Benson Yang1, Pei-Shan Wei1, and Simon J Graham1,2
1Physical Sciences, Sunnybrook Research Institute, Toronto, ON, Canada, 2Medical Biophysics, University of Toronto, Toronto, ON, Canada
Synopsis
Magnetic resonance imaging (MRI)
offers many benefits for deep brain stimulation (DBS) patients. However, radiofrequency
(RF) heating remains a concern for MRI at 3T. Parallel transmission technology (pTx)
has the potential to lower the risk by generating low electric-field regions at
target locations. As pTx technology proceeds to imaging DBS patient, the impact
of head motion becomes an important consideration. The present study
investigates the impact of phantom mis-positioning when using pTx technology to
suppress RF heating with results indicating head motion can potentially cause unsafe
temperature elevations.
Introduction
Deep
brain stimulation (DBS), where therapeutic electrical impulses are delivered
through implanted leads to deep brain nuclei, is an effective treatment to
improve quality of life for patients with neurological disorders (eg.
Parkinson’s Disease). DBS treatments utilize MRI for precise electrode
positioning and to evaluate post-operative effects. Functional MRI of DBS
patients is also of research interest to study and help optimize the mechanism
of effect. However, electrical charge can build up along the leads and causes localized
radiofrequency (RF) heating effects, potentially damaging nearby tissue. Consequently,
there is no regulatory approval for clinical MRI of DBS patients at 3 T and MRI
safety remains an active topic of research in this context. A previous phantom study1
shows that parallel RF transmission (pTx) technology has the potential to produce
a "suppression mode" that strongly attenuates electric-field strength
to lower the risks of RF heating at specific target locations along DBS leads. Another
study shows that the patient-specific lead trajectory is an important factor in
the locations and amounts of RF heating2. As pTx technology proceeds
to imaging DBS patients at 3 T, it is important to consider how the utility of
suppression mode is affected if the zone of minimal electric field is
mis-positioned away from the intended target. This could potentially occur due
to inaccurate positioning of the head during patient setup, or due to head
motion during imaging. The first of these two issues is examined here using 4-channel
pTx MRI of a phantom at 3 T, using the suppression mode method developed by McElcheran
et al.1.Methods
The
effect of head displacement was studied using a custom 4-channel pTx system3
with 4 kW of peak output power, integrated on a research-dedicated 3 T MRI
system (Siemens Magnetom Prisma). PTx suppression mode inputs to each channel were
determined by previous FEKO (Altair Engineering Inc., Troy, USA)
electromagnetic simulation results (amplitude/phase: ch1: -4.0dB/251⁰
ch2: -1.9dB/0⁰, ch3: 0dB/126.6⁰ and ch4: -1.3dB/112.2⁰) on
a uniform head-shaped gel phantom with an insulated cooper wire implant3.
First, MRI demonstrations of pTx technology were conducted (turbo spin echo
scans, TR/TE/FA = 516ms / 6.7ms / 150⁰, 3m 45s scan time) in
circular-polarized (CP) mode to induce RF heating, and in pTx suppression mode for
reduced heating. This was followed by imaging in pTx suppression mode, where
the head phantom was mis-positioned separately in all ‘six degrees of freedom’ (eg.
x, y, z, yaw, pitch and roll directions), as depicted in Fig. 1 for the maximum
displacement available in the pTx coil. Temperature measurements were recorded
at the tip of the implanted wire using a fibre-optic temperature sensor (Opsens
Inc., Quebec City, QC).Results
Fig.
2 summarizes the heating results that were obtained. RF heating in CP mode
resulted in a temperature increase at the tip of the wire of 2.6 ± 0.3⁰C
(from 23.0⁰C to 25.6⁰C), whereas suppression mode
produced an increase of only 0.2 ± 0.3⁰C (from 23.1⁰C
to 23.3⁰C). Head phantom mis-positioning (as per Fig. 1),
recorded temperature changes ranging from 0 to 0.8 ± 0.3⁰C.
Figs. 3 and 4 display MR images of the head phantom reconstructed in pTx
suppression mode at the reference position and for a -50 ± 5⁰ pitch
rotation that resulted in the highest temperature elevation, respectively.
Despite that different cross-sections of the phantom are depicted in these two
figures, it is evident that Fig. 4 shows substantially more RF coupling.Discussion and Conclusion
The
pTx optimization method provides good suppression of RF heating at the desired
target, with reasonable image uniformity. Promising robustness to
mis-positioning was also demonstrated. A maximum temperature change of 0.8 ±
0.3⁰C
was obtained for a large rotation in the pitch direction, and overall the
displacements investigated were much larger than is likely to occur in clinical
practice. However, strong RF coupling was observed in Fig. 4 along the wire,
and it should be emphasized that temperature elevations in clinical practice
are expected to be patient-specific, pulse sequence-specific and complex.
Furthermore, it was observed that thermal equilibrium was not reached and
temperature is likely to rise further with longer scan duration. The present
study recorded temperature at a tip of the wire but other locations along its
trajectory are susceptible to RF heating if the suppression mode is
mis-positioned. Considering the FDA limit of 1.0⁰C heating for head MRI and
the results presented in this study, head motion effects may pose imaging risks
and further research will be required to investigate and establish safe pTx MRI
protocols for DBS patients.Acknowledgements
No acknowledgement found.References
[1]
McElcheran et al., “Parallel Radiofrequency Transmission at 3 Tesla to Improve
Safety in Bilateral Implanted Wires in a Heterogeneous Model” MRM 2017
[2]
Golestanirad et al., “RF-induced heating in tissue near bilateral DBS implants
during MRI at 1.5 T
and
3T: The role of surgical lead management” NeuroImage 2018
[3]
Yang et al., “A Prototype Four-Channel Parallel Transmission System to
Investigate MRI Safety at 3 T” ISMRM 2018