen, Germany, 3Nuclear Molecular Imaging, Bruker BioSpin Preclinical Imaging Division, Bruker BioSpin, Ettlingen, Germany, 4Biomedical MRI group, Department of Imaging and Pathology, KU Leuven, Leuven, Belgium, 5Nuclear Medicine and Molecular Imaging, Department of Imaging and Pathology, KU Leuven, Leuven, Belgium, 6Instituto de Instrumentación para Imagen Molecular (I3M), Universitat Politècnica de València, Valencia, Spain
Synopsis
While we already reported
indexes of the performance of one of the first commercially available SiPM-based
MR-compatible PET Insert, no NEMA testing has been
reported until now. Moreover the
NEMA protocol for evaluating the performance of preclinical PET scanner, do not
cover the integration and potential interference between PET and MRI. We hereby
propose and report an extended NEMA protocol demonstrating the performance of
the MRI (Biospec USR 70/30) with and without the presence of the PET Insert as
well as the performance of the PET with simultaneous MRI acquisition.
Introduction:
While many preclinical PET
insert have already been developed and tested1-5, there is still no
consensus in what metrics should be reported to evaluate the performance of the
PET and MRI hardware when operated simultaneously. Based on the NEMA NU 4-2008 protocol6 and on the
MRI acceptance protocol from the manufacturer, we tested one of the first
commercially available SiPM-based MR-compatible high-resolution PET Inserts
inside a pre-clinical 7T MRI scanner. Our aim was to report the performance of
the aforementioned hardware but also to propose a testing protocol as a
guideline for preclinical PET insert testing. Materials and Methods:
MRI performance:
Image quality was assessed with and without the
PET Insert. The test performed comprised signal to noise ratio,
image ghosting, field homogeneity, radiofrequency noise and signal stability.
All MRI sequences and parameters are summarized in table 1.
PET
performance:
All measurements and processing were performed
according to the NEMA NU 4-2008 protocol6. For PET performance characterization,
the Quad86 coil was used for all testing requiring simultaneous PET/MRI
acquisitions.
Sensitivity: 22Na NEMA compliant point source (234
kBq), sensitivity profile along the axial axis of the FOV was determined with
increment steps of 1mm for the first 1 centimeter from the center and then by
step of 5mm until the edge of the FOV. Three acquisitions of 1 minute were acquired
at each position with (EPI and RARE sequences) and without MRI. Background measurements (with
no activity in the PET FOV) were also acquired in the same conditions. Resolution:
22Na point source. Acquisitions of 10 min were performed at FOV center and at 20 mm radial
offset. The images were reconstructed using MLEM, 35 iterations, 0.25 mm voxel
size and a PSF-based partial volume correction. Work is progress to measure the
resolution is all locations as per NEMA standard. Noise
equivalent count rate (NECR): mouse- and rat-sized count-rate phantoms. During
both acquisitions, the PET insert was in position inside the MRI scanner with gradient
powered ON but without radiofrequency coil inside the FOV. PET image
quality: NU4 image quality phantom under different conditions: no MRI, FLASH,
RARE, UTE and EPI acquisitions.
Results:
The field homogeneity was well
preserved when the PET insert was in the MRI FOV (Fig.1). The FWHM linewidth of
the PRESS acquisitions were 6.7 Hz (0.022 ppm) without and with the PET Insert.
The SNR measurement showed no major differences. For the rat quadrature volume coil (86mm internal diameter), SNR changed by
-5.6% and -2% for the short and long TR spin-echo acquisitions, while for the
mouse quadrature volume coil (40mm internal diameter), these values changed by -1.5 and +1.8%. Image ghosting was well within
the manufacturer specifications in all cases (<2.5%) and no RF noise was
visible. The EPI signal stability was ± 0.11 % in both conditions.
Maximum
NEMA sensitivity of the PET insert is 11.0 % at the center of the FOV. Sensibly
the same performance was achieved while MRI was running simultaneously (11.0
and 11.0% for EPI and RARE respectively). PET spatial resolution (FWHM) in
(Transverse; Coronal; Sagittal) planes measured with the NEMA source was (1.09;
1.04; 0.94) mm at the center of the FOV and (1.18; 1.11; 1.09) mm at 20 mm
radial offset. For the EPI sequence, running in parallel the results were (0.96;
1.08; 1.03) mm at the FOV center and (1.14; 1.19; 1.11) mm at 20 mm; and (0.94; 1.17; 1.02) mm at the FOV center and (1.14;
1.19; 1.11) mm at 20 mm for the RARE sequence. The peaks for true and noise equivalent count rates
were 628 kcps at 28 MBq and 486 kcps at 23 MBq respectively in the mouse
phantom whereas for the rat were 410 kcps at 30 MBq and 239 kcps at 23 MBq. Results on the PET image quality are summarized in Figure3. NEMA
uniformity was only slightly altered by the MRI with a maximum difference seen
for the FLASH sequence (% Std of 6.79% versus 5.87% for the MRI-off condition).
Spill-over ratios were similar in all conditions but interestingly the recovery
coefficient for the different rod sizes were slightly improved when MRI was ON.Conclusion and Discussion:
We hereby report the full
characterization of the PET and MRI hardware using an extended NEMA protocol.
The PET insert has only little effect on the MRI performance which were also only minimally affected by the different MRI sequences
retaining a high sensitivity (>11%) and image quality in all conditions
leading to high in-vivo image quality (Fig.4). Further investigations are
currently ongoing to investigate the quantification accuracy with simultaneous
MRI acquisitions.Acknowledgements
No acknowledgement found.References
-
Greg Stortz, Jonathan D. Thiessen, Daryl Bishop et al., Performance of a PET Insert for
High-Resolution Small-Animal PET/MRI at 7 Tesla. The Journal of Nuclear Medicine • Vol.
59 • No. 3 • March 2018, p 536-542.
- David Schug, Christoph
Lerche, Bjoern Weissler et al. Initial PET performance evaluation of a preclinical insert for PET/MRI
with digital SiPM technology. Phys. Med. Biol. 61 (2016) 2851-2878.
- Bjoern Weissler, Pierre Gebhardt, Peter M.
Dueppenbecker et al. A Digital
Preclinical PET/MRI Insert and Initial Results. IEEE Trans Med Imaging.
2015 Nov;34(11):2258-70.
- Parl C., Kolb A., Schmid A. M. et al. A novel optically transparent RF shielding for
fully integrated PET/MRI systems. Phys. Med. Biol. 62 (2017) 7357–7378.
- Negar Omidvari, Jorge
Cabello, Geoffrey Topping et al. PET performance evaluation of MADPET4: a small animal PET insert for a 7
T MRI scanner. Phys. Med. Biol. 62
(2017) 8671–8692
- National Electrical Manufacturers Association. NEMA
Standard Publication NU 4-2008: Performance Measurements of Small Animal
Positron Emission Tomographs. Rosslyn, VA: National Electrical Manufacturers Association;
2008.