Jiahui Li1, Bogdan Dzyubak1, Kevin J. Glaser1, Ziying Yin1, Jun Chen1, Alina Allen2, Sudhakar K. Venkatesh1, Armando Manduca1, Vijay Shah2, Richard L. Ehman1, and Meng Yin1
1Radiology, Mayo Clinic, Rochester, MN, United States, 2Gastroenterology, Mayo Clinic, Rochester, MN, United States
Synopsis
While conventional MRE is easily performed during
suspended respiration, a free breathing MRE technique would be advantageous in
pediatric and other patients. We
assessed the repeatability of free-breathing MRE in comparison to conventional
breath held acquisitions in volunteers and the clinical performance of
free-breathing MRE in a cohort of 56 patients with chronic liver diseases.
Results demonstrated comparable repeatability and excellent agreement of the
averaged liver stiffnesses measured using the two techniques. In summary, free-breathing MRE provides highly
repeatable and accurate liver stiffness values.
Introduction
Liver MR elastography (MRE), a quantitative imaging method for noninvasively assessing hepatic
fibrosis, is conventionally performed using a breath-held, multi-slice, 2D, spin-echo
echo-planar imaging (SE-EPI) or gradient-recalled echo (GRE) acquisition1. Both SE-EPI and GRE MRE
are performed using expiration breath holds to avoid respiratory motion
artifacts in the images and because end-expiration is considered a more
reproducible breath-hold location than end-inspiration. However, some patients have
difficulty performing adequate end-expiration breath holds (e.g., pediatric and
sedated patients). Preliminary studies have demonstrated that a non-gated,
free-breathing, single-shot, multi-slice 2D EPI-MRE technique with a view-sharing-based
reconstruction strategy can
eliminate the need for breath holds2 and
is capable of using the respiratory cycle to measure liver stiffness and other third-order
mechanical parameters that may be helpful in disease diagnosis3. The
purpose of this study is to evaluate the
test-retest repeatability of this rapid,
free-breathing MRE technique and to assess the accuracy of the liver stiffness measurements
compared to conventional breath-held MRE.Methods
All
activities related to human subjects were reviewed and approved by our
institutional review board. Six patients and one healthy volunteer underwent
test-retest scans to evaluate the repeatability of free-breathing MRE. The exam
consisted of a breath-held 2D SE-EPI MRE scan, a free-breathing 2D SE-EPI
MRE scan, and a conventional Dixon acquisition for hepatic fat quantification. The
exam was performed and then the subject was removed from the scan room,
repositioned, and rescanned approximately 15 minutes later. The same data were
collected in another cohort of 56 patients with chronic liver disease without the retest scan to evaluate
the clinical performance of free-breathing MRE. The examinations were carried out on multiple whole-body
GE 1.5T scanners (Excite HDxt and Optima MR 450w; GE Healthcare, Waukesha, WI). In this implementation of free-breathing
MRE, a 4-offset MRE data set was collected every 4 seconds and repeated 25
times (~1.5 min acquisition time), allowing for a view-shared reconstruction of
100 liver stiffness images over ~20-30 respiratory cycles. We calculated the baseline liver stiffness
(LSB) from the conventional breath-held MRE stiffness images,
and the average liver stiffness (LSA) and stiffness variation ΔLS (ΔLS=75%-25% quantiles) from the 100 free-breathing
stiffness images, within anatomically
co-registered ROIs (Figure 1). The
repeatability coefficients (RC)4 and intra-class correlation
coefficients (ICC) were used to analyze the test-retest repeatability of
breath-held and free-breathing MRE. Spearman correlations were used to evaluate the relationships
between LSB
and LSA, fat fraction, and ΔLS. We also classified the
patients into two groups, one group with fatty liver disease and the other
group with other chronic liver diseases (alcoholic hepatitis, cirrhosis,
primary sclerosing cholangitis, etc.). For all statistical analysis, a significance level of less
than 0.05 was used and a correlation
coefficient (ρ) greater than 0.90 was considered excellent.Results
Table 1 illustrates that the RC of free-breathing MRE (24.5%) is comparable with that
of breath-held MRE (26%). The ICC of free-breathing MRE (0.9571) is also comparable
with that of breath-held MRE (0.9500). Figure 2 shows example images of a
healthy volunteer and a patient. With Spearman’s correlation analysis, we found
that LSA excellently agrees with LSB (LSA=0.9849×LSB,
ρ=0.9562, p<0.0001*)( Figure 3). Figure 4 illustrates that there is no significant
correlation between fat fraction and ΔLS or LSB in the 25 patients
with fatty liver disease. Discussion
The
RC is a percentage of the measured hepatic stiffness change and indicates that
a measured change larger than that can be considered a true change with 95% confidence.
The RC of 2D MRE in healthy volunteers from QIBA is 19%4, while in
another study in 94 patients it was 25.3%5. In our study, the RC calculated from
free-breathing MRE (24.5%) was similar to breath-held MRE (26%) and was similar
to these other studies. This suggests that different breath-held position could
be a substantial source of measurement variation in breath-held MRE.
Free-breathing MRE has potentials to mitigate this variation if calculating LS
at different respiratory states. The LSA calculated from
free-breathing MRE excellently agrees with LSB from conventional
breath-held MRE in patients. The stiffness variation across the breathing
cycle, ΔLS, did not appear to be related to stiffness or hepatic fat fraction. Additional
studies are required to determine if this variation may be another promising
parameter for patient assessment.Conclusion
Free-breathing MRE has comparable repeatability to conventional breath-held MRE.
It also provides an accurate averaged liver stiffness measurement that has
potential for the assessment of patients with chronic liver diseases. This technique
will be very beneficial for pediatric and
sedated patients and could improve the comfort
and patient experience for the general population as well.Acknowledgements
This study
is funded by NIH grant EB017197, EB001981, and Mayo Clinic Center for Individualized Medicine Imaging Biomarker
Discovery Program.References
1.
Yong Seek Kim, Yu Na Jang, Ji Soo Song. Comparison
of gradient-recalled echo and spin-echo echo-planar imaging MR elastography in
staging liver fibrosis: a meta-analysis. Eur Radiol. 2018;28:1709-1718.
2.
Glaser K, Chen J, Ehman R. Fast 2D hepatic MR
elastography for free-breathing and short breath hold applications. The 23rd Annual Meeting of International
Society of Magnetic Resonance in Medicine (ISMRM); May; Toronto, Ontario,
Canada 2015. p. 6579.
3.
Ziying Yin, Bogdan Dzyubak, Jiahui Li, Kevin
J. Glaser, Sudhakar K. Venkatesh, Armando Manduca, Richard L. Ehman, Meng Yin.
A Feasibility Study of Nonlinear Mechanical Response Assessment of the
Liver with MR Elastography (MRE). The 26rd Annual Meeting of International
Society of Magnetic Resonance in Medicine (ISMRM); June; Paris, France 2018.
p.0606.
4.
QIBA MR Elastography Biomarker Committee. MR
Elastography of the Liver, Quantitative Imaging Biomarkers Alliance. Profile
Stage: Consensus. QIBA, May 2, 2018. Available from: http://qibawiki.rsna.org/index.php/Profiles
5.
Ye ji lee, Jeong Min Lee, Jeong Eun Lee, et
al. MR Elastography for Noninvasive Assessment of Hepatic Fibrosis:
Reproducibility of the Examination and Reproducibility and Repeatability of the
Liver Stiffness Value Measurement. Journal of Magnetic Resonance Imaging.
2014;39:326-331