Johannes B Erhardt1, Thomas Lottner2, Jessica Martinez3, Ali C Özen2, Martin Schuettler4, Thomas Stieglitz5, Daniel B Ennis6, and Michael Bock2
1University of Freiburg, Freiburg, Germany, 2University Medical Center Freiburg, Freiburg, Germany, 3University of California Los Angles, Los Angeles, CA, United States, 4CorTec, Freiburg, Germany, 5Dep. of Microsystems Engineering, University of Freiburg, Freiburg, Germany, 6Stanford University, Stanford, CA, United States
Synopsis
Patients with implanted electrodes often
require an MRI exam. This exposes them to the risk of radio-frequency induced
heating. In this study heating of implantable electrodes of diameters between
0.3-4mm were evaluated in a 1.5T MRI system. In situ temperature measurements were compared to simulations of
the specific absorption rate to assess local and total dissipated power.
Measurements showed temperature increases between 0.8-53K. Compared to large
electrodes, small electrodes are subject to less dissipated power, but more
localized power density. Thus, smaller electrodes might be classified as safe
in current certification procedures but may be more likely to burn tissue.
Introduction
Planar multi-electrode arrays are commonly implanted onto
the brain surface to record the electro-corticogram of epilepsy patients. Accurate
localization of the implanted electrodes with respect to the brain anatomy is often
done using MRI. However, the radio frequency (RF) excitation during MRI can
cause severe temperature increase (ΔT) in tissue surrounding the electrodes1.
In the recent years there was a trend towards smaller electrode diameters, but their
RF-induced heating properties in MRI have not been investigated. Therefore, in
this work 4 electrode strips with different electrode diameters were
investigated to identify geometrical factors influencing RF-induced heating.Methods
Four different electrode diameters
(0.3, 1, 2.7 and 4mm) were analyzed using four electrode strips. Each strip had
four electrodes of the same diameter connected by meandering interconnection
lines (MIL), which are connected to 34cm-long PtIr(90/10) wires that were
dip-coated in silicone rubber for insulation. The sample holder was placed into
an MRI phantom as described in the ASTM-F2182 standard² and filled with 30L of
hydroxyethyl cellulose gel with a conductivity of 0.49S/m. Reproducible sample placement was enabled
by a LEGO (LEGO, Denmark) plate glued to the bottom of the phantom and
consistent placement of the sample holder. Temperature measurements were
performed using a fibre-optic thermometer (Optocon, Germany) with 0.1K
resolution. The distance of the probe tip to the electrode was varied between 0
and 3mm. Figure1-AB shows the experimental setup for phantom experiments and the
fiber optic temperature probes (FOTP). Heating experiments were performed in a 1.5T
MRI system (Tim-Symphony, Siemens). A patient weight of 75kg was assumed and
the energy deposition adjusted to 1W/kg. The most distal electrode was placed
at the center of the bore with 11,3cm lateral offset. Finite difference time
domain simulations of the E-field were performed using Sim4Life (Version3.4, ZMT-AG,
Switzerland) to investigate the local SAR distribution around the electrodes
with and without FOTP. RF excitation was simulated by a harmonic voltage source
(f=64MHz) at the MIL end. The
electrode material was modeled as a perfect electric conductor covered by a
60µm isolating silicone layer (εr=3, σ=0S/m) except for the electrode
contact, all surrounded by a 10x10x17cm³ volume of gel (εr=81, σ=0.47S/m) as in the measurements.
The smallest voxel size was 0.04mm³
and the calculation time was 15-20hours. Figure1-C shows the drawing upon which
simulations of the specific absorption rate (SAR) have been performed on.Results
Figure2 and Figure3 show measurement results of the most
distal electrodes as an example. SAR maps and maximum local SAR obtain from
simulations are shown in Figure4 and Figure5 respectively.Discussion
As expected the temperature decreases with increasing FOTP
distance, except for the two smaller diameters with the FOTP in contact. The
highly localized SAR hot spots predicted by the simulations could not be
detected in the measurements possibly due to the displacement of gel by the
FOTP - which is the same size or larger than the electrode diameter. It is
interesting to note that ΔTmax
of the 4mm electrodes is lower than for the 2.7mm electrodes. A possible
explanation is a larger volume in which the power is dissipated. This relationship
between the input energy and the distribution volume is also seen in the
simulations (Figure5). This suggests that a significantly larger volume is
heated in the vicinity of the larger electrodes while the local SAR is reduced,
whereas SAR is locally concentrated and increased for the smaller electrodes. The
total deposited power is three times higher for the two larger diameters than
for the smallest. The simulations also suggest that for implanted electrodes
most of the energy deposition occurs in a volume equivalent to 1g tissue, and
that this value or even less should be used for safety assessment. Furthermore,
for clinical applications it is necessary to decide whether a large electrode
diameter with an associated higher total energy deposition over a larger area
is preferable over a smaller electrode with a reduced total energy deposition
but a much higher energy density in a small volume.Conclusion
Heating measurements of electrodes with different diameters
(0.3mm-4mm) showed maximum temperature increases between 0.8K and 53K.
Comparing simulations of 0.3mm to 4mm electrode contacts showed 92-fold higher
local energy density, but one third of the total energy dissipation.
Accordingly, the diameter of planar electrodes has a strong influence on the
local distribution and the total amount of dissipated RF-induced heating around
the electrode contact. Small electrodes may thus be more prone to exceed safe
tissue temperature thresholds, while being more likely to stay within
regulatory safety limits that average over a volume.Acknowledgements
This work was financially supported by the BrainLinks-BrainTools
Cluster of Excellence funded by the German Research Foundation (DFG, grant
number EXC 1086) and NIH R21 HL127433 to DBE.References
1 Erhardt,
J.B., Fuhrer, E., Gruschke, O.G., Leupold, J., Wapler, M.C., Hennig, J.,
Stieglitz, T., Korvink, J.G., 2018. Should patients with brain implants undergo
MRI? J. Neural Eng. 15, 041002. https://doi.org/10.1088/1741-2552/aab4e4
2 ASTM standard F 2182-2011a, 2011. ASTM 2182 Standard test method for
measurement of radio frequency induced heating near passive implants during
magnetic resonance imaging. ASTM Int. 1–14. https://doi.org/10.1520/F2182-11A.1.7