Chemical Shift Saturation Recovery (CSSR) MR spectroscopy and imaging are powerful techniques for deriving a number of important physiological pulmonary parameters. We investigated the impact of measurement echo time on the spectral composition of the acquisition and the subsequent numerical data analysis in rabbits. We found that lengthening the TE resulted in a vastly improved separation between the dissolved-phase resonances by suppressing short T2* signal components. As consequence of this phenomenon, the measured septal wall thickness changed with TE by 1.0 ± 0.13 μm/ms. Similarly, we hypothesize that CSSR-derived measures of pulmonary physiology might be field strength-dependent.
Imaging experiments were performed in 4 sedated New Zealand rabbits (3.5 – 4.5 kg). Animals were ventilated with room air until imaging began, at which point the gas mix was switched to 20% oxygen and 80% HXe for 3 breaths (6 ml/kg tidal volume) followed by a 10-s breath-hold. All studies were approved by the Institutional Animal Care and Use Committee.
Gaussian RF pulses (1.2-ms duration) were applied during the CSSR acquisition to saturate the dissolved-phase resonances around 200 ppm. Following a variable delay time ranging from 2.5 to 500 ms, a 0.9-ms Gaussian RF excitation pulse was used to generate a free induction decay. This sequence was repeated 40 times during a single breath hold. The signal was sampled for 30.72 ms with 1024 sampling points, apodized by a squared cosine function and zero-filled to 2048 points. The native echo time of the measurement was 0.55 ms. However, to simulate the effect of an increase in echo time, a corresponding number of sampling points were discarded prior to Fourier transformation and phasing of the spectra. The dissolved- and gas-phase resonances were integrated numerically. The dissolved-to-gas-phase ratio was subsequently fitted to an analytical gas-uptake model9. All MR studies were performed at 1.5T (Avanto; Siemens), using a custom xenon-129 transmit/receive birdcage coil (Stark Contrast, Erlangen, Germany). Enriched xenon gas (87% xenon-129) was polarized using a prototype commercial system (XeBox-E10, Xemed LLC, Durham, NH).
Figure 1 shows representative dissolved-phase spectra for delay times of 10 ms (Fig. 1A) and 500 ms (Fig. 1B) reconstructed with 8 different echo times. Typically, only a single dissolved-phase resonance can be resolved in xenon spectra of the rabbit lung, but with increasing echo time a second peak emerged approximately 8 ppm downfield from the more prominent tissue-plasma resonance at 197 ppm. This phenomenon is much more apparent at longer delay times, most likely because the dissolved-phase resonance is a composite peak consisting of signal contributions from various locations and with different T2*. At long delay times, the dissolved-phase signal has a larger component of xenon that has been carried downstream by the blood to magnetically more homogeneous regions. As the TE is increased, the short T2* components are selectively suppressed, revealing the long T2* signal components underneath.
As illustrated by the TE-dependence of the derived apparent septal wall thickness in Fig. 2, lengthening the echo time to suppress the short T2* signal components, which presumably originate closer to the magnetically highly inhomogeneous gas exchange sites, has a considerable impact. In our preliminary measurements, we found the septal wall thickness to increase with TE by 1.0 ± 0.13 μm/ms. Using this slope to extrapolate the measurement curves in Fig. 2, we were able to obtain the “true” septal wall thickness for a hypothetical zero-TE acquisition. A corollary of our findings would be that the measured septal wall thickness should also be dependent on the main field strength due to its effect on T2*. On the positive side, deliberately varying the echo time of spectroscopic measurements might help to reveal or more precisely quantify additional xenon resonances that are obscured by strong but rapidly decaying dissolved-phase signal sources in the lung.