Laurie J Smith1,2, Guilhem J Collier1, Helen Marshall1, Paul J.C Hughes1, Alberto Biancardi1, Graham Norquay1, Jody Bray1, Oliver Rodgers1, Martin Wildman3, Noreen West2, Alex Horsley4, and Jim Wild1
1POLARIS, Academic Radiology, University of Sheffield, Sheffield, United Kingdom, 2Sheffield Children's Hospital NHS Foundation Trust, Sheffield, United Kingdom, 3Sheffield Teaching Hospitals NHS Foundation Trust, Sheffield, United Kingdom, 4Respiratory research group, Division of infection, immunity and respiratory medicine, University of Manchester, Manchester, United Kingdom
Synopsis
Ventilation MRI using hyperpolarised gases is
highly sensitive to early cystic fibrosis lung disease. With the recent move
towards 129Xe from 3He, we aimed to assess both gases in
parallel in patients with CF, at baseline and follow-up, to determine whether
any inherent bias was present when assessing lung disease. We found that there
was no inherent bias for VDP between the two gases present although at an
individual level differences were evident. Despite this, when followed up at a
later date both gases similarly reflected changes in ventilation, suggesting
both are capable of reflecting CF lung disease severity.
Introduction
Ventilation MRI using inhaled hyperpolarised
helium-3 (3He) and xenon-129 (129Xe), is highly sensitive
to early lung disease in cystic fibrosis (CF)1,2 and offers a promising method for quantitative assessment of regional lung
disease. There has been a recent move towards 129Xe as a clinically
viable ventilation-imaging agent, however the two gases have not been assessed
in parallel in patients with CF lung disease. The gases have different inherent
properties and therefore will not necessarily provide the same results3. Here
we assess a cohort of children and adults with a range of CF lung disease using
both hyperpolarised 3He and 129Xe MRI on the same day and
at a second visit approximately one year later. We aimed to evaluate the
quantitative information derived from both gases and assess their relative
sensitivities to longitudinal change.Methods
31 children and adults were recruited from three
specialist CF centres in the UK. Hyperpolarised 3He and 129Xe
ventilation MRI were performed on the same day using a bSSFP sequence on a 1.5T
GE HDx scanner. 3He voxel resolution ranged from 3.3*3.3*5 - 4*4*5mm3,
whilst 129Xe ranged from 3.3*3.3*10 - 4*4*10mm3. 3He
sequence parameters; bandwidth=167kHz, TE/TR=0.6/1.9ms, flip angle=10°. 129Xe bandwidth=16kHz,
TE/TR=2.2/6.7ms, flip angle=10°. Ventilation imaging was acquired by inhaling a
titrated volume of either 3He or 129Xe with a balance of
nitrogen, from a lung volume of functional residual capacity. The bag volume
ranged from 0.4-1.0L and was calculated based on subjects’ height. For 3He
acquisition, a 1H anatomical MR image was acquired during the same
breath-hold in order to calculate the thoracic cavity volume and for 129Xe
this was performed in a separate breath-hold. For quantitative analysis of the
images, the 1H and ventilation images were segmented using a semi-automated
method4, from which the ventilation defect percentage
(VDP), the number (Ndefects) and the size of individual ventilation
defects in 3D and the ventilation heterogeneity index (VHI) were
calculated as recently described5. At the same visit patients also performed
spirometry and multiple breath washout (using 0.2% SF6) in order to
calculate lung clearance index (LCI). These methods were then repeated at a
second visit at least one year later.Results
Baseline demographics, MRI metrics and lung function
are detailed in Table 1. 14 subjects were followed up at a second time point at
a median (range) duration of 1.3 (1.1,1.8) years. At baseline 3He
and 129Xe ventilation images were qualitatively similar in all but
one subject (Figure 1). There was no significant group difference between 3He
and 129Xe MRI for VDP (p>0.05) and there was a strong correlation
between gases (r=0.97, p<0.001). Bland-Altman analysis (3He-129Xe)
showed good agreement with minimal bias (bias=0.7%, limits of agreement (LoA)=8.9,-7.4%
(Figure 2)). Similarly B-A analysis for Ndefects (bias=-0.2,
LoA=-6.8,6.4) and the largest individual defect (bias=0.3%, LoA=-5.1,5.7%) demonstrated
no significant bias. VHI however was significantly higher for 3He
when compared to 129Xe (p<0.001) (B-A bias=1.4%, LoA=-2.6,5.3%). At
baseline VDP and VHI from both gases demonstrated similar strong
correlations with LCI and FEV1 (Figure 3). Of the 14 subjects
followed up (using a 1% change in VDP as a threshold), for both gases four had increased
VDP, three had reduced VDP and two had no change in VDP. Four subjects demonstrated
change in VDP for 129Xe only (two increased, two decreased) and only
one subject demonstrated reduced VDP for 3He despite increased VDP
for 129Xe. As seen at baseline, qualitative assessment of change in
ventilation distribution with time in an individual was similar for both gases
(Figure 4).Discussion/Conclusions
3He
and 129Xe ventilation MRI in CF give similar qualitative results
with minimal quantitative difference. For patients with normal FEV1,
VDP values tend to be lower in 129Xe when compared to 3He,
however similar ventilation abnormalities are evident for both gases. Some of
these differences relate to gas diffusivity (D0 in air = 0.86/0.14cm2s-1
for 3He/129Xe respectively6) and to image resolution. 129Xe
image slices are double the thickness of 3He (10mm vs 5mm), despite
the same in-plane resolution, and therefore smaller unventilated defects are
more likely to appear unventilated for 3He. VDP and VHI
for both gases strongly correlate with lung function markers of disease and
therefore both reflect CF lung disease severity. Further to this both gases demonstrated
similar quantitative and qualitative change at follow-up, suggesting that
despite any subtle differences at a single time point, both gases are
reflecting the underlying level of lung disease. Acknowledgements
The authors would like to acknowledge all members of the
POLARIS research group at the University of Sheffield for the support. In particular
we would like to thank Mrs Leanne Armstrong for administrative support. We would also like to
thank the Cystic Fibrosis clinical teams at Sheffield Children’s Hospital,
Sheffield Teaching Hospital and Manchester CF Centre for their support. Finally
we would like to thank all of the participants for their time in taking part in
this research.References
1. Marshall H, Horsley A, Taylor CJ, et al. Detection of early subclinical
lung disease in children with cystic fibrosis by lung ventilation imaging with
hyperpolarised gas MRI. Thorax
2017;72(8):760-62. doi: 10.1136/thoraxjnl-2016-208948
2. Thomen RP, Walkup LL, Roach DJ, et al.
Hyperpolarized (129)Xe for investigation of mild cystic fibrosis lung disease
in pediatric patients. Journal of cystic
fibrosis : official journal of the European Cystic Fibrosis Society
2017;16(2):275-82. doi: 10.1016/j.jcf.2016.07.008
3. Stewart NJ, Chan HF, Hughes PJC, et al.
Comparison of (3) He and (129) Xe MRI for evaluation of lung microstructure and
ventilation at 1.5T. J Magn Reson Imaging
2018 doi: 10.1002/jmri.25992
4. Hughes PJC, Horn FC, Collier GJ, et al.
Spatial fuzzy c-means thresholding for semiautomated calculation of percentage
lung ventilated volume from hyperpolarized gas and (1) H MRI. J Magn Reson Imaging 2018;47(3):640-46.
doi: 10.1002/jmri.25804
5. Smith LJ, Collier GJ, Marshall H, et al.
Patterns of regional lung physiology in cystic fibrosis using ventilation MRI
and MBW. European Respiratory Journal
2018. In Press. DOI: 10.1183/13993003.00821-2018
6. Chen XJ, Moller HE, Chawla MS, et al.
Spatially resolved measurements of hyperpolarized gas properties in the lung in
vivo. Part I: diffusion coefficient. Magn
Reson Med 1999;42(4):721-8.