Kristen Zakian1, Andreas Wibmer2, Hebert Vargas Alvarez2, Eveline Alberts3, Mo Kadbi3, Borys Mychalczak4, Marisa Kollmeier4, Daniel Gorovets4, Sean McBride4, Margie Hunt1, Michael Zelefsky4, and Neelam Tyagi1
1Medical Physics, Memorial Sloan Kettering Cancer Center, New York, NY, United States, 2Radiology, Memorial Sloan Kettering Cancer Center, New York, NY, United States, 3Philips Medical Systems, Best, Netherlands, 4Radiation Oncology, Memorial Sloan Kettering Cancer Center, New York, NY, United States
Synopsis
At our institution, a Foley catheter is inserted during prostate
SBRT simulation to ensure visualization of the prostatic urethra. Our goal was to find an optimal motion-robust
T2-weighted pulse sequence that would allow for accurate catheter-free
visualization and segmentation of the urethra. We compared Multivane XD (MVXD),
a PROPELLER-based multislice TSE sequence with SENSE acceleration to single-shot
TSE with SENSE (SSTSE). Two expert GU radiologists scored urethra visibility in
MVXD and SSTSE series of equivalent spatial resolution and duration. Urethra
visibility scores were significantly higher in the MVXD series and SNR was
found to be superior in all MVXD series.
Introduction
High radiation doses employed for prostate radiotherapy can
result in urethral side effects. Stereotactic body radiation therapy (SBRT)
employs very high radiation doses in limited fractions and may further increase
the probability of urethral toxicity1. At our institution, a Foley
catheter is inserted during the simulation process for prostate SBRT patients to
ensure visualization of the prostatic urethra for contouring. However, catheterization is particularly unpleasant
for patients. Our goal was to find an optimal motion-robust T2-weighted pulse
sequence that would allow for accurate catheter-free visualization and
segmentation of the urethra. While isotropic T2-weighted 3D-MRI is excellent
for high-resolution soft tissue differentiation, artifacts in two dimensions due
to respiration, bladder and bowel motion can obscure the urethra. Therefore, we
compared Multivane XD (MVXD), a 2D multislice radial fast-spin-echo sequence based
on PROPELLER2 where SENSE3 parallel imaging acceleration
is incorporated within each blade to single-shot turbo spin echo with SENSE (SSTSE).Methods
All scans were performed on a 3T Philips Ingenia (Philips,
Best, The Netherlands) equipped with a flat tabletop for MR simulation. Twelve prostate
cancer patients underwent MRI as part of simulation for conventionally
fractionated or moderately hypofractionated prostate radiotherapy. Our standard procedure is to require Foley
catheterization for patients undergoing ultra-hypofractionated therapy while urethral
location for patients treated with other fractionations is estimated using
other anatomical features.
Standard imaging
series include large FOV 3D Dixon-based T1-weighted gradient-echo,
small FOV multislice axial T2 fast-spin-echo, and sagittal multislice fast spin
echo or 3D balanced fast field echo (bFFE). As patients are scanned with a full
bladder, exam time must be limited and the MVXD and SSTSE sequences were
optimized to provide sub-millimeter in-plane resolution and adequate SNR in 2-3
minutes. Parameters common to both sequences were: FOV = 220 mm2-300
mm2, in-plane resolution = 0.9x0.9 mm2, slice thickness =
2.5 mm, TR = 3500 ms, TEeff = 80 ms. MVXD parameters included: 35 echoes per shot/blade, 1 average, percent
of Cartesian sampling = 220, SENSE factor = 1.8. SSTSE parameters included echo
train length 79-96 and SENSE factor = 2.0. Two-averages were used for the SSTSE
to provide comparable SNR compared to MVXD.
Two
radiologists with GU experience > 10 years and > 5 years assessed the MVXD
and SSTSE series. A 1-5 scale was used
to score visibility of the prostatic urethra where 1 = urethra not visible and
5 = urethra completely visible. In addition, readers were asked to comment on
signal-to-noise ratio and the presence of artifacts.
Results
Figure 1 contains MVXD and SSTSE series for subject 2. Both
readers gave MVXD a higher urethral visibility score than SSTSE (4 vs. 3 and 5
vs. 2). As seen in Table 1, both readers scored urethra visibility higher in MVXD images than in SSTSE images in 10/12 subjects. In the remaining 2 subjects, urethral
visibility was good (score ≥3) for
either method as ranked by both readers. SSTSE was rated slightly superior (4
vs 3) in 1 case (Reader 2). Using a paired Student’s t-test for comparison, urethra
visibility scores were significantly higher in the MVXD series (See table). Qualitatively,
both readers rated MVXD as having superior SNR in all patients. Readers noted
streak artifacts in 7/12 and 6/12 MVXD data sets, respectively; however, these
were near the periphery and did not interfere with urethra visualization. No
artifacts were noted in the SSTSE images. Despite identical effective echo
times, the different sampling schemes resulted in subtle differences in T2
contrast; this did not affect urethra visualization.Discussion
Single shot fast spin-echo is widely used for motion-insensitive
T2-weighted imaging, but can be SNR-limited. We found that two averaged single-shot
TSE acquisitions were needed to achieve sufficient SNR and spatial resolution, increasing the potential for between-shot
motion. Optimized Multivane XD provided superior urethra visualization in 10 of
12 cases. Limiting scan time to 2-3 minutes and incorporating SENSE acceleration of 1.8 permitted sampling of 220% over conventional Cartesian k-space, mitigating streaking artifacts and providing superior SNR. In this patient group, motion artifacts were not observed in either series suggesting that the SNR advantage of MVXD was the source of enhanced urethra visibility.Conclusions
Mutlivane XD, a PROPELLER-based pulse sequence with SENSE encoding,
provided superior visibility of the prostatic urethra over a single-shot TSE
sequence of equivalent resolution and duration. At our institution, Multivane
XD has been incorporated into the MR-only simulation workflow for the prostate,
and elimination of the Foley catheter is being studied.Acknowledgements
The authors would like to thank the MRI technologists and
therapists in the MSKCC Department of Radiation Oncology for their valuable
assistance.References
1. Repka, MC, Guleria, S. et. al. Acute Urinary Morbidity Following Stereotactic Body Radiation Therapy for
Prostate Cancer with Prophylactic Alpha-Adrenergic Antagonist and Urethral Dose
Reduction. Front Oncol 2016 May 18; 6:122.
2. Pipe, JG.
Motion correction
with PROPELLER MRI: application to head
motion and free-breathing cardiac imaging. Magn Reson Med 1999 Nov;42(5):963-9.
3. Pruessmann, KP, Weiger, M., et. al. SENSE: Sensitivity Encoding for
Fast MRI. Magn Reson Med 1999 Nov; 42(5): 952-962.