Krzysztof Wawrzyn1, Jack Hendriks1, Dereck Gignac1, William B. Handler1, and Blaine A. Chronik1
1The xMR Labs, Department of Physics and Astronomy, Western University, London, ON, Canada
Synopsis
In this work we demonstrate a procedure for the use of a custom-built, semi-automated robotic positioning system to map EM fields in commercially available RF exposure systems and compare the results against FDTD simulations. All field
measurements were obtained independently with calibrated E-field
or H-field probes. Results show that the robot can
effectively operate within the complicated EM environment of the RF exposure
platforms with good agreement between simulated and measured results.
Future work will continue to improve the
techniques discussed here.
INTRODUCTION
Semi-automated
robotic mapping of RF exposure in device test platforms is of great value
within an ISO 17025-compliant test facility for both verification and
validation processes. Establishing and evaluating consistence between measured
fields and simulated fields is critical when either is used as part of a
medical device testing procedure. In this paper, we show results from
proof-of-concept work that demonstrates a procedure for the use of a custom-built,
semi-automated robotic positioning system to map electromagnetic fields within
and around RF exposure systems and quantitatively compare the results against
detailed simulation of those same systems.METHODS
All
measurements were performed on two different RF bench top exposure systems, commercially available as
“Medical Implant Testing Systems”, or MITS 1.5 and 3.0, corresponding to
frequencies of 64 and 128 MHz [1] . The RF exposure parameters for the MITS 1.5
and 3.0 were (respectively): pulse type = sinc2π, duty cycle = 40 %, pulse repetition
rate = 1 kHz, polarization = circular 270 & 90 °, frequency = 63.8 &
127.7 MHz, input power = 50 dBm. All
field measurements were obtained independently with either a calibrated E-field
(EX3DV4) or H-field (H3DV7) RMS probe (EASY4MRI standalone data acquisition
system, [1]). The field probes and data
collection were fully integrated into a custom built automated
robotics system (Figure 1). A 3-axis
stepper motor driven gantry robotic field mapping system built in-house was
used to gather data. Data collection was
taken at points in the unloaded MITS bore at constant spatial increments (3.0
cm) along all directions in an area of XYZ=42×21×42 cm (limit = 50 cm3). A passive algorithm implemented into Labview was
used to pursue arbitrary motion objectives and integration with the data
acquisition system for data collection. Finite difference time domain (FDTD) computations were carried out using
commercially available EMPro 3D EM simulation software [2]. The model was based
on the geometry of the physical coils, without lumped elements, using 48
generic ports with sinusoidal excitation at corresponding frequencies. The
phase of the signal feeding the source was equal to its azimuthal position and
the ports at the same azimuthal position in the two rings were 180-degree out
of phase. The simulated field was normalized using a scaling factor determined
from a fixed reference H-field location. MATLAB was used to linearly
interpolate 3D data onto a 1 mm grid (scatteredInterpolant) and measured data
was compared to simulated data using a linear regression model. The reported R-squared values
approximate the amount of variation in measurements that is explained by our
simulated model.RESULTS AND DISCUSSION
Figure 1 shows
a photograph of the automated robotic system with a field sensor at the
isocenter of the MITS. A representative 2D plane profile of collected measured
and simulated data for MITS 3.0, with percent difference of 4 %, is shown in Figure 2.
An example of a line plot through the middle (0 mm) of the 2D
profile is shown in Figure 3. The global percent differences in both MITS
1.5 and 3.0 ranged from 3 to 25 % and largely depended on the determined scaling
factor. To better quantify the variation
of measured to simulated values, a linear regression fit was applied, as shown
in the example on Figure 4. Table
1 shows reported R-squared values ranging from 0.83 to 0.97. Lower R-squared values are associated higher
variation and may be due to uncertainties in systematic measurement and inaccuracies
in the simulation model. Future work
will (1) investigate the effect of correcting the correlation with a rigid body
transform (2) analyze the accuracy of this method using uncertainty propagation
and budgeting, (3) optimize the workflow, (4) compare mapped SAR values within
an ASTM phantom [3], and (5) validate the methods against those described in
ISO DTS10974 [4].CONCLUSION
In this work, we have presented a protocol and results
from proof-of-concept work showing that a semi-automated robotic system can effectively
operate within the complicated EM environment of the RF exposure platforms with
good agreement between simulated and measured results.Acknowledgements
This
work was funded by NSERC, Ontario Research Fund, and the Canadian Foundation
for Innovation.References
[1]
(ZMT/Speag, Zurich, Switzerland). [2] (Keysight, Santa Rosa, CA). [3] ASTM F2182-11a. [4] ISO/TS 10974:2018(E).