Lena V. Gast1, Anke Henning2, Bernhard Hensel3, Michael Uder1, and Armin M. Nagel1,4
1Institute of Radiology, University Hospital Erlangen, Friedrich-Alexander-Universität Erlangen-Nürnberg (FAU), Erlangen, Germany, 2Max Planck Institute for Biological Cybernetics, Tübingen, Germany, 3Center for Medical Physics and Engineering, Friedrich-Alexander-Universität Erlangen-Nürnberg (FAU), Erlangen, Germany, 4Division of Medical Physics in Radiology, German Cancer Research Center (DKFZ), Heidelberg, Germany
Synopsis
B0 inhomogeneities are a major challenge at ultra-high
magnetic field strengths (B0 ≥ 7 T), as they can lead
to strong image artifacts. In this work, a B0 shimming routine that
is based on 23Na B0 maps was implemented for applications
in X-nuclei imaging where no 1H MRI data can be acquired and
therefore, no vendor-provided shimming routines can be used. The proposed 23Na
ConsTru shimming routine showed an improvement in B0 homogeneity
comparable to the vendor-provided GRE Brain shimming routine both in phantom and in vivo
measurements using
23Na B0 map acquisition times less than 1 minute.
Introduction
At ultra-high
magnetic field strengths (B0≥ 7 T), B0
inhomogeneities are a major challenge as they can lead to strong image
artifacts. Especially for quantitative measurements as
well as advanced acquisition techniques such as multiple quantum filtration,1,2 a homogeneous B0 field is indispensable. In X-nuclei imaging, B0
shimming is usually performed using 1H MRI based shimming routines provided
by the manufacturers of the MRI system. However, if an X-nuclei RF coil without
1H channel is used, B0
shimming usually cannot be performed using vendor-provided shimming routines.
The aim of
this work was to implement a B0 shimming routine that is based on 23Na
B0 maps and to evaluate its performance.Methods
The implemented B0 shimming routine consists of four steps:
- 23Na image acquisition using a double-echo 3D density-adapted radial readout (DA-3D-RAD),3 resulting in two phase images $$$\Phi_1$$$ and $$$\Phi_2$$$ corresponding to echo times TE1 and TE2.
- Phase unwrapping 4 of the images ($$$\Phi_{1,unwrapped}$$$, $$$\Phi_{2,unwrapped}$$$).
- Calculation of the B0 deviation map according to $$\Delta B_0 = \frac{\Phi_{2,unwrapped}- \Phi_{1,unwrapped}}{\gamma_{Na}\left(TE_2-TE_1\right)}\qquad\text{(Eq. 1)}$$ with the
gyromagnetic ratio of sodium ($$$\gamma_{Na} = $$$11.27 MHz/T).
- Solution of the shim problem 5 $$\left(A\cdot C\right)\cdot b = B_0\qquad\text{(Eq. 2)}$$ Here, $$$A$$$ is
the matrix of the ideal shim fields described by spherical Harmonics, $$$C$$$ is the decomposition coefficient
matrix modeling the real shim fields,5 $$$b$$$ is the vector of the shim currents to be determined and $$$B_0$$$ is the map calculated in
Step 3. To solve Eq. 2,
the ConsTru algorithm as proposed by Nassirpour et al. 6 was chosen
due to its robustness with respect to low SNR data as expected for in vivo 23Na B0
maps.
Measurements
were performed at a 7 T Magnetom Terra system (Siemens Healthineers, Erlangen,
Germany) equipped with third order shim coils. For the validation of the
implemented shimming routine, a double-resonant 32Na/1H
head coil (Rapid Biomedical, Rimpar, Germany) was used. B0 maps
of a spherical phantom containing 137 mM NaCl in 5% Agarose were acquired both
with the 23Na DA-3D-RAD sequence and a 1H GRE-B0-mapping
sequence. Parameters: 23Na:
TR = 50 ms, TE1/2 = 0.3/5.8 ms, FA = 49°, nominal spatial resolution
Δx= (5 mm)3, Gaussian Filter; 1H: TR = 304 ms, TE1/2 = 2.99/4.60 ms, FA = 17°, Δx = (4 mm)3,
50 slices, FOV = 240x240x200 mm3,
TAcq = 38 s. The dependency of the 23Na shim on the B0
map acquisition duration was examined by varying the number of radial
projections between 500 and 12,000, corresponding to acquisition times of 25 s
to 10 min. To assess the 23Na shimming results, the vendor-provided GRE Brain shimming routine was used as reference (TR = 4.3 ms, TE1/2 =
1.02/3.06 ms, FA = 10°, Δx = (4.4 mm)3, 52 slices, FOV =
282x282x274 mm3, TAcq = 10 s). Additionally, shimming with
B0 maps acquired using 23Na MRI (TAcq = 50 s
and 5 min) and shimming using the vendor-provided shim were performed on a
healthy volunteer. Reconstruction and post-processing of the sodium data sets was performed using MATLAB (TheMathworks, Natick, USA).
Results
B0
maps of the spherical phantom acquired using 23Na MRI show similar
quality for all acquisition times (10 min to 25 s; Figure 1). Even when
acquiring B0 maps with strong radial undersampling, reasonable B0
maps can be reconstructed. 1H and 23Na B0 maps
acquired using the Tune Up shim values, as well as shim values calculated using
the 1H GRE Brain shim and the 23Na ConsTru shim algorithm
can be found in Figure 2. For the 23Na shim, B0 maps
with acquisition durations of 10 min and 25 s were used. B0
variations over the phantom were reduced from 0.99 µT to 0.40 µT using the
vendor-provided 1H GRE brain shim. The 23Na MRI based
ConsTru shim showed a similar performance (0.85 µT to 0.26 µT (TAcq
= 10 min)/0.32 µT (TAcq = 25 s)). Results for the in vivo measurement are shown in Figure
3. Again, the 23Na ConsTru algorithm results in comparable
homogeneity as the GRE Brain shim, even for the short acquisition time (50 s).Discussion and Conclusion
B0
maps acquired with 23Na imaging show a B0 variation similar
to 1H B0 maps and can therefore be used for the
calculation of shim values. With the proposed 23Na ConsTru shimming
routine, an improvement of the B0 homogeneity comparable to the vendor-provided
GRE Brain shim routine could be achieved. Even when reducing the acquisition
time of the 23Na B0 maps to less than 1 minute, similar
ΔB0 distributions and resulting shim values were found. These
results are promising for future applications, where no 1H MRI data
can be acquired (e.g. single-tuned X-nuclei RF coils). Acknowledgements
No acknowledgement found.References
- Gast LV, Gerhalter T, Hensel B, Uder M, Nagel AM. Double quantum filtered 23Na MRI with magic angle excitation of human skeletal muscle in the presence of B0 and B1 inhomogeneities. NMR in
Biomed 2018;e4010.
- Tanase
C, Boada FE. Triple‐quantum‐filtered imaging of sodium in presence of B0
inhomogeneities. J Magn Reson 2005; 174(2):270‐278.
-
Nagel AM, Laun FB, Weber MA, et al. Sodium MRI using a density-adapted
3D radial acquisition technique. Magn Reson Med 2009; 62:1565–1573.
- Maier F, Fuentes D, Weinberg JS, Hazle JD, Stafford RS. Robust Phase Unwrapping for MR Temperature Imaging Using a Magnitude-Sorted List, Multi-Clustering Algorithm. Magn Reson Med 2015; 73(4):1662–1668.
-
Chang P,
Nassirpour S, Henning A. Modeling Real Shim Fields for Very High Degree (and
Order) B0 Shimming of the Human Brain at 9.4 T. Magn Reson Med 2018;
79:529-540.
-
Nassirpour
S, Chang P, Fillmer A, Henning A. A Comparison of Optimization Algorithms for
Localized In Vivo B0 Shimming. Magn Reson Med 2018; 79:1145-1156.