In this abstract a semi-automated verification and validation protocol for a dB/dt system are presented. A dB/dt coil was simulated, fabricated, and mapped with both a hall probe and dB/dt probe using a positioning robot. All measured values, both DC and dB/dt, matched the expected values over the exposure volume. In addition, dB/dt measurements in air and in saline verify the consistency of the magnetic field in a typical device testing media.
A large-bore split solenoid dB/dt coil was modeled using methods previously described by Martire et al. [2] and fabricated in our laboratories to be able to handle all ISO 10974 requirements for gradient (dB/dt) testing. The dimensions were set to have a coil inner winding radius of 24.8 cm and height of 34.1 cm to maximize the exposure region within the coil. The coil was a split design with 24 windings on layer 1 and 23 on layer 2, on each side of a 7.7 cm gap. The DC magnetic field produced by the system (when carrying DC current from a 50 A DC power supply (6032A, Agilent, California, USA)) was measured using a calibrated, NIST-traceable 3-axis Hall probe (F3A-03A05F-A50M25KM, Senis AG, Zug, Switzerland) using 1 cm increments to obtain a magnetic field map and the corresponding coil efficiency map.
At frequencies less than approximately 1 MHz, the magnetic field within an electromagnet is simply the DC field map modulated by the current waveform; however, to verify this and demonstrate the dB/dt values produced within the system, the dB/dt coil was connected to a gradient amplifier (PCI 2100) and a 270 Hz waveform was generated with a calculated dB/dt of 35 T/s peak. The dB/dt field was mapped with a single loop inductive sense coil, radius of 1 cm, fabricated on PCB followed by an instrumentation amplifier with two stages to provide differential amplification, followed by a difference amplifier stage to remove the common-mode voltage. The amplifier was designed to have a gain of 100 and overall probe antenna factor determined to be 30.83 (T/s)/ V. The probe was encased in a 3D printed holder and sealed with a silicone coating to allow for additional mapping inside saline. Field probe data was collected using an oscilloscope to determine the peak dB/dt at each location inside an empty 21 cm diameter phantom. Measurements were repeated with the phantom filled with saline (conductivity 0.47 S/m) to ensure the dB/dt magnetic field remains consistent within the test medium.
For all field mapping, a 3 axis in-house positioning robot was used to automatically move a probe (either Hall probe or dB/dt probe) in 1 cm increments throughout the volume of interest. The robot was equipped with three Nema 23 stepper motors and driven with a controller (DMC-4133, Galil, California, USA). A custom Labview control software was designed to automatically record the time domain signal for each corresponding location which was further analyzed in Matlab (2018b) for comparisons against simulations. Figure 1 shows the field mapping robot and gradient coil setup.
[1] ISO/TS 10974. Assessment of the safety of magnetic resonance imaging for patients with an active implantable medical device; 2018
[2] DJ Martire et al. Design of an MRI Gradient Field Exposure System for Medical Device Testing. ISMRM 2017. Abstract # 4334.