4107

Feasibility of 1 mm isotropic 3D T2-weighted breast imaging accelerated using Compressed SENSE
Yajing Zhang1, Jiazheng Wang2, and Yishi Wang2

1Philips Healthcare, Suzhou, China, 2Philips Healthcare Greater China, Beijing, China

Synopsis

We demonstrated the feasibility of 5.3 times acceleration for 3D T2-weighted breast imaging with a 1 mm isotropic resolution at 1.5 T, using Compressed SENSE. This technique showed either better image quality or enhanced imaging speed when compared to traditional SENSE with different imaging schemes.

Introduction

Compressed sensing is one of the state-of-the-art acceleration technologies for MRI acquisition and reconstruction [1]. Compressed SENSE technology is built on the combination of compressed sensing and SENSE [2] theories by leveraging a balanced under-sampling acquisition scheme with variable density and incoherent sampling pattern [3]. Iterative reconstruction is used to solve an inverse problem with the priori information of coil sensitivities and a sparsity constraint. This study aimed to demonstrate the feasibility of compressed SENSE for 3D T2-weighted breast imaging.

Method

The whole-breast imaging was performed on an Ingenia Prodiva 1.5 T MR scanner (Philips Healthcare, Suzhou, China) on healthy volunteers with informed consent, study approved by a local IRB. A 7-channel phase-arrayed breast coil was used for a 3D turbo spin echo (TSE) acquisition with TSE factor 66. The imaging FOV = 360x360 mm2, voxel size = 1mm isotropic, number of slices = 200, and TE/TR = 170/1300 ms. Compressed SENSE was applied with 5.3 times acceleration to enhance the imaging efficiency (hereafter CS-SENSE). To examine the image quality with comparable acceleration, scans were also performed with 1) 5 times SENSE acceleration in phase encoding direction (hereafter SENSE-1), 2) 2.6 and 2.0 times SENSE acceleration in both phase encoding and slice encoding directions (hence a total acceleration factor of 5.2, hereafter SENSE-2), and 3) 2.6 and 2.0 times SENSE acceleration in both phase encoding and slice encoding directions with oversampling in the slice encoding direction (hereafter SENSE-3). Other imaging parameters were the same as abovementioned.

Results

Imaging time for CS-SENSE, SENSE-1, SENSE-2, and SENSE-3 were 2m39s, 2m40s, 2m32s, and 5m17s, respectively. Figure 1 illustrates a comparison of CS-SENSE and SENSE-1 using the same net acceleration factor of about 5. Severe aliasing artifact was observed in the SENSE-1 image while not significant in the image from CS-SENSE. In-plane artifacts were reduced in SENSE-2 when compared to that in SENSE-1, while the through-plane artifact started to present (Figure 2).

Discussion

Our results showed that, under the same acceleration factor of about 5, 3D T2-weighted images can be obtained using compressed SENSE without observable imaging artifacts, at a relatively low field (1.5 T) with a moderate quality coil (7 channels). At the same time, it posed a challenge for traditional SENSE to achieve such a high acceleration factor, where the acceleration factor is heavily limited by the number and alignment of coil elements. When the SENSE acceleration was distributed into two directions, folding-artifacts started to appear in the slice direction, which could be removed by oversampling in the same direction at the cost of doubled imaging time. A further study should be performed to verify the clinical significance of compressed SENSE for T2-weighted breast imaging.

Conclusion

For 3D T2-weigted breast imaging, image acquisition can be highly accelerated by compressed SENSE at 1.5 T.

Acknowledgements

No acknowledgement found.

References

[1] Lustig M, Donoho D, Pauly JM. Sparse MRI: the application of compressed sensing for rapid MR imaging. Magn Reson Med. 2007; 58: 1182-1195

[2] Pruessmann KP, Weiger M, Scheidegger MB, Boesiger P. SENSE: sensitivity encoding for fast MRI. Magn Reson Med. 1999; 42: 952–962

[3] Bratke G, Rau R, Weiss K, et al. Accelerated MRI of the lumbar spine using compressed sensing: quality and efficiency. J Magn Reson Imaging 2018; DOI: 10.1002/jmri.26526

Figures

Figure 1: (A) Compressed SENSE with 5.3 times acceleration; (B) SENSE with 5 times acceleration in the phase encoding direction. In-plane folding artifact was observed in the SENSE image (arrowed).

Figure 2: (A) Compressed SENSE with 5.3 times acceleration; (B) SENSE with 2.6 times acceleration in the phase encoding direction and 2.0 times in the slice encoding direction. Through-plane artifact was observed in the SENSE image (yellow arrow).

Proc. Intl. Soc. Mag. Reson. Med. 27 (2019)
4107